Low-dose cone-beam CT in implant dentistry: a clinical validation of diagnostic sufficiency.

IF 4 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Ruiyang Ni, Xuhui Luo, Nuo Han, Ziyang Hu, Zitong Lin
{"title":"Low-dose cone-beam CT in implant dentistry: a clinical validation of diagnostic sufficiency.","authors":"Ruiyang Ni, Xuhui Luo, Nuo Han, Ziyang Hu, Zitong Lin","doi":"10.1093/dmfr/twag035","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To validate the diagnostic sufficiency of a low-dose (LD) cone-beam CT (CBCT) protocol for dental implant assessment by comparing it to a standard-dose (SD) protocol in a clinical setting. The study evaluated linear measurement accuracy, subjective clarity of anatomical landmarks, and the impact on bone microarchitecture assessment.</p><p><strong>Methods: </strong>This retrospective study analyzed preoperative (SD) and postoperative (LD) CBCT scans from 37 patients (37 SD and 37 LD scans for both maxillary and mandibular sites). The LD protocol achieved a 76.7% dose reduction by decreasing tube current and exposure time. The diagnostic utility was assessed: (1) subjective clarity scoring of the mandibular canal and maxillary sinus floor; (2) objective linear measurements of key anatomical dimensions; and (3) quantitative analysis of bone microstructural parameters (trabecular number [Tb.N], trabecular separation [Tb.Sp], trabecular thickness [Tb.Th], bone volume fraction [BV/TV]). Reliability assessment was also performed for clarity scoring and quantitative measurements. Statistical comparisons were performed using Chi-square and paired-samples t-tests.</p><p><strong>Results: </strong>No statistically significant differences were found between the LD and SD protocols in the subjective clarity of anatomical structures (P > .05) or in the accuracy of linear measurements (P > .05). However, all 4 bone microstructural parameters showed statistically significant differences between the 2 protocols in both the maxilla and mandible (P < .05 for all). Intra- and inter-examiner assessments demonstrated substantial to good consistency across all qualitative (weighted Kappa >0.611) and quantitative (intra-class correlation coefficient [ICC] >0.702) parameters.</p><p><strong>Conclusion: </strong>This study provides clinical evidence that for the specific CBCT unit tested, the optimized LD CBCT protocol reduces patient radiation exposure by over 76.7% for dental implant assessment without compromising the diagnostic accuracy required for linear measurements and visualization of key anatomical landmarks. Therefore, for the tested model, the LD protocol is sufficient for preoperative CBCT scanning in implant dentistry. This offers a clinically validated method for enhancing patient safety in this specific context, effectively applying the As Low As Diagnostically Acceptable, Indication-oriented, and Patient-specific (ALADAIP) principle.</p>","PeriodicalId":11261,"journal":{"name":"Dento maxillo facial radiology","volume":" ","pages":"751-761"},"PeriodicalIF":4.0000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dento maxillo facial radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/dmfr/twag035","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To validate the diagnostic sufficiency of a low-dose (LD) cone-beam CT (CBCT) protocol for dental implant assessment by comparing it to a standard-dose (SD) protocol in a clinical setting. The study evaluated linear measurement accuracy, subjective clarity of anatomical landmarks, and the impact on bone microarchitecture assessment.

Methods: This retrospective study analyzed preoperative (SD) and postoperative (LD) CBCT scans from 37 patients (37 SD and 37 LD scans for both maxillary and mandibular sites). The LD protocol achieved a 76.7% dose reduction by decreasing tube current and exposure time. The diagnostic utility was assessed: (1) subjective clarity scoring of the mandibular canal and maxillary sinus floor; (2) objective linear measurements of key anatomical dimensions; and (3) quantitative analysis of bone microstructural parameters (trabecular number [Tb.N], trabecular separation [Tb.Sp], trabecular thickness [Tb.Th], bone volume fraction [BV/TV]). Reliability assessment was also performed for clarity scoring and quantitative measurements. Statistical comparisons were performed using Chi-square and paired-samples t-tests.

Results: No statistically significant differences were found between the LD and SD protocols in the subjective clarity of anatomical structures (P > .05) or in the accuracy of linear measurements (P > .05). However, all 4 bone microstructural parameters showed statistically significant differences between the 2 protocols in both the maxilla and mandible (P < .05 for all). Intra- and inter-examiner assessments demonstrated substantial to good consistency across all qualitative (weighted Kappa >0.611) and quantitative (intra-class correlation coefficient [ICC] >0.702) parameters.

Conclusion: This study provides clinical evidence that for the specific CBCT unit tested, the optimized LD CBCT protocol reduces patient radiation exposure by over 76.7% for dental implant assessment without compromising the diagnostic accuracy required for linear measurements and visualization of key anatomical landmarks. Therefore, for the tested model, the LD protocol is sufficient for preoperative CBCT scanning in implant dentistry. This offers a clinically validated method for enhancing patient safety in this specific context, effectively applying the As Low As Diagnostically Acceptable, Indication-oriented, and Patient-specific (ALADAIP) principle.

低剂量锥束计算机断层扫描在种植牙科:诊断充分性的临床验证。
目的:通过将低剂量锥形束计算机断层扫描(CBCT)方案与临床标准剂量方案进行比较,验证其诊断充分性。该研究评估了线性测量的准确性、解剖标志的主观清晰度以及对骨微结构评估的影响。材料和方法:本回顾性研究分析了37例患者的术前(标准剂量)和术后(低剂量)CBCT扫描(上颌和下颌部位的37个标准剂量和37个低剂量扫描)。低剂量方案通过减少管电流和暴露时间实现了76.7%的剂量减少。评估其诊断价值:1)下颌管及上颌窦底主观清晰度评分;2)关键解剖尺寸的客观线性测量;3)骨显微结构参数定量分析(Tb。N,结核病。Sp,结核病。Th, BV /电视)。可靠性评估也进行了清晰度评分和定量测量。统计学比较采用卡方检验和配对样本t检验。结果:低剂量与标准剂量方案在解剖结构的主观清晰度(P > 0.05)和线性测量的准确性(P > 0.05)方面无统计学差异。然而,两种方案在上颌骨和下颌骨的四种骨显微结构参数(P = 0.611)和定量参数(ICC = 0.702)上均有统计学差异。结论:本研究提供了临床证据,对于特定的CBCT测试单元,优化的低剂量CBCT方案在不影响线性测量和关键解剖标志可视化所需的诊断准确性的情况下,减少了76.7%以上的牙科种植评估患者辐射暴露。因此,对于所测试的模型,低剂量方案足以用于种植牙科的术前CBCT扫描。这为在这种特殊情况下提高患者安全性提供了一种临床验证的方法,有效地应用了ALADAIP原理。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.60
自引率
9.10%
发文量
65
审稿时长
4-8 weeks
期刊介绍: Dentomaxillofacial Radiology (DMFR) is the journal of the International Association of Dentomaxillofacial Radiology (IADMFR) and covers the closely related fields of oral radiology and head and neck imaging. Established in 1972, DMFR is a key resource keeping dentists, radiologists and clinicians and scientists with an interest in Head and Neck imaging abreast of important research and developments in oral and maxillofacial radiology. The DMFR editorial board features a panel of international experts including Editor-in-Chief Professor Ralf Schulze. Our editorial board provide their expertise and guidance in shaping the content and direction of the journal. Quick Facts: - 2015 Impact Factor - 1.919 - Receipt to first decision - average of 3 weeks - Acceptance to online publication - average of 3 weeks - Open access option - ISSN: 0250-832X - eISSN: 1476-542X
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书