{"title":"Reliability of a Scheimpflug-based tomographer compared with a Placido-based topographer and specular microscopy in healthy eyes.","authors":"Camyla Lemos Budib, Caio Henrique Peres Oliani, Guilherme de Oliveira Ferreira, Ulysses Tachibana, José Álvaro Pereira Gomes, Bernardo Kaplan Moscovici, Luiz Roberto Colombo-Barboza, Marcello Novoa Colombo-Barboza, Guilherme Novoa Colombo-Barboza","doi":"10.5935/0004-2749.2025-0259","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the reliability and comparability of a Scheimpflug-based tomographer relative to a Placido-based topographer and specular microscopy in healthy eyes.</p><p><strong>Methods: </strong>This cross-sectional study included 40 patients (80 eyes). Each eye underwent randomized imaging with a Scheimpflug-based tomographer, a Placido-based topographer, and Tomey EM-4000 specular microscopy. Three acquisitions per device were obtained. For interdevice comparisons, the best-quality scan per eye/device was selected, whereas all three scans were used for intradevice repeatability analyses. Unreliable scans were repeated (up to five attempts) and excluded if acceptable quality was not achieved, resulting in variable denominators. Between-device comparisons were performed using generalized estimating equations with participant-level clustering and robust standard errors and were supplemented by Bland-Altman analysis.</p><p><strong>Results: </strong>The effective sample size varied by parameter (independent summaries: 59-67 eyes; paired comparisons: 48-51 eyes). In paired-eye analyses, the Scheimpflug-based tomographer measured slightly higher keratometry values than the Placido-based topographer (K1: 43.95 vs. 43.78 D, p=0.003; K2: 44.91 vs 44.73 D, p=0.002), more negative Q-values (p=0.001), smaller photopic pupil diameter (p<0.001), and shorter kappa distance (p<0.001). Mean absolute differences were 0.32 D for K1 and 0.30 D for K2, with high dispersion for angular metrics (kappa angle coefficient of variation: 195%).</p><p><strong>Conclusions: </strong>The Scheimpflug-based tomographer provides reproducible corneal measurements in healthy eyes. However, systematic differences relative to the Placido-based topographer-particularly for keratometry, asphericity, and pupil and kappa metrics-suggest limited interchangeability. Consistent device use is recommended when these parameters inform clinical decision-making.</p>","PeriodicalId":8397,"journal":{"name":"Arquivos brasileiros de oftalmologia","volume":"89 3","pages":"e20250259"},"PeriodicalIF":1.0000,"publicationDate":"2026-05-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arquivos brasileiros de oftalmologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5935/0004-2749.2025-0259","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the reliability and comparability of a Scheimpflug-based tomographer relative to a Placido-based topographer and specular microscopy in healthy eyes.
Methods: This cross-sectional study included 40 patients (80 eyes). Each eye underwent randomized imaging with a Scheimpflug-based tomographer, a Placido-based topographer, and Tomey EM-4000 specular microscopy. Three acquisitions per device were obtained. For interdevice comparisons, the best-quality scan per eye/device was selected, whereas all three scans were used for intradevice repeatability analyses. Unreliable scans were repeated (up to five attempts) and excluded if acceptable quality was not achieved, resulting in variable denominators. Between-device comparisons were performed using generalized estimating equations with participant-level clustering and robust standard errors and were supplemented by Bland-Altman analysis.
Results: The effective sample size varied by parameter (independent summaries: 59-67 eyes; paired comparisons: 48-51 eyes). In paired-eye analyses, the Scheimpflug-based tomographer measured slightly higher keratometry values than the Placido-based topographer (K1: 43.95 vs. 43.78 D, p=0.003; K2: 44.91 vs 44.73 D, p=0.002), more negative Q-values (p=0.001), smaller photopic pupil diameter (p<0.001), and shorter kappa distance (p<0.001). Mean absolute differences were 0.32 D for K1 and 0.30 D for K2, with high dispersion for angular metrics (kappa angle coefficient of variation: 195%).
Conclusions: The Scheimpflug-based tomographer provides reproducible corneal measurements in healthy eyes. However, systematic differences relative to the Placido-based topographer-particularly for keratometry, asphericity, and pupil and kappa metrics-suggest limited interchangeability. Consistent device use is recommended when these parameters inform clinical decision-making.
期刊介绍:
The ABO-ARQUIVOS BRASILEIROS DE OFTALMOLOGIA (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - (ABO, ISSN 0004-2749 - print and ISSN 1678-2925 - electronic version), the official bimonthly publication of the Brazilian Council of Ophthalmology (CBO), aims to disseminate scientific studies in Ophthalmology, Visual Science and Health public, by promoting research, improvement and updating of professionals related to the field.