Diagnostic performance of a novel wireless direct-conversion intraoral sensor versus complementary metal-oxide semiconductor and photostimulable storage phosphor systems for proximal caries detection: an ex vivo study.
Matheus Barros-Costa, Gustavo Machado Santaella, Gabriela Gavilan Hadid, Christiano Oliveira-Santos, Deborah Queiroz Freitas, Francisco Carlos Groppo, William C Scarfe
{"title":"Diagnostic performance of a novel wireless direct-conversion intraoral sensor versus complementary metal-oxide semiconductor and photostimulable storage phosphor systems for proximal caries detection: an ex vivo study.","authors":"Matheus Barros-Costa, Gustavo Machado Santaella, Gabriela Gavilan Hadid, Christiano Oliveira-Santos, Deborah Queiroz Freitas, Francisco Carlos Groppo, William C Scarfe","doi":"10.1093/dmfr/twag031","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To compare the diagnostic accuracy of a new wireless direct-conversion digital intraoral sensor to current indirect-conversion complementary metal-oxide semiconductor (CMOS) and photostimulable storage phosphor (PSP) sensors for detecting proximal caries.</p><p><strong>Methods: </strong>The sample in this ex vivo study were 28 human premolars with 21 micro-CT verified non-cavitated proximal dental carious lesions. Images were obtained using 3 digital intraoral systems: a wireless direct-conversion sensor, a wired indirect-conversion CMOS sensor, and a PSP sensor. A custom phantom was constructed providing a limited dental arch segment comprising a molar, first premolar, and canine. The premolar was substituted and the phantom imaged using each intraoral system with (filtered) and without (raw) the application of a manufacturer-recommended enhancements, providing 168 images. Eight oral radiologists assessed the images using a 5-point scale. Receiver operating characteristic analysis was performed to calculate sensitivity, specificity, and the area under the curve (AUC) and compared across modalities using 2-way analysis of variance (ANOVA) with post hoc Tukey's test at an a priori level of significance of 0.05.</p><p><strong>Results: </strong>The AUC ranged from 0.58 to 0.73 (sensitivity range: 0.21-0.68, specificity range: 0.69-0.92). The wireless direct-conversion sensor (0.72 ± 0.06) showed diagnostic performance comparable to CMOS (0.72 ± 0.06) and superior to PSP (0.58 ± 0.07) for raw images (P < .001). The direct-conversion sensor without image enhancement (0.68 ± 0.18) demonstrated the highest sensitivity, whereas the PSP without image enhancement (0.21 ± 0.09) had the lowest sensitivity. Application of image enhancement improved PSP performance (0.48 ± 0.16), approximating that of the other systems.</p><p><strong>Conclusions: </strong>Based on this ex vivo study, the wireless direct-conversion and CMOS sensors demonstrated similar diagnostic performance, superior to PSP in unfiltered conditions. These findings should be interpreted in light of the ex vivo design and modest sample size.</p>","PeriodicalId":11261,"journal":{"name":"Dento maxillo facial radiology","volume":" ","pages":"743-750"},"PeriodicalIF":4.0000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dento maxillo facial radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/dmfr/twag031","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: To compare the diagnostic accuracy of a new wireless direct-conversion digital intraoral sensor to current indirect-conversion complementary metal-oxide semiconductor (CMOS) and photostimulable storage phosphor (PSP) sensors for detecting proximal caries.
Methods: The sample in this ex vivo study were 28 human premolars with 21 micro-CT verified non-cavitated proximal dental carious lesions. Images were obtained using 3 digital intraoral systems: a wireless direct-conversion sensor, a wired indirect-conversion CMOS sensor, and a PSP sensor. A custom phantom was constructed providing a limited dental arch segment comprising a molar, first premolar, and canine. The premolar was substituted and the phantom imaged using each intraoral system with (filtered) and without (raw) the application of a manufacturer-recommended enhancements, providing 168 images. Eight oral radiologists assessed the images using a 5-point scale. Receiver operating characteristic analysis was performed to calculate sensitivity, specificity, and the area under the curve (AUC) and compared across modalities using 2-way analysis of variance (ANOVA) with post hoc Tukey's test at an a priori level of significance of 0.05.
Results: The AUC ranged from 0.58 to 0.73 (sensitivity range: 0.21-0.68, specificity range: 0.69-0.92). The wireless direct-conversion sensor (0.72 ± 0.06) showed diagnostic performance comparable to CMOS (0.72 ± 0.06) and superior to PSP (0.58 ± 0.07) for raw images (P < .001). The direct-conversion sensor without image enhancement (0.68 ± 0.18) demonstrated the highest sensitivity, whereas the PSP without image enhancement (0.21 ± 0.09) had the lowest sensitivity. Application of image enhancement improved PSP performance (0.48 ± 0.16), approximating that of the other systems.
Conclusions: Based on this ex vivo study, the wireless direct-conversion and CMOS sensors demonstrated similar diagnostic performance, superior to PSP in unfiltered conditions. These findings should be interpreted in light of the ex vivo design and modest sample size.
期刊介绍:
Dentomaxillofacial Radiology (DMFR) is the journal of the International Association of Dentomaxillofacial Radiology (IADMFR) and covers the closely related fields of oral radiology and head and neck imaging.
Established in 1972, DMFR is a key resource keeping dentists, radiologists and clinicians and scientists with an interest in Head and Neck imaging abreast of important research and developments in oral and maxillofacial radiology.
The DMFR editorial board features a panel of international experts including Editor-in-Chief Professor Ralf Schulze. Our editorial board provide their expertise and guidance in shaping the content and direction of the journal.
Quick Facts:
- 2015 Impact Factor - 1.919
- Receipt to first decision - average of 3 weeks
- Acceptance to online publication - average of 3 weeks
- Open access option
- ISSN: 0250-832X
- eISSN: 1476-542X