Reliability of a Scheimpflug-based tomographer compared with a Placido-based topographer and specular microscopy in healthy eyes.

IF 1 4区 医学 Q3 OPHTHALMOLOGY
Arquivos brasileiros de oftalmologia Pub Date : 2026-05-25 eCollection Date: 2026-01-01 DOI:10.5935/0004-2749.2025-0259
Camyla Lemos Budib, Caio Henrique Peres Oliani, Guilherme de Oliveira Ferreira, Ulysses Tachibana, José Álvaro Pereira Gomes, Bernardo Kaplan Moscovici, Luiz Roberto Colombo-Barboza, Marcello Novoa Colombo-Barboza, Guilherme Novoa Colombo-Barboza
{"title":"Reliability of a Scheimpflug-based tomographer compared with a Placido-based topographer and specular microscopy in healthy eyes.","authors":"Camyla Lemos Budib, Caio Henrique Peres Oliani, Guilherme de Oliveira Ferreira, Ulysses Tachibana, José Álvaro Pereira Gomes, Bernardo Kaplan Moscovici, Luiz Roberto Colombo-Barboza, Marcello Novoa Colombo-Barboza, Guilherme Novoa Colombo-Barboza","doi":"10.5935/0004-2749.2025-0259","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the reliability and comparability of a Scheimpflug-based tomographer relative to a Placido-based topographer and specular microscopy in healthy eyes.</p><p><strong>Methods: </strong>This cross-sectional study included 40 patients (80 eyes). Each eye underwent randomized imaging with a Scheimpflug-based tomographer, a Placido-based topographer, and Tomey EM-4000 specular microscopy. Three acquisitions per device were obtained. For interdevice comparisons, the best-quality scan per eye/device was selected, whereas all three scans were used for intradevice repeatability analyses. Unreliable scans were repeated (up to five attempts) and excluded if acceptable quality was not achieved, resulting in variable denominators. Between-device comparisons were performed using generalized estimating equations with participant-level clustering and robust standard errors and were supplemented by Bland-Altman analysis.</p><p><strong>Results: </strong>The effective sample size varied by parameter (independent summaries: 59-67 eyes; paired comparisons: 48-51 eyes). In paired-eye analyses, the Scheimpflug-based tomographer measured slightly higher keratometry values than the Placido-based topographer (K1: 43.95 vs. 43.78 D, p=0.003; K2: 44.91 vs 44.73 D, p=0.002), more negative Q-values (p=0.001), smaller photopic pupil diameter (p<0.001), and shorter kappa distance (p<0.001). Mean absolute differences were 0.32 D for K1 and 0.30 D for K2, with high dispersion for angular metrics (kappa angle coefficient of variation: 195%).</p><p><strong>Conclusions: </strong>The Scheimpflug-based tomographer provides reproducible corneal measurements in healthy eyes. However, systematic differences relative to the Placido-based topographer-particularly for keratometry, asphericity, and pupil and kappa metrics-suggest limited interchangeability. Consistent device use is recommended when these parameters inform clinical decision-making.</p>","PeriodicalId":8397,"journal":{"name":"Arquivos brasileiros de oftalmologia","volume":"89 3","pages":"e20250259"},"PeriodicalIF":1.0000,"publicationDate":"2026-05-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arquivos brasileiros de oftalmologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5935/0004-2749.2025-0259","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the reliability and comparability of a Scheimpflug-based tomographer relative to a Placido-based topographer and specular microscopy in healthy eyes.

Methods: This cross-sectional study included 40 patients (80 eyes). Each eye underwent randomized imaging with a Scheimpflug-based tomographer, a Placido-based topographer, and Tomey EM-4000 specular microscopy. Three acquisitions per device were obtained. For interdevice comparisons, the best-quality scan per eye/device was selected, whereas all three scans were used for intradevice repeatability analyses. Unreliable scans were repeated (up to five attempts) and excluded if acceptable quality was not achieved, resulting in variable denominators. Between-device comparisons were performed using generalized estimating equations with participant-level clustering and robust standard errors and were supplemented by Bland-Altman analysis.

Results: The effective sample size varied by parameter (independent summaries: 59-67 eyes; paired comparisons: 48-51 eyes). In paired-eye analyses, the Scheimpflug-based tomographer measured slightly higher keratometry values than the Placido-based topographer (K1: 43.95 vs. 43.78 D, p=0.003; K2: 44.91 vs 44.73 D, p=0.002), more negative Q-values (p=0.001), smaller photopic pupil diameter (p<0.001), and shorter kappa distance (p<0.001). Mean absolute differences were 0.32 D for K1 and 0.30 D for K2, with high dispersion for angular metrics (kappa angle coefficient of variation: 195%).

Conclusions: The Scheimpflug-based tomographer provides reproducible corneal measurements in healthy eyes. However, systematic differences relative to the Placido-based topographer-particularly for keratometry, asphericity, and pupil and kappa metrics-suggest limited interchangeability. Consistent device use is recommended when these parameters inform clinical decision-making.

在健康眼睛中,基于scheimpflug的断层扫描仪与基于placido的地形扫描仪和镜面显微镜的可靠性比较。
目的:评估健康眼睛中基于scheimpflug的断层成像相对于基于placido的地形成像和镜面显微镜的可靠性和可比性。方法:本横断面研究纳入40例患者(80眼)。每只眼睛接受随机成像,分别使用基于scheimpflug的断层扫描仪、基于placido的地形扫描仪和Tomey EM-4000镜面显微镜。每台设备获得三次收购。对于设备间比较,选择每只眼睛/设备的最佳质量扫描,而所有三次扫描用于设备内重复性分析。重复不可靠的扫描(多达五次尝试),如果没有达到可接受的质量,则排除,导致变量分母。设备间比较使用具有参与者水平聚类和稳健标准误差的广义估计方程进行,并辅以Bland-Altman分析。结果:有效样本量因参数而异(独立总结:59-67眼;成对比较:48-51眼)。在配对眼分析中,基于scheimpflug的断层扫描仪比基于placido的地形仪测量的角膜测量值略高(K1: 43.95 vs 43.78 D, p=0.003; K2: 44.91 vs 44.73 D, p=0.002),更多的负q值(p=0.001),更小的光瞳直径(p结论:基于scheimpflug的断层扫描仪在健康眼睛中提供了可重复的角膜测量。然而,相对于基于placido的地形仪的系统差异-特别是角膜测量,非球面,瞳孔和kappa度量-表明有限的互换性。当这些参数告知临床决策时,建议一致使用器械。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.60
自引率
0.00%
发文量
200
审稿时长
6-12 weeks
期刊介绍: The ABO-ARQUIVOS BRASILEIROS DE OFTALMOLOGIA (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - electronic version), the official bimonthly publication of the Brazilian Council of Ophthalmology (CBO), aims to disseminate scientific studies in Ophthalmology, Visual Science and Health public, by promoting research, improvement and updating of professionals related to the field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书