Impact of application repetition on Joule heating and lesion formation using a circular-shaped pulsed field catheter.

IF 2.6
Mizuki Okumura, Shintaro Yamagami, Masaya Akiyama, Yuta Nakano, Yu Amano, Toyoki Okuda, Tsukasa Motoyoshi, Ryuki Chatani, Hirokazu Kondo, Toshihiro Tamura
{"title":"Impact of application repetition on Joule heating and lesion formation using a circular-shaped pulsed field catheter.","authors":"Mizuki Okumura, Shintaro Yamagami, Masaya Akiyama, Yuta Nakano, Yu Amano, Toyoki Okuda, Tsukasa Motoyoshi, Ryuki Chatani, Hirokazu Kondo, Toshihiro Tamura","doi":"10.1007/s10840-026-02379-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Pulsed field ablation (PFA) is a non-thermal ablation modality, but repeated high-voltage applications can still cause Joule heating at the electrode-tissue interface. We evaluated the effects of application repetition and inter-application intervals on temperature rise and lesion depth using a circular-shaped PFA catheter.</p><p><strong>Methods and results: </strong>In an in vitro potato model, a circular-shaped PFA catheter was tested under five settings: 1×, 2×, 2× with a 10-s interval, 4×, and 4× with a 10-s interval. Maximum electrode surface temperature was measured using a thermal camera, and lesion depth was assessed by TTC staining (n = 5 per setting). Temperature increased significantly with application repetition and was attenuated by a 10-s interval (overall P < 0.0001). Lesion depth also increased with repetition (overall P < 0.0001), but did not differ significantly between interval and no-interval conditions at equal application numbers.</p><p><strong>Conclusion: </strong>In this simplified in vitro model, a short inter-application interval was associated with attenuation of temperature rise during repeated PFA delivery, while lesion depth was not significantly reduced.</p>","PeriodicalId":520675,"journal":{"name":"Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing","volume":" ","pages":"1563-1566"},"PeriodicalIF":2.6000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s10840-026-02379-7","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/23 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Pulsed field ablation (PFA) is a non-thermal ablation modality, but repeated high-voltage applications can still cause Joule heating at the electrode-tissue interface. We evaluated the effects of application repetition and inter-application intervals on temperature rise and lesion depth using a circular-shaped PFA catheter.

Methods and results: In an in vitro potato model, a circular-shaped PFA catheter was tested under five settings: 1×, 2×, 2× with a 10-s interval, 4×, and 4× with a 10-s interval. Maximum electrode surface temperature was measured using a thermal camera, and lesion depth was assessed by TTC staining (n = 5 per setting). Temperature increased significantly with application repetition and was attenuated by a 10-s interval (overall P < 0.0001). Lesion depth also increased with repetition (overall P < 0.0001), but did not differ significantly between interval and no-interval conditions at equal application numbers.

Conclusion: In this simplified in vitro model, a short inter-application interval was associated with attenuation of temperature rise during repeated PFA delivery, while lesion depth was not significantly reduced.

应用重复对焦耳加热和使用圆形脉冲场导管病变形成的影响。
简介:脉冲场烧蚀(PFA)是一种非热烧蚀方式,但重复的高压应用仍然会在电极-组织界面产生焦耳加热。我们使用圆形PFA导管评估了重复应用和间隔应用对温升和病变深度的影响。方法与结果:在离体马铃薯模型中,采用1×、2×、2×(间隔10-s)、4×、4×(间隔10-s) 5种设置对圆形PFA导管进行测试。使用热像仪测量最高电极表面温度,通过TTC染色评估病变深度(每次n = 5)。结论:在简化的体外模型中,较短的间隔时间与PFA重复给药过程中温升的衰减有关,而病变深度没有明显降低。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书