Ruturaj R. Masvekar, Gabriela Mansano Do Nascimento, Aliyah Hannah A. Remoroza, Shweta Jain, Troy J. Kemp, Bruce K. Brown, Ligia A. Pinto
{"title":"A semi-automated ELISA workflow for SARS-CoV-2 Nucleocapsid IgG detection significantly improves throughput and reduces hands-on time","authors":"Ruturaj R. Masvekar, Gabriela Mansano Do Nascimento, Aliyah Hannah A. Remoroza, Shweta Jain, Troy J. Kemp, Bruce K. Brown, Ligia A. Pinto","doi":"10.1016/j.jim.2026.114075","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Enzyme-linked immunosorbent assay (ELISA) is widely used in clinical and research laboratories for applications such as disease surveillance, immunogenicity assessment, and vaccine evaluation. Traditional manual ELISA workflows, however, are labor-intensive, time-consuming, and susceptible to operator-dependent variability, which can limit throughput and reproducibility. Incorporating semi-automation into established protocols offers a potential strategy to improve efficiency while maintaining analytical performance.</div></div><div><h3>Methods</h3><div>A validated manual ELISA protocol for SARS-CoV-2 Nucleocapsid IgG detection was adapted into a semi-automated workflow using an automated multichannel pipette, microplate washer-dispenser, and plate stacker to streamline key steps. Performance was compared across 10 manual and 30 semi-automated plates. Sixteen pre-characterized serum samples were analyzed to evaluate precision, accuracy, and overall assay validity, and internal controls were monitored to ensure compliance with predefined acceptance criteria.</div></div><div><h3>Results</h3><div>The semi-automated workflow doubled assay throughput from 6 to 12 plates per analyst per day and reduced hands-on time by 75% (from 5.5 h to 1 h). Internal controls for all 40 plates met established acceptance ranges. Precision, measured as relative standard deviation, remained below 25% for all samples, and accuracy, expressed as percent error, remained below 20%. No loss of analytical performance was observed when transitioning from manual to semi-automated processing.</div></div><div><h3>Conclusion</h3><div>Semi-automation markedly increases ELISA throughput and reduces hands-on time while maintaining precision, accuracy, and overall assay validity. These results underscore its feasibility as an efficient workflow enhancement with potential for broader application across diverse ELISA platforms pending further validation.</div></div>","PeriodicalId":16000,"journal":{"name":"Journal of immunological methods","volume":"549 ","pages":"Article 114075"},"PeriodicalIF":1.7000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of immunological methods","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0022175926000517","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/6/7 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Enzyme-linked immunosorbent assay (ELISA) is widely used in clinical and research laboratories for applications such as disease surveillance, immunogenicity assessment, and vaccine evaluation. Traditional manual ELISA workflows, however, are labor-intensive, time-consuming, and susceptible to operator-dependent variability, which can limit throughput and reproducibility. Incorporating semi-automation into established protocols offers a potential strategy to improve efficiency while maintaining analytical performance.
Methods
A validated manual ELISA protocol for SARS-CoV-2 Nucleocapsid IgG detection was adapted into a semi-automated workflow using an automated multichannel pipette, microplate washer-dispenser, and plate stacker to streamline key steps. Performance was compared across 10 manual and 30 semi-automated plates. Sixteen pre-characterized serum samples were analyzed to evaluate precision, accuracy, and overall assay validity, and internal controls were monitored to ensure compliance with predefined acceptance criteria.
Results
The semi-automated workflow doubled assay throughput from 6 to 12 plates per analyst per day and reduced hands-on time by 75% (from 5.5 h to 1 h). Internal controls for all 40 plates met established acceptance ranges. Precision, measured as relative standard deviation, remained below 25% for all samples, and accuracy, expressed as percent error, remained below 20%. No loss of analytical performance was observed when transitioning from manual to semi-automated processing.
Conclusion
Semi-automation markedly increases ELISA throughput and reduces hands-on time while maintaining precision, accuracy, and overall assay validity. These results underscore its feasibility as an efficient workflow enhancement with potential for broader application across diverse ELISA platforms pending further validation.
期刊介绍:
The Journal of Immunological Methods is devoted to covering techniques for: (1) Quantitating and detecting antibodies and/or antigens. (2) Purifying immunoglobulins, lymphokines and other molecules of the immune system. (3) Isolating antigens and other substances important in immunological processes. (4) Labelling antigens and antibodies. (5) Localizing antigens and/or antibodies in tissues and cells. (6) Detecting, and fractionating immunocompetent cells. (7) Assaying for cellular immunity. (8) Documenting cell-cell interactions. (9) Initiating immunity and unresponsiveness. (10) Transplanting tissues. (11) Studying items closely related to immunity such as complement, reticuloendothelial system and others. (12) Molecular techniques for studying immune cells and their receptors. (13) Imaging of the immune system. (14) Methods for production or their fragments in eukaryotic and prokaryotic cells.
In addition the journal will publish articles on novel methods for analysing the organization, structure and expression of genes for immunologically important molecules such as immunoglobulins, T cell receptors and accessory molecules involved in antigen recognition, processing and presentation. Submitted full length manuscripts should describe new methods of broad applicability to immunology and not simply the application of an established method to a particular substance - although papers describing such applications may be considered for publication as a short Technical Note. Review articles will also be published by the Journal of Immunological Methods. In general these manuscripts are by solicitation however anyone interested in submitting a review can contact the Reviews Editor and provide an outline of the proposed review.