[Development and validation of high-performance liquid chromatography methodology for quantify determination of azodicarbonamide in the workplace air].

IF 0.9
Medycyna pracy Pub Date : 2026-04-28 Epub Date: 2026-04-15 DOI:10.13075/mp.5893.01685
Joanna Kowalska, Małgorzata Pośniak
{"title":"[Development and validation of high-performance liquid chromatography methodology for quantify determination of azodicarbonamide in the workplace air].","authors":"Joanna Kowalska, Małgorzata Pośniak","doi":"10.13075/mp.5893.01685","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Occupational exposure to azodicarbonamide (ADCA) can occur via the respiratory tract or through skin in workplaces, e.g., in the production of plastics and rubber where it is used as a foaming agent or a porosity-increasing agent. Due to the irritant nature of ADCA on the respiratory tract, a concentration limit for it in workplace air has been established. This article presents a method for determining this substance in air.</p><p><strong>Material and methods: </strong>The determination method consists of the following steps: passing air containing ADCA through a glass fiber filter, extracting the substance adsorbed on the filter with acetonitrile, adding 1-naphthyl isothiocyanate and analyzing the obtained derivative using liquid chromatography. The studies were performed using an Agilent Technologies 1200 series chromatograph with a diode array detector at an analytical wavelength of 220 nm and an Ultra C18 column with a length of 25 cm and a diameter of 4.6 mm.</p><p><strong>Results: </strong>The method is linear (r = 0.9998) in the working range of 0.08-1.6 μg/ml. This range, for an air sample of 360 l, allows for the determination of 0.002-0.04 mg of ADCA per cubic meter of air. The instrument detection limit was 0.03 ng/ml, and the instrument quantification limit was 0.10 ng/ml.</p><p><strong>Conclusions: </strong>The obtained values, including precision and expanded uncertainty for measurements, meet the requirements of the European standard PN-EN 482. The usefulness of the method for estimating workers' inhalation exposure to ADCA was confirmed. Med Pr Work Health Saf. 2026;77(2):137-146.</p>","PeriodicalId":101460,"journal":{"name":"Medycyna pracy","volume":" ","pages":"137-146"},"PeriodicalIF":0.9000,"publicationDate":"2026-04-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medycyna pracy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.13075/mp.5893.01685","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/4/15 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Occupational exposure to azodicarbonamide (ADCA) can occur via the respiratory tract or through skin in workplaces, e.g., in the production of plastics and rubber where it is used as a foaming agent or a porosity-increasing agent. Due to the irritant nature of ADCA on the respiratory tract, a concentration limit for it in workplace air has been established. This article presents a method for determining this substance in air.

Material and methods: The determination method consists of the following steps: passing air containing ADCA through a glass fiber filter, extracting the substance adsorbed on the filter with acetonitrile, adding 1-naphthyl isothiocyanate and analyzing the obtained derivative using liquid chromatography. The studies were performed using an Agilent Technologies 1200 series chromatograph with a diode array detector at an analytical wavelength of 220 nm and an Ultra C18 column with a length of 25 cm and a diameter of 4.6 mm.

Results: The method is linear (r = 0.9998) in the working range of 0.08-1.6 μg/ml. This range, for an air sample of 360 l, allows for the determination of 0.002-0.04 mg of ADCA per cubic meter of air. The instrument detection limit was 0.03 ng/ml, and the instrument quantification limit was 0.10 ng/ml.

Conclusions: The obtained values, including precision and expanded uncertainty for measurements, meet the requirements of the European standard PN-EN 482. The usefulness of the method for estimating workers' inhalation exposure to ADCA was confirmed. Med Pr Work Health Saf. 2026;77(2):137-146.

[高效液相色谱法定量测定工作场所空气中偶氮二甲酰胺的开发和验证]。
背景:偶氮二甲酰胺(ADCA)的职业暴露可通过呼吸道或工作场所的皮肤发生,例如,在塑料和橡胶的生产中,偶氮二甲酰胺被用作发泡剂或增气孔剂。由于ADCA对呼吸道的刺激性,工作场所空气中ADCA的浓度限值已被确定。本文介绍了一种测定空气中这种物质的方法。材料和方法:测定方法包括:将含有ADCA的空气通过玻璃纤维过滤器,用乙腈提取吸附在过滤器上的物质,加入1-异硫氰酸萘酯,用液相色谱法分析得到的衍生物。采用Agilent Technologies 1200系列色谱仪,二极管阵列检测器,分析波长为220 nm, Ultra C18色谱柱长为25 cm,直径为4.6 mm。结果:在0.08 ~ 1.6 μg/ml范围内,该方法线性(r = 0.9998)良好。这个范围,对于360升的空气样品,允许测定0.002-0.04毫克ADCA每立方米空气。仪器检测限为0.03 ng/ml,仪器定量限为0.10 ng/ml。结论:所得测量值的精密度和扩展不确定度均符合欧洲标准PN-EN 482的要求。证实了该方法对估计工人吸入ADCA暴露量的有效性。中华医学杂志,2009;27(2)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书