Ilja S Deurloo, Milou M Steemers-de Boer, Johannes H van Snick, Klaas Pieter Koopmans, Riemer H J A Slart, Walter Noordzij, Adrienne H Brouwers, Gilles N Stormezand, Ton Velleman, Erik H J G Aarntzen, Berlinda J de Wit-van der Veen, Wouter V Vogel, Sanne M Jansen, Arthur J A T Braat, Andor W J M Glaudemans, Oleksandra V Ivashchenko
{"title":"Smart scanning for two: [<sup>18</sup>F]FDG dose optimization for pregnant patients using LAFOV PET/CT technology.","authors":"Ilja S Deurloo, Milou M Steemers-de Boer, Johannes H van Snick, Klaas Pieter Koopmans, Riemer H J A Slart, Walter Noordzij, Adrienne H Brouwers, Gilles N Stormezand, Ton Velleman, Erik H J G Aarntzen, Berlinda J de Wit-van der Veen, Wouter V Vogel, Sanne M Jansen, Arthur J A T Braat, Andor W J M Glaudemans, Oleksandra V Ivashchenko","doi":"10.1186/s40658-026-00855-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to evaluate the image quality-dose trade-off in pregnant patients imaged with long-axial field-of-view [<sup>18</sup>F]FDG PET/CT and to identify the most predictive body composition metric for image quality to develop a pregnancy-tailored dosage model.</p><p><strong>Methods: </strong>Patients imaged with [<sup>18</sup>F]FDG PET/CT according to local pregnancy protocols were included in this study. Using raw PET data, images of various degrees of image quality were reconstructed. Acceptable image quality was identified using signal-to-noise ratio (SNR) in the liver and Likert scores. The minimum required scan statistics was modelled based on SNR and patient body composition. F-tests were used to find the best-fitting model parameter out of weight, weight-to-height-ratio, body-mass-index, and body surface area (BSA). Foetal dose was estimated with PET conversion factors and size-specific CT dose index values.</p><p><strong>Results: </strong>Eleven patients were included in image quality analysis and dosage model optimization. SNR strongly correlated with Likert scores (R² = 0.80), with 10.72 SNR indicating acceptable image quality. BSA best predicted image quality (R² = 0.85), outperforming weight (R² = 0.78), weight-to-height ratio (R² = 0.63), and body mass index (R² = 0.38). The proposed dosage model reduces activity by 41-96% compared to current local pregnancy and adult protocols, resulting in estimated foetal radiation doses of 0.066 mGy (PET) and 0.31 mGy (CT).</p><p><strong>Conclusion: </strong>BSA accurately predicted [<sup>18</sup>F]FDG PET/CT image quality in pregnant patients. The proposed dosage regimen allows significant dose reduction and can be used as a foundation for the development of pregnancy dosage protocols.</p>","PeriodicalId":11559,"journal":{"name":"EJNMMI Physics","volume":" ","pages":""},"PeriodicalIF":3.5000,"publicationDate":"2026-03-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13106102/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40658-026-00855-7","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study aimed to evaluate the image quality-dose trade-off in pregnant patients imaged with long-axial field-of-view [18F]FDG PET/CT and to identify the most predictive body composition metric for image quality to develop a pregnancy-tailored dosage model.
Methods: Patients imaged with [18F]FDG PET/CT according to local pregnancy protocols were included in this study. Using raw PET data, images of various degrees of image quality were reconstructed. Acceptable image quality was identified using signal-to-noise ratio (SNR) in the liver and Likert scores. The minimum required scan statistics was modelled based on SNR and patient body composition. F-tests were used to find the best-fitting model parameter out of weight, weight-to-height-ratio, body-mass-index, and body surface area (BSA). Foetal dose was estimated with PET conversion factors and size-specific CT dose index values.
Results: Eleven patients were included in image quality analysis and dosage model optimization. SNR strongly correlated with Likert scores (R² = 0.80), with 10.72 SNR indicating acceptable image quality. BSA best predicted image quality (R² = 0.85), outperforming weight (R² = 0.78), weight-to-height ratio (R² = 0.63), and body mass index (R² = 0.38). The proposed dosage model reduces activity by 41-96% compared to current local pregnancy and adult protocols, resulting in estimated foetal radiation doses of 0.066 mGy (PET) and 0.31 mGy (CT).
Conclusion: BSA accurately predicted [18F]FDG PET/CT image quality in pregnant patients. The proposed dosage regimen allows significant dose reduction and can be used as a foundation for the development of pregnancy dosage protocols.
期刊介绍:
EJNMMI Physics is an international platform for scientists, users and adopters of nuclear medicine with a particular interest in physics matters. As a companion journal to the European Journal of Nuclear Medicine and Molecular Imaging, this journal has a multi-disciplinary approach and welcomes original materials and studies with a focus on applied physics and mathematics as well as imaging systems engineering and prototyping in nuclear medicine. This includes physics-driven approaches or algorithms supported by physics that foster early clinical adoption of nuclear medicine imaging and therapy.