[Dose Management and Estimation Model Development Using a Dose Area Product Meter during kV Planar Imaging in Image-guided Radiotherapy].

Hideharu Miura, Toshiya Okazue, Kenji Kanemoto
{"title":"[Dose Management and Estimation Model Development Using a Dose Area Product Meter during kV Planar Imaging in Image-guided Radiotherapy].","authors":"Hideharu Miura, Toshiya Okazue, Kenji Kanemoto","doi":"10.6009/jjrt.26-1614","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to measure patient exposure dose during kilovoltage (kV) planar imaging in image-guided radiotherapy (IGRT) using a dose area product (DAP) meter and to estimate the DAP under unmeasured imaging conditions using two-dimensional cubic interpolation.</p><p><strong>Methods: </strong>Dose measurements during kV planar imaging were performed using a DAP meter (VacuDAP duo; VacuTec Meßtechnik, Dresden, Germany) attached to the kV X-ray tube of the On-Board Imager mounted on a TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA). DAP was measured under various imaging conditions, with tube voltages ranging from 60 to 140 kV and tube currents from 2.5 to 20 mAs. A two-dimensional cubic interpolation based on the measured data was implemented to estimate the DAP for unmeasured imaging conditions.</p><p><strong>Results: </strong>The deviation from linearity was relatively larger under low-dose conditions, with a maximum deviation of 1.8% at 100 kV and 2.5 mAs. Nevertheless, the overall DAP for each mAs value exhibited high linearity (R<sup>2</sup>=1.00). The relative error of the estimated DAP was within 2.0%.</p><p><strong>Conclusion: </strong>The DAP meter was demonstrated to be effective for dose management in kV planar imaging in IGRT, and use of the two-dimensional cubic interpolation enabled accurate dose estimation even under unmeasured imaging conditions.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":"82 4","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.26-1614","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study aimed to measure patient exposure dose during kilovoltage (kV) planar imaging in image-guided radiotherapy (IGRT) using a dose area product (DAP) meter and to estimate the DAP under unmeasured imaging conditions using two-dimensional cubic interpolation.

Methods: Dose measurements during kV planar imaging were performed using a DAP meter (VacuDAP duo; VacuTec Meßtechnik, Dresden, Germany) attached to the kV X-ray tube of the On-Board Imager mounted on a TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA). DAP was measured under various imaging conditions, with tube voltages ranging from 60 to 140 kV and tube currents from 2.5 to 20 mAs. A two-dimensional cubic interpolation based on the measured data was implemented to estimate the DAP for unmeasured imaging conditions.

Results: The deviation from linearity was relatively larger under low-dose conditions, with a maximum deviation of 1.8% at 100 kV and 2.5 mAs. Nevertheless, the overall DAP for each mAs value exhibited high linearity (R2=1.00). The relative error of the estimated DAP was within 2.0%.

Conclusion: The DAP meter was demonstrated to be effective for dose management in kV planar imaging in IGRT, and use of the two-dimensional cubic interpolation enabled accurate dose estimation even under unmeasured imaging conditions.

[使用剂量面积积计在图像引导放疗中kV平面成像中的剂量管理和估计模型的建立]。
目的:本研究旨在使用剂量面积积(DAP)计测量图像引导放疗(IGRT)中千伏(kV)平面成像时患者的暴露剂量,并使用二维三次插值估计未测量成像条件下的DAP。方法:使用DAP仪表(VacuDAP duo; VacuTec Meßtechnik, Dresden, Germany)连接到安装在TrueBeam直线加速器(Varian Medical Systems, Palo Alto, CA, USA)上的机载成像仪的kV x射线管,进行kV平面成像期间的剂量测量。在各种成像条件下测量DAP,管电压为60至140 kV,管电流为2.5至20 ma。基于测量数据的二维三次插值方法估计了未测量成像条件下的DAP。结果:低剂量条件下与线性的偏差较大,在100 kV、2.5 ma时最大偏差为1.8%。然而,每个mAs值的总体DAP表现出高度线性(R2=1.00)。估计DAP的相对误差在2.0%以内。结论:在IGRT中,DAP仪对kV平面成像的剂量管理是有效的,即使在未测量的成像条件下,使用二维三次插值也能准确估计剂量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书