Design and feasibility testing of an automated custom finger orthosis for virtual therapy.

IF 1.8 4区 医学 Q2 ORTHOPEDICS
Drew Miles, Yonas Tadesse
{"title":"Design and feasibility testing of an automated custom finger orthosis for virtual therapy.","authors":"Drew Miles, Yonas Tadesse","doi":"10.1016/j.jht.2026.01.004","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>3D printing has revolutionized orthoses manufacturing in assistive technologies, enabling development of custom-fitted orthoses in under 24 hours. Physician-engineer collaboration is essential to addressing the mechanical complexities of automated orthoses design. Single-step manufacturing of these devices is necessary for simplistic and translational technologies.</p><p><strong>Purpose: </strong>This work proposes a novel automated program, application programing interface, to design a custom single-finger orthosis using only basic biometric data (body weight, height, and age) as input to enable fully virtual rehabilitation.</p><p><strong>Study design: </strong>Feasibility study, cross-sectional design, and survey study.</p><p><strong>Methods: </strong>Customization without body scans streamlines programming and enables gross estimate of hand orthosis size. Computer-aided design models are automatically generated using the Automated Custom Finger Orthoses system that was created using several macros with SolidWorks and Visual Basic Editor. A military database was used for the biometric data for the program to establish a relationship between finger size and biometric data. Survey data were taken from 10 human healthy participants (six male and four female, aged between 18 and 60) using Likert scales to check fitness of 3D-printed soft index finger orthosis part.</p><p><strong>Results: </strong>Computer-aided design models were successfully created using the application programing interface using military database, the models were 3D-printed and fitted on subjects. The fitness of the device varies significantly under the current parameters. Eight of the 10 test devices fit their test subjects with overall average offsets of 7.00 mm, 6.14 mm, and 8.89 mm for DIP (L1), PIP (L2), and MCP joints (L3 offset), respectively and the subjects were fully donned. The biasing of current data toward male subjects results in slightly better fit accuracy in males than in females.</p><p><strong>Conclusions: </strong>Anthropometric studies can aid in discerning fundamental ratios for orthoses design, allowing for custom-fitted designs and manufacturing with ease and efficiency.</p>","PeriodicalId":54814,"journal":{"name":"Journal of Hand Therapy","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2026-04-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Hand Therapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jht.2026.01.004","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: 3D printing has revolutionized orthoses manufacturing in assistive technologies, enabling development of custom-fitted orthoses in under 24 hours. Physician-engineer collaboration is essential to addressing the mechanical complexities of automated orthoses design. Single-step manufacturing of these devices is necessary for simplistic and translational technologies.

Purpose: This work proposes a novel automated program, application programing interface, to design a custom single-finger orthosis using only basic biometric data (body weight, height, and age) as input to enable fully virtual rehabilitation.

Study design: Feasibility study, cross-sectional design, and survey study.

Methods: Customization without body scans streamlines programming and enables gross estimate of hand orthosis size. Computer-aided design models are automatically generated using the Automated Custom Finger Orthoses system that was created using several macros with SolidWorks and Visual Basic Editor. A military database was used for the biometric data for the program to establish a relationship between finger size and biometric data. Survey data were taken from 10 human healthy participants (six male and four female, aged between 18 and 60) using Likert scales to check fitness of 3D-printed soft index finger orthosis part.

Results: Computer-aided design models were successfully created using the application programing interface using military database, the models were 3D-printed and fitted on subjects. The fitness of the device varies significantly under the current parameters. Eight of the 10 test devices fit their test subjects with overall average offsets of 7.00 mm, 6.14 mm, and 8.89 mm for DIP (L1), PIP (L2), and MCP joints (L3 offset), respectively and the subjects were fully donned. The biasing of current data toward male subjects results in slightly better fit accuracy in males than in females.

Conclusions: Anthropometric studies can aid in discerning fundamental ratios for orthoses design, allowing for custom-fitted designs and manufacturing with ease and efficiency.

一种用于虚拟治疗的自动定制手指矫形器的设计和可行性测试。
背景:3D打印彻底改变了辅助技术中的矫形器制造,使定制矫形器的开发在24小时内成为可能。医师与工程师的合作对于解决自动化矫形器设计的机械复杂性至关重要。这些设备的单步制造对于简化和转化技术是必要的。目的:本工作提出了一种新颖的自动化程序,应用程序编程接口,仅使用基本的生物特征数据(体重,身高和年龄)作为输入来设计定制的单指矫形器,以实现完全虚拟康复。研究设计:可行性研究、横断面设计、调查研究。方法:无需身体扫描的定制简化了编程,并使手部矫形器尺寸的总体估计成为可能。使用SolidWorks和Visual Basic编辑器使用几个宏创建的自动定制手指矫形器系统自动生成计算机辅助设计模型。该程序的生物特征数据使用了一个军事数据库,以建立手指尺寸和生物特征数据之间的关系。采用Likert量表对10名健康受试者(男性6名,女性4名,年龄在18 - 60岁之间)进行调查数据,检测3d打印软性食指矫形器部件的适合度。结果:利用军用数据库的应用程序编程界面,成功地建立了计算机辅助设计模型,并对模型进行了3d打印和拟合。在当前参数下,器件的适应度变化很大。10个测试装置中的8个分别适合DIP (L1)、PIP (L2)和MCP关节(L3偏移)的总体平均偏移量为7.00 mm、6.14 mm和8.89 mm,并且受试者完全戴上。当前数据偏向男性受试者导致男性的拟合精度略高于女性。结论:人体测量学研究可以帮助识别矫形器设计的基本比例,允许定制设计和制造轻松高效。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Hand Therapy
Journal of Hand Therapy 医学-外科
CiteScore
3.50
自引率
10.00%
发文量
65
审稿时长
19.2 weeks
期刊介绍: The Journal of Hand Therapy is designed for hand therapists, occupational and physical therapists, and other hand specialists involved in the rehabilitation of disabling hand problems. The Journal functions as a source of education and information by publishing scientific and clinical articles. Regular features include original reports, clinical reviews, case studies, editorials, and book reviews.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书