Determination and use of patient curvature correction factors by measurement in electron monitor unit calculations

IF 1.7 4区 医学 Q4 ONCOLOGY
Medical Dosimetry Pub Date : 2026-01-01 Epub Date: 2026-04-22 DOI:10.1016/j.meddos.2026.03.007
Neil Richmond MSc, Katie Chester MSc, Liam Walker PhD
{"title":"Determination and use of patient curvature correction factors by measurement in electron monitor unit calculations","authors":"Neil Richmond MSc,&nbsp;Katie Chester MSc,&nbsp;Liam Walker PhD","doi":"10.1016/j.meddos.2026.03.007","DOIUrl":null,"url":null,"abstract":"<div><div>In our department, monitor units (MU) for electron treatments are calculated using a Monte Carlo algorithm in the RayStation treatment planning system and independently checked using RadCalc software. An audit found that 22 of 50 independent checks failed the ± 5% local tolerance level. One of the causes for the discrepancy was thought to be that the RadCalc calculation takes no account of the patient surface curvature. Three-dimensional printed moulds were used to create wax “domes” and “bowls” with curvatures ranging from +0.33 cm<sup>−1</sup> to −0.13 cm<sup>−1</sup>. The dose per MU delivered at d<sub>max</sub><sup>,</sup> under the wax phantoms, to a calibrated NACP type chamber was determined by measurement. The fields were delivered using 6 to 18 MeV electrons from a Varian TrueBeam linac. The measured data shows that the dose varies linearly, for a given electron energy, with the curvature of the wax mould. This information was used to create a set of curvature correction factors to be applied to the 50 independent MU checks. When the appropriate correction factor was applied to the RadCalc MU only 3 of the 50 calculations were beyond the tolerance level, improving agreement and reducing the number of out of tolerance results.</div></div>","PeriodicalId":49837,"journal":{"name":"Medical Dosimetry","volume":"51 3","pages":"Pages 350-354"},"PeriodicalIF":1.7000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Dosimetry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0958394726000233","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/4/22 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

In our department, monitor units (MU) for electron treatments are calculated using a Monte Carlo algorithm in the RayStation treatment planning system and independently checked using RadCalc software. An audit found that 22 of 50 independent checks failed the ± 5% local tolerance level. One of the causes for the discrepancy was thought to be that the RadCalc calculation takes no account of the patient surface curvature. Three-dimensional printed moulds were used to create wax “domes” and “bowls” with curvatures ranging from +0.33 cm−1 to −0.13 cm−1. The dose per MU delivered at dmax, under the wax phantoms, to a calibrated NACP type chamber was determined by measurement. The fields were delivered using 6 to 18 MeV electrons from a Varian TrueBeam linac. The measured data shows that the dose varies linearly, for a given electron energy, with the curvature of the wax mould. This information was used to create a set of curvature correction factors to be applied to the 50 independent MU checks. When the appropriate correction factor was applied to the RadCalc MU only 3 of the 50 calculations were beyond the tolerance level, improving agreement and reducing the number of out of tolerance results.
电子监护装置计算中测量病人曲率修正系数的测定和使用。
在我们部门,电子治疗的监测单位(MU)是在RayStation治疗计划系统中使用蒙特卡罗算法计算的,并使用RadCalc软件进行独立检查。审核发现50个独立检查中有22个未能达到±5%的本地公差水平。造成差异的原因之一被认为是RadCalc计算没有考虑病人的表面曲率。三维打印模具被用来制作蜡制的“圆顶”和“碗”,曲率范围从+0.33 cm-1到-0.13 cm-1。在蜡模下,以dmax向校准的NACP型腔室递送的每MU剂量通过测量确定。这些电场是用瓦里安TrueBeam直线加速器中的6到18 MeV电子传输的。测量数据表明,对于给定的电子能量,剂量随蜡模的曲率呈线性变化。该信息用于创建一组曲率校正因子,应用于50个独立的MU检查。当对RadCalc MU应用适当的校正因子时,50个计算中只有3个超出公差水平,提高了一致性并减少了超出公差结果的数量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Medical Dosimetry
Medical Dosimetry 医学-核医学
CiteScore
2.40
自引率
0.00%
发文量
51
审稿时长
34 days
期刊介绍: Medical Dosimetry, the official journal of the American Association of Medical Dosimetrists, is the key source of information on new developments for the medical dosimetrist. Practical and comprehensive in coverage, the journal features original contributions and review articles by medical dosimetrists, oncologists, physicists, and radiation therapy technologists on clinical applications and techniques of external beam, interstitial, intracavitary and intraluminal irradiation in cancer management. Articles dealing primarily with physics will be reviewed by a specially appointed team of experts in the field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书