Universal Fiducial Markers for Multi-Modal Radiotherapy.

IF 2 Q3 ONCOLOGY
International Journal of Particle Therapy Pub Date : 2026-02-10 eCollection Date: 2026-03-01 DOI:10.1016/j.ijpt.2026.101308
Jie Liu, Caroline Bamberger, Naveed Iqbal, Kevin Reynolds, Smit Shah, Xing Li, Peng Wang, Jiajin Fan, Ashish Chawla, Kevin Choe, Daniel Kim, Robabeh Rahimi
{"title":"Universal Fiducial Markers for Multi-Modal Radiotherapy.","authors":"Jie Liu, Caroline Bamberger, Naveed Iqbal, Kevin Reynolds, Smit Shah, Xing Li, Peng Wang, Jiajin Fan, Ashish Chawla, Kevin Choe, Daniel Kim, Robabeh Rahimi","doi":"10.1016/j.ijpt.2026.101308","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Integrating multiple radiotherapy modalities, including photon and proton therapy, within a single clinical facility requires universal fiducial markers that perform reliably across all imaging and treatment systems. For computed tomography (CT) and image-guided radiotherapy (IGRT), fiducials should be large enough to ensure visibility without introducing significant image artifacts. In CyberKnife treatments, a fiducial diameter of at least 0.75 mm is necessary to maintain accurate tracking, particularly in the abdominal and pelvic regions. However, fiducials of this size are unsuitable for proton therapy due to the considerable beam perturbation they cause. This study investigates the design and implementation of customized fiducials that balance imaging visibility, tracking reliability, and proton beam perturbation, enabling their use across multiple radiotherapy modalities.</p><p><strong>Materials/methods: </strong>Customized coil-shaped platinum fiducials with a 0.75 mm diameter and variable coil gaps were developed to reduce the amount of metal in the proton beam path. For image visibility and trackability assessments, both standard fiducials (0.5 and 0.75 mm diameters with nominally zero coil gaps) and customized versions were inserted in an anthropomorphic phantom. Visibility was evaluated using cone-beam CT (CBCT) and kilovoltage (kV) imaging on both photon and proton treatment systems. Trackability was assessed in the CyberKnife system using an anthropomorphic phantom and a respiratory motion platform. Proton beam perturbation was measured using radiochromic films that were placed downstream of the fiducials and irradiated with a broad spread-out Bragg peak (SOBP) field. A triple-channel film dosimetry method was employed for analysis.</p><p><strong>Results: </strong>Customized fiducials with increased coil gaps demonstrated visibility comparable to standard fiducials in IGRT for both photon and proton systems. In the CyberKnife system, these customized fiducials also provided reliable tracking performance under both static conditions and simulated respiratory motion. The film measurements revealed that the customized fiducials with coil gaps larger than 0.5 mm/coil could significantly reduce proton beam perturbation compared to both the standard 0.75 and 0.5 mm diameter fiducials.</p><p><strong>Conclusions: </strong>This study demonstrates the feasibility of redesigning fiducial markers to maintain a large diameter for reliable imaging visibility and trackability while minimizing proton beam perturbation by increasing the coil gap. Adopting a universal fiducial marker design has the potential to streamline clinical workflows and support seamless integration across various radiotherapy modalities.</p>","PeriodicalId":36923,"journal":{"name":"International Journal of Particle Therapy","volume":"19 ","pages":"101308"},"PeriodicalIF":2.0000,"publicationDate":"2026-02-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12927308/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Particle Therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.ijpt.2026.101308","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/3/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Integrating multiple radiotherapy modalities, including photon and proton therapy, within a single clinical facility requires universal fiducial markers that perform reliably across all imaging and treatment systems. For computed tomography (CT) and image-guided radiotherapy (IGRT), fiducials should be large enough to ensure visibility without introducing significant image artifacts. In CyberKnife treatments, a fiducial diameter of at least 0.75 mm is necessary to maintain accurate tracking, particularly in the abdominal and pelvic regions. However, fiducials of this size are unsuitable for proton therapy due to the considerable beam perturbation they cause. This study investigates the design and implementation of customized fiducials that balance imaging visibility, tracking reliability, and proton beam perturbation, enabling their use across multiple radiotherapy modalities.

Materials/methods: Customized coil-shaped platinum fiducials with a 0.75 mm diameter and variable coil gaps were developed to reduce the amount of metal in the proton beam path. For image visibility and trackability assessments, both standard fiducials (0.5 and 0.75 mm diameters with nominally zero coil gaps) and customized versions were inserted in an anthropomorphic phantom. Visibility was evaluated using cone-beam CT (CBCT) and kilovoltage (kV) imaging on both photon and proton treatment systems. Trackability was assessed in the CyberKnife system using an anthropomorphic phantom and a respiratory motion platform. Proton beam perturbation was measured using radiochromic films that were placed downstream of the fiducials and irradiated with a broad spread-out Bragg peak (SOBP) field. A triple-channel film dosimetry method was employed for analysis.

Results: Customized fiducials with increased coil gaps demonstrated visibility comparable to standard fiducials in IGRT for both photon and proton systems. In the CyberKnife system, these customized fiducials also provided reliable tracking performance under both static conditions and simulated respiratory motion. The film measurements revealed that the customized fiducials with coil gaps larger than 0.5 mm/coil could significantly reduce proton beam perturbation compared to both the standard 0.75 and 0.5 mm diameter fiducials.

Conclusions: This study demonstrates the feasibility of redesigning fiducial markers to maintain a large diameter for reliable imaging visibility and trackability while minimizing proton beam perturbation by increasing the coil gap. Adopting a universal fiducial marker design has the potential to streamline clinical workflows and support seamless integration across various radiotherapy modalities.

多模态放射治疗的通用基准标志物。
目的:整合多种放射治疗方式,包括光子和质子治疗,在一个单一的临床设施中需要通用的基准标记,在所有成像和治疗系统中可靠地执行。对于计算机断层扫描(CT)和图像引导放射治疗(IGRT),基准应该足够大,以确保可视性,而不会引入明显的图像伪影。在射波刀治疗中,至少0.75毫米的基准直径是保持准确跟踪的必要条件,特别是在腹部和骨盆区域。然而,这种尺寸的基准不适合质子治疗,因为它们会引起相当大的光束扰动。本研究探讨了平衡成像可见性、跟踪可靠性和质子束扰动的定制基准的设计和实现,使其能够跨多种放射治疗模式使用。材料/方法:开发了直径为0.75 mm和可变线圈间隙的定制线圈形铂基准,以减少质子束路径中的金属量。为了评估图像可见性和可追踪性,将标准基准(0.5和0.75毫米直径,名义上线圈间隙为零)和定制版本插入拟人化幻影中。使用锥束CT (CBCT)和千伏(kV)成像对光子和质子处理系统进行可视性评估。在射波刀系统中使用拟人化幻影和呼吸运动平台评估可追踪性。质子束摄动是用放色膜测量的,放色膜放置在基准的下游,用宽展布喇格峰(SOBP)场照射。采用三通道膜剂量法进行分析。结果:在光子和质子系统的IGRT中,增加线圈间隙的定制基准显示出与标准基准相当的可见性。在射波刀系统中,这些定制的基准在静态条件和模拟呼吸运动下都能提供可靠的跟踪性能。薄膜测量表明,与标准的0.75和0.5 mm直径基准相比,线圈间隙大于0.5 mm/圈的定制基准可以显著减少质子束微扰。结论:本研究证明了重新设计基准标记的可行性,以保持可靠的成像可见性和可追踪性的大直径,同时通过增加线圈间隙来最小化质子束扰动。采用通用基准标记物设计有可能简化临床工作流程,并支持各种放疗方式的无缝集成。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal of Particle Therapy
International Journal of Particle Therapy Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
3.70
自引率
5.90%
发文量
23
审稿时长
20 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书