Calibration of diffusion MRI measurements using aqueous glycerin phantoms with controlled viscosity.

IF 1.6 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Radiological Physics and Technology Pub Date : 2026-09-01 Epub Date: 2026-06-19 DOI:10.1007/s12194-026-01083-5
Mayuka Seguchi, Yuki Kanazawa, Tosiaki Miyati, Masafumi Harada, Mitsuharu Miyoshi, Hiroaki Hayashi, Tetsuya Wakayama, Akihiro Haga
{"title":"Calibration of diffusion MRI measurements using aqueous glycerin phantoms with controlled viscosity.","authors":"Mayuka Seguchi, Yuki Kanazawa, Tosiaki Miyati, Masafumi Harada, Mitsuharu Miyoshi, Hiroaki Hayashi, Tetsuya Wakayama, Akihiro Haga","doi":"10.1007/s12194-026-01083-5","DOIUrl":null,"url":null,"abstract":"<p><p>To characterize the relationship between the apparent diffusion coefficient (ADC) and dynamic viscosity in aqueous glycerin phantoms across different concentrations and temperatures, and to establish a controlled calibration framework for diffusion MRI measurements. Eight glycerin-water solutions (10-60 wt%) were prepared, and viscosities were measured with capillary viscometers at 24, 37, and 40 °C. Diffusion-weighted imaging (DWI) was performed on a 3.0-T scanner using 15 b-values (0-1500 s/mm²). Voxel-wise ADC maps were calculated by mono-exponential fitting, and ADC-viscosity relationships were modeled using linear and inverse fitting approaches. Viscosity increased with concentration and decreased with temperature, consistent with the known physical behavior of glycerin-water mixtures. ADC values were inversely related to viscosity and showed strong relationships at all temperatures (R² > 0.98). High-concentration samples exhibited reduced signal-to-noise ratio and greater fitting instability, limiting reliable analysis in the highest-viscosity range. Aqueous glycerin solutions provide practical phantom materials for evaluating ADC-viscosity relationships under controlled conditions. By anchoring MRI-derived ADC measurements to standard viscometer data, this study established a reproducible calibration framework for diffusion MRI system evaluation across a defined viscosity range.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":"1068-1076"},"PeriodicalIF":1.6000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-026-01083-5","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/19 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

To characterize the relationship between the apparent diffusion coefficient (ADC) and dynamic viscosity in aqueous glycerin phantoms across different concentrations and temperatures, and to establish a controlled calibration framework for diffusion MRI measurements. Eight glycerin-water solutions (10-60 wt%) were prepared, and viscosities were measured with capillary viscometers at 24, 37, and 40 °C. Diffusion-weighted imaging (DWI) was performed on a 3.0-T scanner using 15 b-values (0-1500 s/mm²). Voxel-wise ADC maps were calculated by mono-exponential fitting, and ADC-viscosity relationships were modeled using linear and inverse fitting approaches. Viscosity increased with concentration and decreased with temperature, consistent with the known physical behavior of glycerin-water mixtures. ADC values were inversely related to viscosity and showed strong relationships at all temperatures (R² > 0.98). High-concentration samples exhibited reduced signal-to-noise ratio and greater fitting instability, limiting reliable analysis in the highest-viscosity range. Aqueous glycerin solutions provide practical phantom materials for evaluating ADC-viscosity relationships under controlled conditions. By anchoring MRI-derived ADC measurements to standard viscometer data, this study established a reproducible calibration framework for diffusion MRI system evaluation across a defined viscosity range.

校正扩散核磁共振测量使用含水甘油幻影与控制粘度。
表征不同浓度和温度下甘油水溶液幻影中表观扩散系数(ADC)与动态粘度之间的关系,并建立扩散MRI测量的受控校准框架。制备8种甘油-水溶液(10-60 wt%),在24、37和40°C下用毛细管粘度计测量粘度。在3.0 t扫描仪上使用15个b值(0-1500 s/mm²)进行弥散加权成像(DWI)。通过单指数拟合计算体素ADC图,并使用线性和逆拟合方法建模ADC-粘度关系。粘度随浓度增加而降低,与已知的甘油-水混合物的物理行为一致。ADC值与粘度呈负相关,在所有温度下均表现出很强的相关性(R²> 0.98)。高浓度样品表现出较低的信噪比和较大的拟合不稳定性,限制了在最高粘度范围内的可靠分析。甘油水溶液为在受控条件下评估adc -粘度关系提供了实用的幻影材料。通过将MRI衍生的ADC测量值锚定到标准粘度计数据,本研究建立了一个可重复的校准框架,用于在定义的粘度范围内对扩散MRI系统进行评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书