[Simulation Analysis and Radiation Protection Study of Induced Radioactivity in Carbon Ion Therapy Systems].

Q4 Medicine
Siyang Liu, Chunying Jiao, Shibing Xie, Guotao Fu, Peichen Wang, Yulin Deng
{"title":"[Simulation Analysis and Radiation Protection Study of Induced Radioactivity in Carbon Ion Therapy Systems].","authors":"Siyang Liu, Chunying Jiao, Shibing Xie, Guotao Fu, Peichen Wang, Yulin Deng","doi":"10.12455/j.issn.1671-7104.250269","DOIUrl":null,"url":null,"abstract":"<p><p><b>Objective</b> Addressing the issue of induced radioactivity in carbon ion therapy system operation, to systematically evaluate the impact of beam parameters on the distribution of induced radiation field in carbon ion therapy room and develop radiation protection strategies. <b>Methods</b> A geometric model integrating a multi-leaf collimator (MLC), compensator, patient phantom, and concrete shielding structure was developed using the FLUKA Monte Carlo simulation code. Systematic simulations were conducted to analyze residual dose rate decay kinetics and radionuclide activation profiles at critical components under 400 MeV/u carbon ion irradiation with beam intensities and irradiation durations. <b>Results</b> The induced radioactivity generated by components exhibited a linear response to beam intensity. The MLC surface exhibited a residual dose rate of 15.8 μSv/h at 1 min post-irradiation (1×10 <sup>8</sup> pps, 15-minute irradiation), requiring 45 min cooling for the safety threshold. The activation nuclides produced in patient phantoms were dominated by short-lived nuclides, with radiation dose decreased to acceptable level within 10 min post-irradiation. Occupational exposure assessments demonstrated annual effective doses of about 174 μSv (air immersion) and 0.31 μSv (inhalation). <b>Conclusion</b> The induced radioactivity dose produced by the carbon ion radiotherapy system is mainly generated by MLC and patient phantoms. Personnel should avoid direct contact within 10 minutes post-irradiation. Occupational exposure levels of the air activation remain compliant with occupational dose constraints.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":"50 2","pages":"160-166"},"PeriodicalIF":0.0000,"publicationDate":"2026-03-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.250269","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective Addressing the issue of induced radioactivity in carbon ion therapy system operation, to systematically evaluate the impact of beam parameters on the distribution of induced radiation field in carbon ion therapy room and develop radiation protection strategies. Methods A geometric model integrating a multi-leaf collimator (MLC), compensator, patient phantom, and concrete shielding structure was developed using the FLUKA Monte Carlo simulation code. Systematic simulations were conducted to analyze residual dose rate decay kinetics and radionuclide activation profiles at critical components under 400 MeV/u carbon ion irradiation with beam intensities and irradiation durations. Results The induced radioactivity generated by components exhibited a linear response to beam intensity. The MLC surface exhibited a residual dose rate of 15.8 μSv/h at 1 min post-irradiation (1×10 8 pps, 15-minute irradiation), requiring 45 min cooling for the safety threshold. The activation nuclides produced in patient phantoms were dominated by short-lived nuclides, with radiation dose decreased to acceptable level within 10 min post-irradiation. Occupational exposure assessments demonstrated annual effective doses of about 174 μSv (air immersion) and 0.31 μSv (inhalation). Conclusion The induced radioactivity dose produced by the carbon ion radiotherapy system is mainly generated by MLC and patient phantoms. Personnel should avoid direct contact within 10 minutes post-irradiation. Occupational exposure levels of the air activation remain compliant with occupational dose constraints.

碳离子治疗系统诱导放射性的模拟分析与辐射防护研究
目的针对碳离子治疗系统运行中的诱导放射性问题,系统评价束流参数对碳离子治疗室诱导辐射场分布的影响,制定辐射防护策略。方法利用FLUKA蒙特卡罗仿真程序建立多叶准直器(MLC)、补偿器、患者幻影和混凝土屏蔽结构的几何模型。系统模拟分析了400 MeV/u碳离子辐照下关键组分的残余剂量率衰减动力学和放射性核素活化谱。结果各组分的诱导放射性随束流强度呈线性变化。辐照后1 min (1×10 8 pps,辐照15 min), MLC表面的残留剂量率为15.8 μSv/h,需要冷却45 min才能达到安全阈值。患者幻影中产生的活化核素以短寿命核素为主,照射后10分钟内辐射剂量降至可接受水平。职业暴露评估显示年有效剂量约为174 μSv(空气浸泡)和0.31 μSv(吸入)。结论碳离子放射治疗系统产生的诱导放射性剂量主要由MLC和患者幻影产生。照射后10分钟内人员应避免直接接触。空气活化的职业暴露水平仍然符合职业剂量限制。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍: Chinese Journal of Medical Instrumentation mainly reports on the development, progress, research and development, production, clinical application, management, and maintenance of medical devices and biomedical engineering. Its aim is to promote the exchange of information on medical devices and biomedical engineering in China and turn the journal into a high-quality academic journal that leads academic directions and advocates academic debates.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书