Ó. Seoane-Casqueiro , N. Vázquez-Agra , A. Cruces-Sande , E. López-Prado , J.E. López-Paz , M. Camafort , E. Casariego-Vales , Á. Hermida-Ameijeiras , A. Pose-Reino
{"title":"Fiabilidad clínica de un monitor ambulatorio de presión arterial oscilométrico de muñeca: una evaluación independiente de concordancia","authors":"Ó. Seoane-Casqueiro , N. Vázquez-Agra , A. Cruces-Sande , E. López-Prado , J.E. López-Paz , M. Camafort , E. Casariego-Vales , Á. Hermida-Ameijeiras , A. Pose-Reino","doi":"10.1016/j.rce.2026.502523","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>To evaluate the concordance between a novel wrist-worn oscillometric device and a validated ambulatory blood pressure monitor (ABPM) for blood pressure (BP) and heart rate (HR) assessment under real 24-hour ABPM conditions.</div></div><div><h3>Material and methods</h3><div>We conducted an observational concordance study in consecutive hypertensive outpatients requiring 24-hour ABPM between May and October 2025. Participants simultaneously wore a validated reference device (SunTech Oscar 2) and the wrist-worn monitor (Huawei Watch D2). Mean BP and HR values for the 24-hour, daytime, and nighttime periods were compared. Agreement was assessed using Bland–Altman analysis, intraclass correlation coefficients [ICC(A,1)], and the proportion of readings within<!--> <!-->±<!--> <!-->5,<!--> <!-->±<!--> <!-->10, and<!--> <!-->±<!--> <!-->15<!--> <!-->mmHg according to the AAMI/ESH/ISO criteria.</div></div><div><h3>Results</h3><div>Forty-one patients were included (63% women; median age: 47 years). The wrist-worn device systematically underestimated systolic BP (SBP) during the 24-hour and daytime periods and exhibited reduced variability compared with the reference device. Bland–Altman analysis showed a mean bias of –6.8<!--> <!-->mmHg for SBP with wide limits of agreement. ICC values indicated poor concordance for SBP (0.38) and moderate concordance for DBP (0.54), while HR agreement was excellent (ICC 0.87). Only 47% of systolic and 61% of DBP readings fell within<!--> <!-->±<!--> <!-->10<!--> <!-->mmHg of the reference values, below the AAMI/ESH/ISO acceptability threshold.</div></div><div><h3>Conclusions</h3><div>The wrist-worn device did not demonstrate sufficient concordance for ambulatory BP monitoring, although HR performance was reliable. These findings support the need for independent evaluations before considering such devices for clinical BP assessment.</div></div>","PeriodicalId":21223,"journal":{"name":"Revista clinica espanola","volume":"226 5","pages":"Article 502523"},"PeriodicalIF":1.7000,"publicationDate":"2026-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista clinica espanola","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S001425652601132X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/4/22 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
To evaluate the concordance between a novel wrist-worn oscillometric device and a validated ambulatory blood pressure monitor (ABPM) for blood pressure (BP) and heart rate (HR) assessment under real 24-hour ABPM conditions.
Material and methods
We conducted an observational concordance study in consecutive hypertensive outpatients requiring 24-hour ABPM between May and October 2025. Participants simultaneously wore a validated reference device (SunTech Oscar 2) and the wrist-worn monitor (Huawei Watch D2). Mean BP and HR values for the 24-hour, daytime, and nighttime periods were compared. Agreement was assessed using Bland–Altman analysis, intraclass correlation coefficients [ICC(A,1)], and the proportion of readings within ± 5, ± 10, and ± 15 mmHg according to the AAMI/ESH/ISO criteria.
Results
Forty-one patients were included (63% women; median age: 47 years). The wrist-worn device systematically underestimated systolic BP (SBP) during the 24-hour and daytime periods and exhibited reduced variability compared with the reference device. Bland–Altman analysis showed a mean bias of –6.8 mmHg for SBP with wide limits of agreement. ICC values indicated poor concordance for SBP (0.38) and moderate concordance for DBP (0.54), while HR agreement was excellent (ICC 0.87). Only 47% of systolic and 61% of DBP readings fell within ± 10 mmHg of the reference values, below the AAMI/ESH/ISO acceptability threshold.
Conclusions
The wrist-worn device did not demonstrate sufficient concordance for ambulatory BP monitoring, although HR performance was reliable. These findings support the need for independent evaluations before considering such devices for clinical BP assessment.
期刊介绍:
Revista Clínica Española published its first issue in 1940 and is the body of expression of the Spanish Society of Internal Medicine (SEMI).
The journal fully endorses the goals of updating knowledge and facilitating the acquisition of key developments in internal medicine applied to clinical practice. Revista Clínica Española is subject to a thorough double blind review of the received articles written in Spanish or English. Nine issues are published each year, including mostly originals, reviews and consensus documents.