Gayani Nilmini Dissanayake, Hanine Fourie, Gordana Fricker, Darmiga Thayabaran, Mayur Patel, Ingrid Granne, Brian Shine
{"title":"False-positive point-of-care urine beta human chorionic gonadotropin testing: Insights from two clinical cases.","authors":"Gayani Nilmini Dissanayake, Hanine Fourie, Gordana Fricker, Darmiga Thayabaran, Mayur Patel, Ingrid Granne, Brian Shine","doi":"10.1177/00045632261433346","DOIUrl":null,"url":null,"abstract":"<p><p>Case oneA young woman presented to the Early Pregnancy Assessment Unit (EPAU) with abdominal pain, amenorrhoea and variably positive home-pregnancy tests. On review, a point-of-care test (POCT) for urine beta-HCG (β-HCG) was negative but a blood test using the Abbott i-STAT β-HCG POCT device was positive. Initial transvaginal ultrasound did not demonstrate an intra- or extra-uterine pregnancy. Over the following 2 months, weekly plasma POCT iSTAT β-HCG checks remained positive. After a further ultrasound suggesting a possible ectopic pregnancy, the patient underwent a diagnostic laparoscopy which was unremarkable. Post-operatively, POCT iSTAT β-HCG levels remained elevated, and a blood sample was sent for laboratory analysis revealing an undetectable β-HCG of <1.20 IU/L (reference interval 0-4).Case twoA middle-aged woman presented to the emergency department with mons pubis pain and swelling and was admitted to the gynaecology ward for drainage of bilateral abscesses. On review, she had raised blood β-HCG levels, measured using the Abbott i-STAT POCT device. A subsequent blood sample sent for laboratory analysis showed β-HCG levels within the post-menopausal reference interval. <b>Discussion:</b> In both cases, POCT immunoassay interference was confirmed by consistent results produced when contemporaneous samples were analysed by different analytical platforms. Immunoassay interference, though rare, can lead to inaccurate results from POCT devices, potentially impacting patient diagnosis and management. Clinical teams should remain vigilant for this possibility; if test results do not align with clinical expectations, it is essential to promptly send a blood sample for laboratory analysis.</p>","PeriodicalId":8005,"journal":{"name":"Annals of Clinical Biochemistry","volume":" ","pages":"458-461"},"PeriodicalIF":1.1000,"publicationDate":"2026-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Clinical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/00045632261433346","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/21 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Case oneA young woman presented to the Early Pregnancy Assessment Unit (EPAU) with abdominal pain, amenorrhoea and variably positive home-pregnancy tests. On review, a point-of-care test (POCT) for urine beta-HCG (β-HCG) was negative but a blood test using the Abbott i-STAT β-HCG POCT device was positive. Initial transvaginal ultrasound did not demonstrate an intra- or extra-uterine pregnancy. Over the following 2 months, weekly plasma POCT iSTAT β-HCG checks remained positive. After a further ultrasound suggesting a possible ectopic pregnancy, the patient underwent a diagnostic laparoscopy which was unremarkable. Post-operatively, POCT iSTAT β-HCG levels remained elevated, and a blood sample was sent for laboratory analysis revealing an undetectable β-HCG of <1.20 IU/L (reference interval 0-4).Case twoA middle-aged woman presented to the emergency department with mons pubis pain and swelling and was admitted to the gynaecology ward for drainage of bilateral abscesses. On review, she had raised blood β-HCG levels, measured using the Abbott i-STAT POCT device. A subsequent blood sample sent for laboratory analysis showed β-HCG levels within the post-menopausal reference interval. Discussion: In both cases, POCT immunoassay interference was confirmed by consistent results produced when contemporaneous samples were analysed by different analytical platforms. Immunoassay interference, though rare, can lead to inaccurate results from POCT devices, potentially impacting patient diagnosis and management. Clinical teams should remain vigilant for this possibility; if test results do not align with clinical expectations, it is essential to promptly send a blood sample for laboratory analysis.
期刊介绍:
Annals of Clinical Biochemistry is the fully peer reviewed international journal of the Association for Clinical Biochemistry and Laboratory Medicine.
Annals of Clinical Biochemistry accepts papers that contribute to knowledge in all fields of laboratory medicine, especially those pertaining to the understanding, diagnosis and treatment of human disease. It publishes papers on clinical biochemistry, clinical audit, metabolic medicine, immunology, genetics, biotechnology, haematology, microbiology, computing and management where they have both biochemical and clinical relevance. Papers describing evaluation or implementation of commercial reagent kits or the performance of new analysers require substantial original information. Unless of exceptional interest and novelty, studies dealing with the redox status in various diseases are not generally considered within the journal''s scope. Studies documenting the association of single nucleotide polymorphisms (SNPs) with particular phenotypes will not normally be considered, given the greater strength of genome wide association studies (GWAS). Research undertaken in non-human animals will not be considered for publication in the Annals.
Annals of Clinical Biochemistry is also the official journal of NVKC (de Nederlandse Vereniging voor Klinische Chemie) and JSCC (Japan Society of Clinical Chemistry).