Effect of femoral component design and quadriceps load on patellofemoral kinematics after total knee arthroplasty: an in vitro cadaveric study.

IF 4.4 Q2 Medicine
Vera Maioli, Emanuele Diquattro, Michele Conconi, Francesco Castagnini, Francesco Traina, Nicola Sancisi, Luca Cristofolini
{"title":"Effect of femoral component design and quadriceps load on patellofemoral kinematics after total knee arthroplasty: an in vitro cadaveric study.","authors":"Vera Maioli, Emanuele Diquattro, Michele Conconi, Francesco Castagnini, Francesco Traina, Nicola Sancisi, Luca Cristofolini","doi":"10.1186/s43019-026-00308-6","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Patellofemoral (PF) complications are a common cause of dissatisfaction and revision following total knee arthroplasty (TKA), often linked to altered kinematics and implant design. \"Patella-friendly\" femoral components with a wider, funnel-shaped trochlear groove may better restore native patellar motion. This study evaluated PF kinematics both before and after TKA performed using kinematic alignment, investigating the role of implant design and quadriceps loading.</p><p><strong>Methods: </strong>In total, 12 paired fresh-frozen cadaveric lower limbs were tested before and after TKA. Within each pair, one limb received a traditional medial pivot femoral component, while the contralateral limb received a \"patella-friendly\" medial pivot femoral component. Native and implanted knees were tested by flexing the knee under the action of an external load applied through the quadriceps tendon, varying its magnitudes (20, 160, 280N) and directions in the frontal (neutral,±6°, ±12°) and sagittal plane (neutral, +5° anterior). Motion was captured using an eight-camera optoelectronic system.</p><p><strong>Results: </strong>In the reference condition (20N, neutral direction), neither design showed statistical differences versus native (p > 0.05). However, the patella excursion in varus-valgus rotation was much higher in the specimens implanted with the traditional femoral component design (35.1° versus 14.6° native) than with the patella-friendly (20.5°). Differences between the designs emerged mainly with quadriceps load variations, especially frontal direction changes, which significantly affected patellar motion in both native and implanted knees (p < 0.05). Overall, the patella-friendly design better reproduced native kinematics under most conditions. However, with extreme medial loading (12°), three out of six specimens implanted with the patella-friendly femoral component were untestable owing to instability, and others exhibited high lateral displacement and trochlear dysplasia. In contrast, all traditional design implants remained stable, though with greater deviation from native kinematics.</p><p><strong>Conclusions: </strong>This study provides foundational insights into PF biomechanics before and after TKA with kinematic alignment. By analyzing the interplay between implant geometry and quadriceps loading direction, it emphasizes the importance of selecting femoral components on the basis of individual patient anatomy. Our findings suggest that patella-friendly femoral components-although capable of better reproducing native motion in some cases-may not be suitable for patients with medially directed quadriceps forces or severely varus morphotypes.</p>","PeriodicalId":36317,"journal":{"name":"Knee Surgery and Related Research","volume":"38 1","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2026-02-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12937521/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Knee Surgery and Related Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s43019-026-00308-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Patellofemoral (PF) complications are a common cause of dissatisfaction and revision following total knee arthroplasty (TKA), often linked to altered kinematics and implant design. "Patella-friendly" femoral components with a wider, funnel-shaped trochlear groove may better restore native patellar motion. This study evaluated PF kinematics both before and after TKA performed using kinematic alignment, investigating the role of implant design and quadriceps loading.

Methods: In total, 12 paired fresh-frozen cadaveric lower limbs were tested before and after TKA. Within each pair, one limb received a traditional medial pivot femoral component, while the contralateral limb received a "patella-friendly" medial pivot femoral component. Native and implanted knees were tested by flexing the knee under the action of an external load applied through the quadriceps tendon, varying its magnitudes (20, 160, 280N) and directions in the frontal (neutral,±6°, ±12°) and sagittal plane (neutral, +5° anterior). Motion was captured using an eight-camera optoelectronic system.

Results: In the reference condition (20N, neutral direction), neither design showed statistical differences versus native (p > 0.05). However, the patella excursion in varus-valgus rotation was much higher in the specimens implanted with the traditional femoral component design (35.1° versus 14.6° native) than with the patella-friendly (20.5°). Differences between the designs emerged mainly with quadriceps load variations, especially frontal direction changes, which significantly affected patellar motion in both native and implanted knees (p < 0.05). Overall, the patella-friendly design better reproduced native kinematics under most conditions. However, with extreme medial loading (12°), three out of six specimens implanted with the patella-friendly femoral component were untestable owing to instability, and others exhibited high lateral displacement and trochlear dysplasia. In contrast, all traditional design implants remained stable, though with greater deviation from native kinematics.

Conclusions: This study provides foundational insights into PF biomechanics before and after TKA with kinematic alignment. By analyzing the interplay between implant geometry and quadriceps loading direction, it emphasizes the importance of selecting femoral components on the basis of individual patient anatomy. Our findings suggest that patella-friendly femoral components-although capable of better reproducing native motion in some cases-may not be suitable for patients with medially directed quadriceps forces or severely varus morphotypes.

股骨假体设计和股四头肌负荷对全膝关节置换术后髌股运动学的影响:一项体外尸体研究。
目的:髌骨股骨(PF)并发症是全膝关节置换术(TKA)后不满意和翻修的常见原因,通常与运动学和植入物设计的改变有关。“髌骨友好型”股骨假体具有更宽的漏斗状滑车沟,可以更好地恢复原有的髌骨运动。本研究评估了使用运动学对齐进行TKA前后的PF运动学,研究了种植体设计和股四头肌负荷的作用。方法:对12对新鲜冷冻尸体下肢进行TKA前后检测。在每对假肢中,一侧肢体接受传统的股骨内侧枢轴假体,而对侧肢体接受“髌骨友好型”股骨内侧枢轴假体。在通过股四头肌肌腱施加的外部负荷作用下,改变膝关节的大小(20、160、280N)和正前方(中性,±6°,±12°)和矢状面(中性,+5°)方向,通过弯曲膝关节来测试天然膝关节和植入膝关节。运动是由一个八摄像头光电系统捕获的。结果:在参考条件(20N,中性方向),两种设计均与自然条件比较,差异均有统计学意义(p < 0.05)。然而,植入传统股骨假体设计的标本(35.1°对14.6°)在内翻旋转中髌骨偏移远高于髌骨友好型(20.5°)。不同设计之间的差异主要体现在股四头肌负荷的变化,尤其是前部方向的变化,这显著影响了原生膝关节和植入膝关节的髌骨运动(结论:本研究为TKA前后的PF生物力学提供了基础见解。通过分析植入物几何形状与股四头肌载荷方向之间的相互作用,强调了根据患者个体解剖结构选择股骨假体的重要性。我们的研究结果表明,髌骨友好型股骨假体-尽管在某些情况下能够更好地再现自然运动-可能不适合具有内侧定向股四头肌力量或严重内翻形态的患者。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.50
自引率
0.00%
发文量
42
审稿时长
19 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书