Red reflex examination for early detection of pediatric cataract using a novel red-reflex screening device.

IF 2.7
Expert review of medical devices Pub Date : 2026-06-01 Epub Date: 2026-03-19 DOI:10.1080/17434440.2026.2646442
Hongzhe Li, Zexin Chen, Li Wan, Zhangliang Li, Yun-E Zhao
{"title":"Red reflex examination for early detection of pediatric cataract using a novel red-reflex screening device.","authors":"Hongzhe Li, Zexin Chen, Li Wan, Zhangliang Li, Yun-E Zhao","doi":"10.1080/17434440.2026.2646442","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Early detection of pediatric cataracts is essential to prevent irreversible visual impairment; however, effective screening remains limited in many settings. This study evaluated the diagnostic performance of the CatScreener, a self-designed handheld red-reflex screening device, when operated by non-expert examiners.</p><p><strong>Methods: </strong>A total of 250 eyes from 250 infants and young children (≤8 years) underwent CatScreener screening and slit-lamp examination. Images were independently interpreted by two medical students and two junior ophthalmologists. Diagnostic outcomes were compared with an expert-defined reference standard. Cumulative summation (CUSUM) analysis was used to assess the learning curve for operational proficiency.</p><p><strong>Results: </strong>Diagnostic accuracy did not differ significantly between CatScreener screening and slit-lamp examination across all examiners (all <i>p</i> > 0.05). The CatScreener demonstrated high sensitivity (84.9%-87.1%) and specificity (89.2%-91.0%), with excellent agreement with slit-lamp examination (Kappa = 0.856-0.952) and good inter-examiner consistency (Kappa = 0.792-0.920). Sensitivity and specificity remained stable across chronological operation stages and age-stratified subgroups. CUSUM analysis indicated that approximately 50 cases were required to achieve operational proficiency.</p><p><strong>Conclusions: </strong>The CatScreener is an effective and user-friendly tool for the early detection of pediatric cataracts, demonstrating high diagnostic accuracy, strong agreement with conventional examination methods, and consistent performance across examiners with varying levels of clinical experience.</p>","PeriodicalId":94006,"journal":{"name":"Expert review of medical devices","volume":" ","pages":"713-721"},"PeriodicalIF":2.7000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Expert review of medical devices","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/17434440.2026.2646442","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/19 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Early detection of pediatric cataracts is essential to prevent irreversible visual impairment; however, effective screening remains limited in many settings. This study evaluated the diagnostic performance of the CatScreener, a self-designed handheld red-reflex screening device, when operated by non-expert examiners.

Methods: A total of 250 eyes from 250 infants and young children (≤8 years) underwent CatScreener screening and slit-lamp examination. Images were independently interpreted by two medical students and two junior ophthalmologists. Diagnostic outcomes were compared with an expert-defined reference standard. Cumulative summation (CUSUM) analysis was used to assess the learning curve for operational proficiency.

Results: Diagnostic accuracy did not differ significantly between CatScreener screening and slit-lamp examination across all examiners (all p > 0.05). The CatScreener demonstrated high sensitivity (84.9%-87.1%) and specificity (89.2%-91.0%), with excellent agreement with slit-lamp examination (Kappa = 0.856-0.952) and good inter-examiner consistency (Kappa = 0.792-0.920). Sensitivity and specificity remained stable across chronological operation stages and age-stratified subgroups. CUSUM analysis indicated that approximately 50 cases were required to achieve operational proficiency.

Conclusions: The CatScreener is an effective and user-friendly tool for the early detection of pediatric cataracts, demonstrating high diagnostic accuracy, strong agreement with conventional examination methods, and consistent performance across examiners with varying levels of clinical experience.

红色反射检查在儿童白内障早期发现中的应用。
背景:儿童白内障的早期发现对于预防不可逆的视力损害至关重要;然而,在许多情况下,有效的筛查仍然有限。本研究评估了CatScreener(一种自行设计的手持式红色反射筛查设备)在由非专业检查人员操作时的诊断性能。方法:对250例婴幼儿(≤8岁)250只眼进行CatScreener筛查和裂隙灯检查。图像由两名医学生和两名初级眼科医生独立解读。将诊断结果与专家定义的参考标准进行比较。累积求和(CUSUM)分析用于评估操作熟练程度的学习曲线。结果:在所有检查者中,CatScreener筛查和裂隙灯检查的诊断准确率无显著差异(p < 0.05)。CatScreener具有较高的灵敏度(84.9% ~ 87.1%)和特异度(89.2% ~ 91.0%),与裂隙灯检查(Kappa = 0.856 ~ 0.952)具有良好的一致性(Kappa = 0.792 ~ 0.920)。敏感性和特异性在手术时间分期和年龄分层亚组中保持稳定。CUSUM分析表明,大约需要50个病例才能达到操作熟练程度。结论:CatScreener是一种有效且用户友好的儿童白内障早期检测工具,具有较高的诊断准确性,与传统检查方法高度一致,并且具有不同临床经验水平的检查人员的表现一致。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书