{"title":"Instrument traceability: Optimisation of Datamatrix readers by an experimental design","authors":"Catherine Guimier Pingault, Clement Richard, Gregoire Murison, Nathalie Sylvoz, Philippe Py, Cordelia Salomez-Ihl, Pierrick Bedouch","doi":"10.1016/j.pharma.2026.02.002","DOIUrl":null,"url":null,"abstract":"<div><div>Implementing traceability at instrument's individual level showed that Datamatrix readers were unable to read the Data Matrix barcodes effectively. This led to slow down medical procedure trays repackaging's stage and a great reluctance from operators to continue the Datamatrix process deployment. To improve reader performance, an experimental design was developed using the Taguchi method. This experimental design made possible, in a limited number of trials, to identify influencing parameters and their optimal settings. The reading time of instruments was reduced by 81% while using the leanest possible production resources. Thanks to this improvement, the Sterilisation department team has been able to continue ramping up individual level instrument traceability with greater efficiency and increased satisfaction of the medical staff.</div></div><div><div>Suite à la mise en place de la traçabilité à l’instrument, il a été mis en évidence que les lecteurs ne parvenaient pas à lire efficacement les Datamatrix. Cela a entraîné des ralentissements de l’étape de recomposition et une grande réticence de la part des opérateurs dans la poursuite du déploiement. Pour améliorer la performance des lecteurs, un Plan d’Expérience a été conçu en suivant la méthode de Taguchi. Ce plan d’expérience a permis, en un nombre d’essais limité, d’identifier les paramètres d’influence ainsi que leurs réglages optimaux. Le temps de lecture des instruments a ainsi été réduit de 81 % en utilisant le minimum de ressources de production possible. Grâce à cette amélioration, l’équipe du service de stérilisation a pu poursuivre la montée en charge de la traçabilité à l’instrument avec plus d’efficacité et de satisfaction des professionnels.</div></div>","PeriodicalId":8332,"journal":{"name":"Annales pharmaceutiques francaises","volume":"84 4","pages":"Pages 696-702"},"PeriodicalIF":1.3000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales pharmaceutiques francaises","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0003450926000271","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/2/14 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0
Abstract
Implementing traceability at instrument's individual level showed that Datamatrix readers were unable to read the Data Matrix barcodes effectively. This led to slow down medical procedure trays repackaging's stage and a great reluctance from operators to continue the Datamatrix process deployment. To improve reader performance, an experimental design was developed using the Taguchi method. This experimental design made possible, in a limited number of trials, to identify influencing parameters and their optimal settings. The reading time of instruments was reduced by 81% while using the leanest possible production resources. Thanks to this improvement, the Sterilisation department team has been able to continue ramping up individual level instrument traceability with greater efficiency and increased satisfaction of the medical staff.
Suite à la mise en place de la traçabilité à l’instrument, il a été mis en évidence que les lecteurs ne parvenaient pas à lire efficacement les Datamatrix. Cela a entraîné des ralentissements de l’étape de recomposition et une grande réticence de la part des opérateurs dans la poursuite du déploiement. Pour améliorer la performance des lecteurs, un Plan d’Expérience a été conçu en suivant la méthode de Taguchi. Ce plan d’expérience a permis, en un nombre d’essais limité, d’identifier les paramètres d’influence ainsi que leurs réglages optimaux. Le temps de lecture des instruments a ainsi été réduit de 81 % en utilisant le minimum de ressources de production possible. Grâce à cette amélioration, l’équipe du service de stérilisation a pu poursuivre la montée en charge de la traçabilité à l’instrument avec plus d’efficacité et de satisfaction des professionnels.
期刊介绍:
This journal proposes a scientific information validated and indexed to be informed about the last research works in all the domains interesting the pharmacy. The original works, general reviews, the focusing, the brief notes, subjected by the best academics and the professionals, propose a synthetic approach of the last progress accomplished in the concerned sectors. The thematic Sessions and the – life of the Academy – resume the communications which, presented in front of the national Academy of pharmacy, are in the heart of the current events.