[Impact of Source Position Error in Tandem-ovoid Applicator on Dose Distribution in Cervical Cancer Brachytherapy].

Junichi Akagawa, Tatsuhiro Sera, Yuki Yuasa
{"title":"[Impact of Source Position Error in Tandem-ovoid Applicator on Dose Distribution in Cervical Cancer Brachytherapy].","authors":"Junichi Akagawa, Tatsuhiro Sera, Yuki Yuasa","doi":"10.6009/jjrt.26-1651","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to determine the discrepancy between planned and actual source positions in the Geneva applicator for cervical cancer brachy therapy and to evaluate its impact on dose distribution.</p><p><strong>Methods: </strong>The Geneva applicator was placed on a computed tomography (CT) couch top, and CT images were acquired at each source dwell position to determine the actual source locations. The discrepancy between the actual source positions and the planned source positions reconstructed using the applicator modeling (AM) method was calculated. These discrepancies were then incorporated into the treatment plans of 13 patients who underwent brachytherapy for cervical cancer at our institution. Dose metrics, including D<sub>90%</sub> for the high-risk clinical target volume (HR-CTV) and D<sub>2cc</sub> for the bladder, rectum, small bowel, and large bowel, were evaluated between the original and the modified treatment plans.</p><p><strong>Results: </strong>The overall three-dimensional displacement of the source was 0.86±0.15 mm. Regarding dose metrics, the D<sub>90%</sub> for the HR-CTV was significantly higher (mean increase, 1.38%) in the plans reflecting the actual source positions, while the D<sub>2cc</sub> for the bladder and rectum was significantly lower (mean decrease, 2.29% and 2.57%, respectively) compared to the original plans.</p><p><strong>Conclusion: </strong>The results suggest that incorporating actual source positions into treatment planning can improve the accuracy of applicator reconstruction in cervical cancer brachytherapy.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":"82 7","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.26-1651","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study aimed to determine the discrepancy between planned and actual source positions in the Geneva applicator for cervical cancer brachy therapy and to evaluate its impact on dose distribution.

Methods: The Geneva applicator was placed on a computed tomography (CT) couch top, and CT images were acquired at each source dwell position to determine the actual source locations. The discrepancy between the actual source positions and the planned source positions reconstructed using the applicator modeling (AM) method was calculated. These discrepancies were then incorporated into the treatment plans of 13 patients who underwent brachytherapy for cervical cancer at our institution. Dose metrics, including D90% for the high-risk clinical target volume (HR-CTV) and D2cc for the bladder, rectum, small bowel, and large bowel, were evaluated between the original and the modified treatment plans.

Results: The overall three-dimensional displacement of the source was 0.86±0.15 mm. Regarding dose metrics, the D90% for the HR-CTV was significantly higher (mean increase, 1.38%) in the plans reflecting the actual source positions, while the D2cc for the bladder and rectum was significantly lower (mean decrease, 2.29% and 2.57%, respectively) compared to the original plans.

Conclusion: The results suggest that incorporating actual source positions into treatment planning can improve the accuracy of applicator reconstruction in cervical cancer brachytherapy.

[串联卵圆器照射源位置误差对宫颈癌近距离放疗剂量分布的影响]。
目的:本研究旨在确定宫颈癌近距离放射治疗日内瓦照射器计划与实际照射位置的差异,并评估其对剂量分布的影响。方法:将日内瓦涂抹器放置在计算机断层扫描(CT)沙发上,并在每个源驻留位置获取CT图像以确定实际源位置。计算了应用应用器建模(AM)方法重建的实际源位置与规划源位置之间的差异。这些差异随后被纳入13名在我院接受近距离宫颈癌治疗的患者的治疗计划。剂量指标,包括高危临床靶体积(HR-CTV)的D90%和膀胱、直肠、小肠和大肠的D2cc,在原治疗方案和修改后的治疗方案之间进行评估。结果:源体整体三维位移为0.86±0.15 mm。在剂量指标方面,反映实际辐射源位置的方案中,HR-CTV的D90%明显高于原方案(平均增加1.38%),而膀胱和直肠的D2cc明显低于原方案(平均减少2.29%和2.57%)。结论:在宫颈癌近距离放射治疗中,将实际源位置纳入治疗方案可提高应用器重建的准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书