Robotic arm-assisted versus conventional jig-based unicompartmental knee arthroplasty achieves better soft-tissue balance : a unicompartmental pressure sensor study.

IF 4.8 1区 医学 Q1 ORTHOPEDICS
Tiejian Li, Xiangpeng Kong, Minzhi Yang, Zheng Cao, Xiang Li, Haifeng Li, Shaokui Nan, Wei Chai
{"title":"Robotic arm-assisted versus conventional jig-based unicompartmental knee arthroplasty achieves better soft-tissue balance : a unicompartmental pressure sensor study.","authors":"Tiejian Li, Xiangpeng Kong, Minzhi Yang, Zheng Cao, Xiang Li, Haifeng Li, Shaokui Nan, Wei Chai","doi":"10.1302/0301-620X.108B3.BJJ-2025-0090.R3","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>We developed a novel unicompartmental pressure sensor to measure intraoperative medial compartment pressureduring robotic arm-assisted unicompartmental knee arthroplasty (UKA) and conventional jig-based UKA in order to assess the effect and impact of soft-tissue balance.</p><p><strong>Methods: </strong>This retrospective study evaluated 80 patients who underwent medial compartment UKA at our institution between November 2022 and June 2023, including 40 patients in conventional jig-based (mobile bearing) UKA and 40 patients in robotic arm-assisted (fixed bearing) UKA. After intraoperative soft-tissue balancing, the medial compartment pressure at 10°, 45°, and 90° of knee flexion was measured using a new unicompartmental pressure sensor. The knee range of motion (ROM), Numerical Rating Scale, and American Knee Society scores were recorded at three months and one year of follow-up.</p><p><strong>Results: </strong>Conventional jig-based UKA did not achieve consistent medial compartment pressures at 10°, 45°, and 90° of knee flexion (p < 0.001), whereas robotic arm-assisted UKA achieved more balanced pressures at these angles (p-values between 10° vs 45°, 10° vs 90°, and 45° vs 90° were 0.132, 0.378, and > 0.999, respectively). In addition, robotic arm-assisted UKA showed better outcome scores and ROM at three months, but no significant clinical differences at the one-year follow-up.</p><p><strong>Conclusion: </strong>Intraoperative pressure measurements using the novel unicompartmental pressure sensor showed that robotic arm-assisted UKA was able to achieve more consistent pressure in the medial compartment than conventional jig-based UKA, which may improve early postoperative rehabilitation.</p>","PeriodicalId":48944,"journal":{"name":"Bone & Joint Journal","volume":"108-B 3","pages":"302-309"},"PeriodicalIF":4.8000,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bone & Joint Journal","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1302/0301-620X.108B3.BJJ-2025-0090.R3","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Aims: We developed a novel unicompartmental pressure sensor to measure intraoperative medial compartment pressureduring robotic arm-assisted unicompartmental knee arthroplasty (UKA) and conventional jig-based UKA in order to assess the effect and impact of soft-tissue balance.

Methods: This retrospective study evaluated 80 patients who underwent medial compartment UKA at our institution between November 2022 and June 2023, including 40 patients in conventional jig-based (mobile bearing) UKA and 40 patients in robotic arm-assisted (fixed bearing) UKA. After intraoperative soft-tissue balancing, the medial compartment pressure at 10°, 45°, and 90° of knee flexion was measured using a new unicompartmental pressure sensor. The knee range of motion (ROM), Numerical Rating Scale, and American Knee Society scores were recorded at three months and one year of follow-up.

Results: Conventional jig-based UKA did not achieve consistent medial compartment pressures at 10°, 45°, and 90° of knee flexion (p < 0.001), whereas robotic arm-assisted UKA achieved more balanced pressures at these angles (p-values between 10° vs 45°, 10° vs 90°, and 45° vs 90° were 0.132, 0.378, and > 0.999, respectively). In addition, robotic arm-assisted UKA showed better outcome scores and ROM at three months, but no significant clinical differences at the one-year follow-up.

Conclusion: Intraoperative pressure measurements using the novel unicompartmental pressure sensor showed that robotic arm-assisted UKA was able to achieve more consistent pressure in the medial compartment than conventional jig-based UKA, which may improve early postoperative rehabilitation.

机械臂辅助与传统的基于夹具的单室膝关节置换术实现更好的软组织平衡:一项单室压力传感器研究。
目的:我们开发了一种新型的单室压力传感器,用于测量机械臂辅助单室膝关节置换术(UKA)和传统的基于夹具的UKA术中内侧室压力,以评估软组织平衡的效果和影响。方法:本回顾性研究评估了我们机构在2022年11月至2023年6月期间接受内侧腔室UKA的80例患者,其中40例为传统的基于夹具(移动轴承)的UKA, 40例为机械臂辅助(固定轴承)的UKA。术中软组织平衡后,使用新的单室压力传感器测量膝关节屈曲10°、45°和90°处的内侧室压力。随访3个月和1年,分别记录膝关节活动度(ROM)、数值评定量表和美国膝关节协会评分。结果:传统的基于夹具的UKA在膝关节屈曲10°、45°和90°处不能实现一致的内侧腔室压力(p < 0.001),而机械臂辅助的UKA在这些角度上实现了更平衡的压力(10°vs 45°、10°vs 90°和45°vs 90°之间的p值分别为0.132、0.378和> 0.999)。此外,机械臂辅助的UKA在3个月时显示出更好的结局评分和ROM,但在1年随访时没有显着的临床差异。结论:使用新型单室压力传感器进行术中压力测量表明,机械臂辅助的UKA比传统的基于夹具的UKA能够在内侧室获得更一致的压力,这可能会改善术后早期康复。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Bone & Joint Journal
Bone & Joint Journal ORTHOPEDICS-SURGERY
CiteScore
9.40
自引率
10.90%
发文量
318
期刊介绍: We welcome original articles from any part of the world. The papers are assessed by members of the Editorial Board and our international panel of expert reviewers, then either accepted for publication or rejected by the Editor. We receive over 2000 submissions each year and accept about 250 for publication, many after revisions recommended by the reviewers, editors or statistical advisers. A decision usually takes between six and eight weeks. Each paper is assessed by two reviewers with a special interest in the subject covered by the paper, and also by members of the editorial team. Controversial papers will be discussed at a full meeting of the Editorial Board. Publication is between four and six months after acceptance.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书