Marieke Hoogewerf, Bart P Van Putte, David Stecher, Wilhelmina Bergmann, Patrick Klein, Pieter A Doevendans
{"title":"Preclinical Feasibility of the ELANA End-to-End Anastomotic Technique in a Porcine Pilot Study.","authors":"Marieke Hoogewerf, Bart P Van Putte, David Stecher, Wilhelmina Bergmann, Patrick Klein, Pieter A Doevendans","doi":"10.2147/MDER.S520029","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To advance minimally invasive surgical techniques, we developed the sutureless ELANA<sup>®</sup> End-to-End anastomotic technique. This preclinical study aimed to evaluate the technical feasibility, hemostatic performance, and short-term patency of the new approach in a porcine model.</p><p><strong>Methods: </strong>Six pigs underwent implantation of the ELANA End-to-End device in a divided in situ right internal mammary artery (RIMA) via partial sternotomy. The pigs were followed-up for either 35 days (N=3) or 90 (N=3) days. Feasibility was evaluated based on technical success, achievement of hemostasis, and anastomotic construction time. Patency was assessed at the end of follow-up through angiography and histological analysis.</p><p><strong>Results: </strong>The anastomosis was successfully constructed in four of six animals. Among these, hemostasis was achieved in three cases without additional intervention; one required a single hemostatic stitch. The median anastomotic construction time was 18 (15-45) minutes. Technical failure in two animals was due to difficulty maintaining vessel alignment during deployment. All four successfully constructed anastomoses were patent at the end of follow-up (N=2, 35 days; N=2, 90 days), with no evidence of stenosis on angiography or histology.</p><p><strong>Conclusion: </strong>The ELANA End-to-End technique demonstrated promising patency outcomes. However, the current procedure is technically demanding. Design improvements aimed at stabilizing vessel positioning are warranted to enhance feasibility and reduce construction time.</p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":"19 ","pages":"520029"},"PeriodicalIF":1.5000,"publicationDate":"2026-05-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13159944/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Devices-Evidence and Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/MDER.S520029","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To advance minimally invasive surgical techniques, we developed the sutureless ELANA® End-to-End anastomotic technique. This preclinical study aimed to evaluate the technical feasibility, hemostatic performance, and short-term patency of the new approach in a porcine model.
Methods: Six pigs underwent implantation of the ELANA End-to-End device in a divided in situ right internal mammary artery (RIMA) via partial sternotomy. The pigs were followed-up for either 35 days (N=3) or 90 (N=3) days. Feasibility was evaluated based on technical success, achievement of hemostasis, and anastomotic construction time. Patency was assessed at the end of follow-up through angiography and histological analysis.
Results: The anastomosis was successfully constructed in four of six animals. Among these, hemostasis was achieved in three cases without additional intervention; one required a single hemostatic stitch. The median anastomotic construction time was 18 (15-45) minutes. Technical failure in two animals was due to difficulty maintaining vessel alignment during deployment. All four successfully constructed anastomoses were patent at the end of follow-up (N=2, 35 days; N=2, 90 days), with no evidence of stenosis on angiography or histology.
Conclusion: The ELANA End-to-End technique demonstrated promising patency outcomes. However, the current procedure is technically demanding. Design improvements aimed at stabilizing vessel positioning are warranted to enhance feasibility and reduce construction time.
目的:为了推进微创手术技术,我们开发了无缝线ELANA®端到端吻合技术。这项临床前研究旨在评估新方法在猪模型中的技术可行性、止血性能和短期通畅性。方法:6头猪通过部分胸骨切开术将ELANA端到端装置植入原位右乳内动脉(RIMA)。随访35 d (N=3)或90 d (N=3)。根据技术成功、止血效果和吻合口建造时间评估可行性。随访结束时通过血管造影和组织学分析评估通畅程度。结果:6只动物中4只吻合成功。其中,3例在没有额外干预的情况下实现了止血;一种只需要一针止血针。正中吻合时间18 (15 ~ 45)min。2只动物的技术故障是由于在部署期间难以保持船只对准。随访结束时(N= 2,35天;N= 2,90天),4例成功构建的吻合口均通畅,血管造影和组织学均无狭窄迹象。结论:ELANA端到端技术具有良好的通畅效果。然而,目前的程序在技术上要求很高。设计改进旨在稳定船舶定位,以提高可行性和缩短施工时间。