{"title":"Towards quality control and harmonization of deep learning CT radiomics: An in-silico feasibility study with virtual colorectal liver metastases.","authors":"Mamatha Venugopal, Sathish Ramani, Jacob J Peoples, Richard K G Do, Amber L Simpson, Ge Wang, Bruno De Man","doi":"10.1002/mp.70500","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Radiomic imaging biomarkers are increasingly studied in oncology as a means to support disease prognosis and personalized treatment planning. While deep learning (DL) offers scalable alternatives to handcrafted radiomic features, DL-derived biomarkers are sensitive to variations in image acquisition protocols and scanner hardware-even within a single imaging modality. To ensure reliable and reproducible biomarker estimation, it is essential to (1) provide clinicians with quantitative uncertainty estimates associated with biomarker predictions, and (2) address acquisition-induced variability through harmonization strategies that render consistent performance across diverse imaging conditions.</p><p><strong>Purpose: </strong>To evaluate uncertainty estimation as a reliability metric for biomarker prediction and to assess the role of image harmonization in improving cross-scanner inference, we develop deep learning radiomics (DLR) models for joint estimation of biomarkers and associated uncertainties and apply linear harmonization filters to standardize imaging conditions.</p><p><strong>Methods: </strong>We constructed hybrid digital phantoms by embedding 20,000 virtual colorectal liver metastases into 20 clinical CT liver images to generate metastases-laden simulated scans. The proposed DLR models jointly estimated a selected set of biomarkers and their associated aleatoric uncertainties from simulated images. Models were trained, validated, and tested using a 70:20:10 data split. Variability in CT image acquisition was modelled using two scanner types, two reconstruction kernels, and three x-ray tube current settings. DLR performance was evaluated under direct inference (matched training and test scanners), cross-scanner inference, and harmonized inference using linear harmonization filters. Quantitative evaluation was based on correlation coefficients and root mean squared errors (RMSEs).</p><p><strong>Results: </strong>The estimated uncertainties were consistently higher for biomarker predictions with larger deviations from ground truth. Excluding high-uncertainty predictions improved concordance between predictions and ground truth biomarkers. The proportion of uncertain predictions increased under cross-scanner inference relative to direct inference, indicating reduced biomarker reliability under heterogeneous imaging conditions. Furthermore, application of harmonization filters reduced RMSEs by an average of 43% across biomarkers and experiments during cross-scanner inference, demonstrating improved cross-scanner consistency.</p><p><strong>Conclusions: </strong>In this controlled in-silico study, uncertainty estimation provided a practical reliability metric for DL-based radiomic biomarker prediction, while image harmonization improved reproducibility across heterogeneous acquisition conditions. These findings demonstrate methodological feasibility within a simulation-based framework but have not yet been validated on clinical CT data, motivating future clinical validation and translation.</p>","PeriodicalId":94136,"journal":{"name":"Medical physics","volume":"53 5","pages":"e70500"},"PeriodicalIF":3.2000,"publicationDate":"2026-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/mp.70500","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Radiomic imaging biomarkers are increasingly studied in oncology as a means to support disease prognosis and personalized treatment planning. While deep learning (DL) offers scalable alternatives to handcrafted radiomic features, DL-derived biomarkers are sensitive to variations in image acquisition protocols and scanner hardware-even within a single imaging modality. To ensure reliable and reproducible biomarker estimation, it is essential to (1) provide clinicians with quantitative uncertainty estimates associated with biomarker predictions, and (2) address acquisition-induced variability through harmonization strategies that render consistent performance across diverse imaging conditions.
Purpose: To evaluate uncertainty estimation as a reliability metric for biomarker prediction and to assess the role of image harmonization in improving cross-scanner inference, we develop deep learning radiomics (DLR) models for joint estimation of biomarkers and associated uncertainties and apply linear harmonization filters to standardize imaging conditions.
Methods: We constructed hybrid digital phantoms by embedding 20,000 virtual colorectal liver metastases into 20 clinical CT liver images to generate metastases-laden simulated scans. The proposed DLR models jointly estimated a selected set of biomarkers and their associated aleatoric uncertainties from simulated images. Models were trained, validated, and tested using a 70:20:10 data split. Variability in CT image acquisition was modelled using two scanner types, two reconstruction kernels, and three x-ray tube current settings. DLR performance was evaluated under direct inference (matched training and test scanners), cross-scanner inference, and harmonized inference using linear harmonization filters. Quantitative evaluation was based on correlation coefficients and root mean squared errors (RMSEs).
Results: The estimated uncertainties were consistently higher for biomarker predictions with larger deviations from ground truth. Excluding high-uncertainty predictions improved concordance between predictions and ground truth biomarkers. The proportion of uncertain predictions increased under cross-scanner inference relative to direct inference, indicating reduced biomarker reliability under heterogeneous imaging conditions. Furthermore, application of harmonization filters reduced RMSEs by an average of 43% across biomarkers and experiments during cross-scanner inference, demonstrating improved cross-scanner consistency.
Conclusions: In this controlled in-silico study, uncertainty estimation provided a practical reliability metric for DL-based radiomic biomarker prediction, while image harmonization improved reproducibility across heterogeneous acquisition conditions. These findings demonstrate methodological feasibility within a simulation-based framework but have not yet been validated on clinical CT data, motivating future clinical validation and translation.