Quantitative accuracy of 177Lu SPECT/CT imaging using ring-shaped CZT versus dual-head NaI systems.

IF 3.5 2区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Rachele Danieli, Clémentine Marin, Anna-Lena Theisen, Johannes Tran-Gia, Carlos Artigas, Bruno Vanderlinden, Nick Reynaert, Patrick Flamen, Hugo Levillain
{"title":"Quantitative accuracy of <sup>177</sup>Lu SPECT/CT imaging using ring-shaped CZT versus dual-head NaI systems.","authors":"Rachele Danieli, Clémentine Marin, Anna-Lena Theisen, Johannes Tran-Gia, Carlos Artigas, Bruno Vanderlinden, Nick Reynaert, Patrick Flamen, Hugo Levillain","doi":"10.1186/s40658-026-00892-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Ring-shaped systems based on CZT are promising for quantitative imaging, but their accuracy may be affected by septal penetration of high-energy photons-given the fixed collimator-and by limitations in scatter correction due to low-energy tailing. This study compares the quantitative performance of a ring-shaped CZT system with that of a conventional dual-head NaI SPECT/CT system (StarGuide™ and NM/CT 870 DR, both GE HealthCare) using the 208 keV photopeak of <sup>177</sup>Lu.</p><p><strong>Methods: </strong>A NEMA phantom with 3D-printed anthropomorphic inserts, including kidneys and lumbar vertebrae, as well as standard spherical inserts, was filled with activity concentrations mimicking the distribution observed in patients 1 day post-injection of 7.4 GBq of [<sup>177</sup>Lu]Lu-PSMA-I&T. The phantom was scanned repeatedly on both systems (scan duration/bed position: 7 min for StarGuide™ and 20 min for NM/CT 870 DR). Quantitative accuracy and reproducibility were evaluated using the mean error and the coefficient of variation (CV) over repeated measurements, respectively. Additionally, a group of patients treated with [<sup>177</sup>Lu]Lu-PSMA-I&T underwent sequential SPECT/CT on both systems to confirm the phantom-based results.</p><p><strong>Results: </strong>The total activity measured within the phantom deviated from the expected value by means of 13% for the StarGuide™ and 0% for the NM/CT 870 DR. After partial-volume-effect correction based on recovery coefficients, the activity concentration in kidney inserts was measured with ~ 5% accuracy on both systems. For the StarGuide™, mean errors in tumour inserts (volumes: 2.6-28 ml) ranged from - 8% to 3%, whereas a consistent underestimation by an average of - 11% was observed for the NM/CT 870 DR. The StarGuide™ significantly overestimated the background activity concentration (59%), while the NM/CT 870 DR showed a smaller error (- 8%). Reproducibility was lower for the StarGuide™, with maximum CV values of 7%, compared with 3% for the NM/CT 870 DR. Clinical findings were consistent with phantom results.</p><p><strong>Conclusions: </strong>The StarGuide™ provided slightly more accurate activity quantification in regions of high relative uptake, such as tumours and kidneys, compared with the NM/CT 870 DR. However, quantification in regions of low relative uptake and total activity estimation were less accurate, likely due to septal penetration. The lower reproducibility suggests that longer scan durations may be considered.</p>","PeriodicalId":11559,"journal":{"name":"EJNMMI Physics","volume":" ","pages":""},"PeriodicalIF":3.5000,"publicationDate":"2026-06-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13493729/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40658-026-00892-2","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Ring-shaped systems based on CZT are promising for quantitative imaging, but their accuracy may be affected by septal penetration of high-energy photons-given the fixed collimator-and by limitations in scatter correction due to low-energy tailing. This study compares the quantitative performance of a ring-shaped CZT system with that of a conventional dual-head NaI SPECT/CT system (StarGuide™ and NM/CT 870 DR, both GE HealthCare) using the 208 keV photopeak of 177Lu.

Methods: A NEMA phantom with 3D-printed anthropomorphic inserts, including kidneys and lumbar vertebrae, as well as standard spherical inserts, was filled with activity concentrations mimicking the distribution observed in patients 1 day post-injection of 7.4 GBq of [177Lu]Lu-PSMA-I&T. The phantom was scanned repeatedly on both systems (scan duration/bed position: 7 min for StarGuide™ and 20 min for NM/CT 870 DR). Quantitative accuracy and reproducibility were evaluated using the mean error and the coefficient of variation (CV) over repeated measurements, respectively. Additionally, a group of patients treated with [177Lu]Lu-PSMA-I&T underwent sequential SPECT/CT on both systems to confirm the phantom-based results.

Results: The total activity measured within the phantom deviated from the expected value by means of 13% for the StarGuide™ and 0% for the NM/CT 870 DR. After partial-volume-effect correction based on recovery coefficients, the activity concentration in kidney inserts was measured with ~ 5% accuracy on both systems. For the StarGuide™, mean errors in tumour inserts (volumes: 2.6-28 ml) ranged from - 8% to 3%, whereas a consistent underestimation by an average of - 11% was observed for the NM/CT 870 DR. The StarGuide™ significantly overestimated the background activity concentration (59%), while the NM/CT 870 DR showed a smaller error (- 8%). Reproducibility was lower for the StarGuide™, with maximum CV values of 7%, compared with 3% for the NM/CT 870 DR. Clinical findings were consistent with phantom results.

Conclusions: The StarGuide™ provided slightly more accurate activity quantification in regions of high relative uptake, such as tumours and kidneys, compared with the NM/CT 870 DR. However, quantification in regions of low relative uptake and total activity estimation were less accurate, likely due to septal penetration. The lower reproducibility suggests that longer scan durations may be considered.

环形CZT与双头NaI系统在177Lu SPECT/CT成像中的定量准确性
背景:基于CZT的环形系统在定量成像方面很有前景,但其精度可能受到高能光子的间隔穿透(给定固定准直器)和低能尾迹对散射校正的限制的影响。本研究使用177Lu的208 keV光峰,比较了环形CZT系统与传统双头NaI SPECT/CT系统(StarGuide™和NM/CT 870 DR,均为GE医疗)的定量性能。方法:在NEMA假体中植入3d打印的拟人植入物,包括肾脏和腰椎,以及标准的球形植入物,填充与注射7.4 GBq [177Lu]Lu-PSMA-I&T后1天观察到的活性浓度分布相似的NEMA假体。在两个系统上重复扫描幻膜(扫描时间/床位:StarGuide™为7分钟,NM/CT 870 DR为20分钟)。分别用重复测量的平均误差和变异系数(CV)评价定量准确性和重复性。此外,一组接受[177Lu]Lu-PSMA-I&T治疗的患者在两个系统上进行了连续的SPECT/CT检查,以确认基于幻像的结果。结果:通过StarGuide™和NM/CT 870 dr,幻影内测量的总活度与期望值的偏差分别为13%和0%。经过基于恢复系数的部分体积效应校正后,两种系统测量的肾植入物活度浓度的准确度均为5%左右。对于StarGuide™,肿瘤植入物(体积:2.6-28 ml)的平均误差范围为- 8%至3%,而NM/CT 870 DR的平均低估幅度为- 11%。StarGuide™显著高估了背景活性浓度(59%),而NM/CT 870 DR的误差较小(- 8%)。StarGuide™的重现性较低,最大CV值为7%,而NM/CT 870 dr的CV值为3%。结论:与NM/CT 870 dr相比,StarGuide™在相对摄取高的区域(如肿瘤和肾脏)提供了略准确的活性量化,然而,在相对摄取低的区域和总活性估计的量化不太准确,可能是由于间隔穿透。较低的再现性表明可以考虑延长扫描时间。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
EJNMMI Physics
EJNMMI Physics Physics and Astronomy-Radiation
CiteScore
6.70
自引率
10.00%
发文量
78
审稿时长
13 weeks
期刊介绍: EJNMMI Physics is an international platform for scientists, users and adopters of nuclear medicine with a particular interest in physics matters. As a companion journal to the European Journal of Nuclear Medicine and Molecular Imaging, this journal has a multi-disciplinary approach and welcomes original materials and studies with a focus on applied physics and mathematics as well as imaging systems engineering and prototyping in nuclear medicine. This includes physics-driven approaches or algorithms supported by physics that foster early clinical adoption of nuclear medicine imaging and therapy.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书