Injection method alters diagnostic confidence and quantitative metrics in [18F]fluorocholine PET/CT for parathyroid adenoma localization.

IF 1.7
Mette Rode, Hanne Reilev Møller, Peymaneh Abadi, Henriette Juel Nissen, Paw Christian Holdgaard
{"title":"Injection method alters diagnostic confidence and quantitative metrics in [<sup>18</sup>F]fluorocholine PET/CT for parathyroid adenoma localization.","authors":"Mette Rode, Hanne Reilev Møller, Peymaneh Abadi, Henriette Juel Nissen, Paw Christian Holdgaard","doi":"10.1186/s41824-026-00285-9","DOIUrl":null,"url":null,"abstract":"<p><p>BACKGROUND: Studies evaluating [18F]fluorocholine PET/CT for parathyroid adenoma localisation fail to report the injection method used. At our department, manual injection method (MIM) was initially used, before converting to an automated injection system (AIS). Afterwards, physicians observed increased retention of the radiotracer within the veins leading from the injection site, which had not been as present previously. The aim of this study was to evaluate whether the injection method had a clinical impact. METHODS: The study compared two groups consisting of 60 patients, where the injection method was the only difference. Two physicians, blinded to injection method, evaluated if retention was present and interpretation was jeopardised. Quantitative measures were determined based on the original diagnostic confidence scores (DCS) for each suspected lesion, as these were compared between patient groups. Additionally, the maximum standardised uptake values (SUVmax) were measured for the suspected lesions along with the thyroid and salivary glands as a physiological reference point. Comparisons of the quantitative parameters were also performed between patients with or without retention. RESULTS: There was significant more occurrence of retention in AIS-group (83.3%) than MIM group (36.7%), (p < 0.00001). Retention did not influence the interpretation of any patient. A significant difference of DCS was observed, as 18% of lesions in the AIS-group had a low score, compared to only 1% in the MIM-group (p < 0.001). SUVmax was significantly lower in the AIS-group than the MIM-group for the reported adenomas (10.1 vs. 11.8, p = 0.008). In addition, SUVmax in the AIS-group was significantly lower in both the thyroid (4.6 vs. 5.2, p = 0.023) and salivary glands (10.8 vs. 12.4, p < 0.001). Comparison between patients with and without retention revealed similar differences in DCS and SUVmax for adenomas. CONCLUSION: The AIS is associated with a significant higher presence of retention, low diagnostic confidence and overall lower SUVmax, which indicates an aspect of clinical impact. It could potentially lead to a parathyroid adenoma being missed or classified as normal parathyroid tissue. Thus, manual injection method might be the method of choice, to ensure the most optimal clinical outcome. </p>","PeriodicalId":519909,"journal":{"name":"EJNMMI reports","volume":"10 1","pages":"3"},"PeriodicalIF":1.7000,"publicationDate":"2026-02-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12876525/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41824-026-00285-9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

BACKGROUND: Studies evaluating [18F]fluorocholine PET/CT for parathyroid adenoma localisation fail to report the injection method used. At our department, manual injection method (MIM) was initially used, before converting to an automated injection system (AIS). Afterwards, physicians observed increased retention of the radiotracer within the veins leading from the injection site, which had not been as present previously. The aim of this study was to evaluate whether the injection method had a clinical impact. METHODS: The study compared two groups consisting of 60 patients, where the injection method was the only difference. Two physicians, blinded to injection method, evaluated if retention was present and interpretation was jeopardised. Quantitative measures were determined based on the original diagnostic confidence scores (DCS) for each suspected lesion, as these were compared between patient groups. Additionally, the maximum standardised uptake values (SUVmax) were measured for the suspected lesions along with the thyroid and salivary glands as a physiological reference point. Comparisons of the quantitative parameters were also performed between patients with or without retention. RESULTS: There was significant more occurrence of retention in AIS-group (83.3%) than MIM group (36.7%), (p < 0.00001). Retention did not influence the interpretation of any patient. A significant difference of DCS was observed, as 18% of lesions in the AIS-group had a low score, compared to only 1% in the MIM-group (p < 0.001). SUVmax was significantly lower in the AIS-group than the MIM-group for the reported adenomas (10.1 vs. 11.8, p = 0.008). In addition, SUVmax in the AIS-group was significantly lower in both the thyroid (4.6 vs. 5.2, p = 0.023) and salivary glands (10.8 vs. 12.4, p < 0.001). Comparison between patients with and without retention revealed similar differences in DCS and SUVmax for adenomas. CONCLUSION: The AIS is associated with a significant higher presence of retention, low diagnostic confidence and overall lower SUVmax, which indicates an aspect of clinical impact. It could potentially lead to a parathyroid adenoma being missed or classified as normal parathyroid tissue. Thus, manual injection method might be the method of choice, to ensure the most optimal clinical outcome.

Abstract Image

Abstract Image

Abstract Image

注射方法改变了[18F]氟胆碱PET/CT对甲状旁腺瘤定位的诊断可信度和定量指标。
背景:评价[18F]氟胆碱PET/CT对甲状旁腺瘤定位的研究未报道所使用的注射方法。在我们部门,最初使用手动注射方法(MIM),然后转换为自动注射系统(AIS)。之后,医生观察到从注射部位通向静脉的放射性示踪剂潴留增加,这在以前是不存在的。本研究的目的是评估注射方法是否有临床影响。方法:研究比较两组60例患者,其中注射方法是唯一的差异。两名医生,对注射方法不知情,评估是否存在保留和解释是否受到影响。根据每个疑似病变的原始诊断置信度评分(DCS)确定定量措施,并在患者组之间进行比较。此外,测量疑似病变的最大标准化摄取值(SUVmax),并将甲状腺和唾液腺作为生理参考点。在有或没有保留的患者之间也进行了定量参数的比较。结果:ais组患者留置发生率(83.3%)明显高于MIM组(36.7%)
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书