{"title":"Robotic arm-assisted versus conventional jig-based unicompartmental knee arthroplasty achieves better soft-tissue balance : a unicompartmental pressure sensor study.","authors":"Tiejian Li, Xiangpeng Kong, Minzhi Yang, Zheng Cao, Xiang Li, Haifeng Li, Shaokui Nan, Wei Chai","doi":"10.1302/0301-620X.108B3.BJJ-2025-0090.R3","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>We developed a novel unicompartmental pressure sensor to measure intraoperative medial compartment pressureduring robotic arm-assisted unicompartmental knee arthroplasty (UKA) and conventional jig-based UKA in order to assess the effect and impact of soft-tissue balance.</p><p><strong>Methods: </strong>This retrospective study evaluated 80 patients who underwent medial compartment UKA at our institution between November 2022 and June 2023, including 40 patients in conventional jig-based (mobile bearing) UKA and 40 patients in robotic arm-assisted (fixed bearing) UKA. After intraoperative soft-tissue balancing, the medial compartment pressure at 10°, 45°, and 90° of knee flexion was measured using a new unicompartmental pressure sensor. The knee range of motion (ROM), Numerical Rating Scale, and American Knee Society scores were recorded at three months and one year of follow-up.</p><p><strong>Results: </strong>Conventional jig-based UKA did not achieve consistent medial compartment pressures at 10°, 45°, and 90° of knee flexion (p < 0.001), whereas robotic arm-assisted UKA achieved more balanced pressures at these angles (p-values between 10° vs 45°, 10° vs 90°, and 45° vs 90° were 0.132, 0.378, and > 0.999, respectively). In addition, robotic arm-assisted UKA showed better outcome scores and ROM at three months, but no significant clinical differences at the one-year follow-up.</p><p><strong>Conclusion: </strong>Intraoperative pressure measurements using the novel unicompartmental pressure sensor showed that robotic arm-assisted UKA was able to achieve more consistent pressure in the medial compartment than conventional jig-based UKA, which may improve early postoperative rehabilitation.</p>","PeriodicalId":48944,"journal":{"name":"Bone & Joint Journal","volume":"108-B 3","pages":"302-309"},"PeriodicalIF":4.8000,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bone & Joint Journal","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1302/0301-620X.108B3.BJJ-2025-0090.R3","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Aims: We developed a novel unicompartmental pressure sensor to measure intraoperative medial compartment pressureduring robotic arm-assisted unicompartmental knee arthroplasty (UKA) and conventional jig-based UKA in order to assess the effect and impact of soft-tissue balance.
Methods: This retrospective study evaluated 80 patients who underwent medial compartment UKA at our institution between November 2022 and June 2023, including 40 patients in conventional jig-based (mobile bearing) UKA and 40 patients in robotic arm-assisted (fixed bearing) UKA. After intraoperative soft-tissue balancing, the medial compartment pressure at 10°, 45°, and 90° of knee flexion was measured using a new unicompartmental pressure sensor. The knee range of motion (ROM), Numerical Rating Scale, and American Knee Society scores were recorded at three months and one year of follow-up.
Results: Conventional jig-based UKA did not achieve consistent medial compartment pressures at 10°, 45°, and 90° of knee flexion (p < 0.001), whereas robotic arm-assisted UKA achieved more balanced pressures at these angles (p-values between 10° vs 45°, 10° vs 90°, and 45° vs 90° were 0.132, 0.378, and > 0.999, respectively). In addition, robotic arm-assisted UKA showed better outcome scores and ROM at three months, but no significant clinical differences at the one-year follow-up.
Conclusion: Intraoperative pressure measurements using the novel unicompartmental pressure sensor showed that robotic arm-assisted UKA was able to achieve more consistent pressure in the medial compartment than conventional jig-based UKA, which may improve early postoperative rehabilitation.
期刊介绍:
We welcome original articles from any part of the world. The papers are assessed by members of the Editorial Board and our international panel of expert reviewers, then either accepted for publication or rejected by the Editor. We receive over 2000 submissions each year and accept about 250 for publication, many after revisions recommended by the reviewers, editors or statistical advisers. A decision usually takes between six and eight weeks. Each paper is assessed by two reviewers with a special interest in the subject covered by the paper, and also by members of the editorial team. Controversial papers will be discussed at a full meeting of the Editorial Board. Publication is between four and six months after acceptance.