Algorithm for Selecting Organic Compounds to Verify the Resolution of Electron Ionization Mass Spectrometers.

Q3 Physics and Astronomy
Mass spectrometry Pub Date : 2025-01-01 Epub Date: 2025-10-28 DOI:10.5702/massspectrometry.A0176
Valentin G Tkachenko, Sergey V Silkin, Alexandr V Sakharov, Vasily A Eliferov, Denis V Kolesnik, Anastasia S Konstantinova, Evgeniy A Filatov, Ekaterina S Shiryaeva, Stanislav I Pekov, Igor A Popov
{"title":"Algorithm for Selecting Organic Compounds to Verify the Resolution of Electron Ionization Mass Spectrometers.","authors":"Valentin G Tkachenko, Sergey V Silkin, Alexandr V Sakharov, Vasily A Eliferov, Denis V Kolesnik, Anastasia S Konstantinova, Evgeniy A Filatov, Ekaterina S Shiryaeva, Stanislav I Pekov, Igor A Popov","doi":"10.5702/massspectrometry.A0176","DOIUrl":null,"url":null,"abstract":"<p><p>The ion coalescence phenomenon complicates the evaluation of the effective resolution of Fourier-transform mass spectrometers. We propose an approach for confirming the resolution of an electron ionization Fourier-transform mass spectrometer using pairs of organic substances identified by automatically generated formula differences. The proposed method is compared with the search for organic substances in the National Institute of Standards and Technology (NIST) database. Under the given conditions of the mass spectrometer resolution range up to 45000-50000 at 100 <i>m/z</i>, 166 pairs of suitable compounds were found using the proposed method, while a search in the NIST database yielded only 88 pairs of compounds. This enabled the selection of six pairs of organic compounds that were most suitable for confirming the resolution of the high-resolution mass spectrometer using molecular ion peaks, and four pairs of compounds that allowed the resolution to be confirmed using fragment ion peaks. The resolution of the Fourier-transform mass spectrometer designed for gas analysis was experimentally evaluated by analyzing the spectra of a mixture of organic compounds selected using the proposed method.</p>","PeriodicalId":18243,"journal":{"name":"Mass spectrometry","volume":"14 1","pages":"A0176"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12583954/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Mass spectrometry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5702/massspectrometry.A0176","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/10/28 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"Physics and Astronomy","Score":null,"Total":0}
引用次数: 0

Abstract

The ion coalescence phenomenon complicates the evaluation of the effective resolution of Fourier-transform mass spectrometers. We propose an approach for confirming the resolution of an electron ionization Fourier-transform mass spectrometer using pairs of organic substances identified by automatically generated formula differences. The proposed method is compared with the search for organic substances in the National Institute of Standards and Technology (NIST) database. Under the given conditions of the mass spectrometer resolution range up to 45000-50000 at 100 m/z, 166 pairs of suitable compounds were found using the proposed method, while a search in the NIST database yielded only 88 pairs of compounds. This enabled the selection of six pairs of organic compounds that were most suitable for confirming the resolution of the high-resolution mass spectrometer using molecular ion peaks, and four pairs of compounds that allowed the resolution to be confirmed using fragment ion peaks. The resolution of the Fourier-transform mass spectrometer designed for gas analysis was experimentally evaluated by analyzing the spectra of a mixture of organic compounds selected using the proposed method.

Abstract Image

Abstract Image

Abstract Image

验证电子电离质谱仪分辨率的有机化合物选择算法。
离子聚结现象使傅里叶变换质谱仪有效分辨率的评估复杂化。我们提出了一种方法来确认电子电离傅里叶变换质谱仪的分辨率使用有机物质对由自动生成的公式差异识别。将提出的方法与美国国家标准与技术研究院(NIST)数据库中有机物质的搜索进行了比较。在给定条件下,在100 m/z的分辨率范围达到45000-50000,使用该方法找到了166对合适的化合物,而在NIST数据库中搜索仅得到88对化合物。这使得选择六对有机化合物最适合使用分子离子峰来确认高分辨率质谱仪的分辨率,以及四对化合物允许使用片段离子峰来确认分辨率。通过对所选有机化合物混合物的光谱分析,对设计用于气体分析的傅里叶变换质谱仪的分辨率进行了实验评价。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Mass spectrometry
Mass spectrometry Physics and Astronomy-Instrumentation
CiteScore
1.90
自引率
0.00%
发文量
3
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书