Hongzhe Li, Zexin Chen, Li Wan, Zhangliang Li, Yun-E Zhao
{"title":"Red reflex examination for early detection of pediatric cataract using a novel red-reflex screening device.","authors":"Hongzhe Li, Zexin Chen, Li Wan, Zhangliang Li, Yun-E Zhao","doi":"10.1080/17434440.2026.2646442","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Early detection of pediatric cataracts is essential to prevent irreversible visual impairment; however, effective screening remains limited in many settings. This study evaluated the diagnostic performance of the CatScreener, a self-designed handheld red-reflex screening device, when operated by non-expert examiners.</p><p><strong>Methods: </strong>A total of 250 eyes from 250 infants and young children (≤8 years) underwent CatScreener screening and slit-lamp examination. Images were independently interpreted by two medical students and two junior ophthalmologists. Diagnostic outcomes were compared with an expert-defined reference standard. Cumulative summation (CUSUM) analysis was used to assess the learning curve for operational proficiency.</p><p><strong>Results: </strong>Diagnostic accuracy did not differ significantly between CatScreener screening and slit-lamp examination across all examiners (all <i>p</i> > 0.05). The CatScreener demonstrated high sensitivity (84.9%-87.1%) and specificity (89.2%-91.0%), with excellent agreement with slit-lamp examination (Kappa = 0.856-0.952) and good inter-examiner consistency (Kappa = 0.792-0.920). Sensitivity and specificity remained stable across chronological operation stages and age-stratified subgroups. CUSUM analysis indicated that approximately 50 cases were required to achieve operational proficiency.</p><p><strong>Conclusions: </strong>The CatScreener is an effective and user-friendly tool for the early detection of pediatric cataracts, demonstrating high diagnostic accuracy, strong agreement with conventional examination methods, and consistent performance across examiners with varying levels of clinical experience.</p>","PeriodicalId":94006,"journal":{"name":"Expert review of medical devices","volume":" ","pages":"713-721"},"PeriodicalIF":2.7000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Expert review of medical devices","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/17434440.2026.2646442","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/19 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Early detection of pediatric cataracts is essential to prevent irreversible visual impairment; however, effective screening remains limited in many settings. This study evaluated the diagnostic performance of the CatScreener, a self-designed handheld red-reflex screening device, when operated by non-expert examiners.
Methods: A total of 250 eyes from 250 infants and young children (≤8 years) underwent CatScreener screening and slit-lamp examination. Images were independently interpreted by two medical students and two junior ophthalmologists. Diagnostic outcomes were compared with an expert-defined reference standard. Cumulative summation (CUSUM) analysis was used to assess the learning curve for operational proficiency.
Results: Diagnostic accuracy did not differ significantly between CatScreener screening and slit-lamp examination across all examiners (all p > 0.05). The CatScreener demonstrated high sensitivity (84.9%-87.1%) and specificity (89.2%-91.0%), with excellent agreement with slit-lamp examination (Kappa = 0.856-0.952) and good inter-examiner consistency (Kappa = 0.792-0.920). Sensitivity and specificity remained stable across chronological operation stages and age-stratified subgroups. CUSUM analysis indicated that approximately 50 cases were required to achieve operational proficiency.
Conclusions: The CatScreener is an effective and user-friendly tool for the early detection of pediatric cataracts, demonstrating high diagnostic accuracy, strong agreement with conventional examination methods, and consistent performance across examiners with varying levels of clinical experience.