Analytical Evaluation of a Point-of-Care Platform for Glycated Hemoglobin and N-Terminal Pro-B-Type Natriuretic Peptide Testing in Cardiometabolic Care.
Perrin Ngougni Pokem, Ali Khatib, Iman Azariouh Kaddouri, Damien Gruson
{"title":"Analytical Evaluation of a Point-of-Care Platform for Glycated Hemoglobin and N-Terminal Pro-B-Type Natriuretic Peptide Testing in Cardiometabolic Care.","authors":"Perrin Ngougni Pokem, Ali Khatib, Iman Azariouh Kaddouri, Damien Gruson","doi":"10.14740/cr2220","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Heart failure is a frequent complication of diabetes, highlighting the importance of integrated cardiometabolic assessment. Glycated hemoglobin (HbA1c) and N-terminal pro-B-type natriuretic peptide (NT-proBNP) represent complementary biomarkers reflecting glycemic control and myocardial stress. Point-of-care testing (POCT) platforms may facilitate decentralized diagnostics and shorten analytical turnaround time. The aim of this study was to evaluate the analytical performance of the AFIAS-3 point-of-care fluorescence immunoassay platform for HbA1c and NT-proBNP measurements in comparison with established central laboratory methods.</p><p><strong>Methods: </strong>This single-center verification study included 50 patient samples for each analyte. HbA1c measurements obtained using AFIAS-3 were compared with Tosoh G8 high-performance liquid chromatography (HPLC), while NT-proBNP measurements were compared with the cobas 8000 electrochemiluminescence immunoassay. Analytical precision and accuracy were assessed using manufacturer quality control materials. Method comparison was performed using Passing-Bablok regression, Bland-Altman analysis, and Spearman correlation. Agreement at clinical thresholds was evaluated using Cohen's kappa statistics.</p><p><strong>Results: </strong>For HbA1c, analytical precision was excellent, with coefficients of variation of 1.9% and 1.8%. Correlation with the reference method was strong (r = 0.98), and the mean bias was minimal (-0.03%). Agreement at the diagnostic threshold of 6.5% was excellent (κ = 0.92). For NT-proBNP, coefficients of variation were 5.3% and 15.4%. Correlation with the central laboratory method was very high (r = 0.99), although proportional bias was observed at higher concentrations (slope 1.24). Agreement at clinically relevant thresholds remained substantial.</p><p><strong>Conclusions: </strong>The AFIAS-3 POCT platform demonstrated excellent analytical agreement for HbA1c and acceptable agreement for NT-proBNP at clinically relevant thresholds. Dual biomarker POCT may support decentralized cardiometabolic diagnostics, although NT-proBNP values at higher concentrations should be interpreted cautiously.</p>","PeriodicalId":9424,"journal":{"name":"Cardiology Research","volume":"17 3","pages":"190-196"},"PeriodicalIF":1.3000,"publicationDate":"2026-06-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13278697/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiology Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14740/cr2220","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Heart failure is a frequent complication of diabetes, highlighting the importance of integrated cardiometabolic assessment. Glycated hemoglobin (HbA1c) and N-terminal pro-B-type natriuretic peptide (NT-proBNP) represent complementary biomarkers reflecting glycemic control and myocardial stress. Point-of-care testing (POCT) platforms may facilitate decentralized diagnostics and shorten analytical turnaround time. The aim of this study was to evaluate the analytical performance of the AFIAS-3 point-of-care fluorescence immunoassay platform for HbA1c and NT-proBNP measurements in comparison with established central laboratory methods.
Methods: This single-center verification study included 50 patient samples for each analyte. HbA1c measurements obtained using AFIAS-3 were compared with Tosoh G8 high-performance liquid chromatography (HPLC), while NT-proBNP measurements were compared with the cobas 8000 electrochemiluminescence immunoassay. Analytical precision and accuracy were assessed using manufacturer quality control materials. Method comparison was performed using Passing-Bablok regression, Bland-Altman analysis, and Spearman correlation. Agreement at clinical thresholds was evaluated using Cohen's kappa statistics.
Results: For HbA1c, analytical precision was excellent, with coefficients of variation of 1.9% and 1.8%. Correlation with the reference method was strong (r = 0.98), and the mean bias was minimal (-0.03%). Agreement at the diagnostic threshold of 6.5% was excellent (κ = 0.92). For NT-proBNP, coefficients of variation were 5.3% and 15.4%. Correlation with the central laboratory method was very high (r = 0.99), although proportional bias was observed at higher concentrations (slope 1.24). Agreement at clinically relevant thresholds remained substantial.
Conclusions: The AFIAS-3 POCT platform demonstrated excellent analytical agreement for HbA1c and acceptable agreement for NT-proBNP at clinically relevant thresholds. Dual biomarker POCT may support decentralized cardiometabolic diagnostics, although NT-proBNP values at higher concentrations should be interpreted cautiously.
期刊介绍:
Cardiology Research is an open access, peer-reviewed, international journal. All submissions relating to basic research and clinical practice of cardiology and cardiovascular medicine are in this journal''s scope. This journal focuses on publishing original research and observations in all cardiovascular medicine aspects. Manuscript types include original article, review, case report, short communication, book review, letter to the editor.