Simultaneous dual-isotope SPECT imaging using a cardiac cadmium-zinc-telluride camera with various Technetium-99 m to Iodine-123 ratios.

IF 1.7
Takayuki Shibutani, Masahisa Onoguchi, Hiroto Yoneyama, Takahiro Konishi, Kenichi Nakajima
{"title":"Simultaneous dual-isotope SPECT imaging using a cardiac cadmium-zinc-telluride camera with various Technetium-99 m to Iodine-123 ratios.","authors":"Takayuki Shibutani, Masahisa Onoguchi, Hiroto Yoneyama, Takahiro Konishi, Kenichi Nakajima","doi":"10.1186/s41824-025-00272-6","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Cadmium-zinc-telluride (CZT) SPECT cameras offer high energy resolution. This allows simultaneous dual-isotope (SDI) acquisition even when <sup>123</sup>I and <sup>99m</sup>Tc photon peaks are close. This study aimed to reveal the impact of image quality for <sup>99m</sup>Tc/<sup>123</sup>I ratios on SDI images.</p><p><strong>Methods: </strong>We created normal and inferior wall defect myocardial models using an anthropomorphic myocardial phantom. The doses of <sup>99m</sup>Tc/<sup>123</sup>I ratios were set to 1.0, 3.5, 5.0 and 6.5 (referred to as Tc/I_1.0, Tc/I_3.5, Tc/I_5.0 and Tc/I_6.5 conditions). We acquired SPECT images using D-SPECT Cardio. The acquisition time was adjusted to achieve 1.5 × 10<sup>6</sup> left ventricular <sup>123</sup>I counts as a clinical reference. Short-axis images were reconstructed with (SC) and without (NC) scatter correction. Kernel and Gaussian-standard values were set to 1 and 1.0 as default conditions. The filter parameters were changed to 1, 3 and 5 for the kernel settings, and 0.25-1.0 for Gaussian standards. The image quality of normal and defective myocardia at different <sup>99m</sup>Tc/<sup>123</sup>I ratios was evaluated as cavity contrast on short-axis images, the percent coefficient of variance (%CV) and %uptake on polar maps. In addition, these reconstruction conditions were applied to a patient with acute coronary syndrome, supporting the phantom-based findings.</p><p><strong>Results: </strong>In the normal myocardial phantom, %uptake in the inferior and inferolateral myocardial walls was significantly lower in SC images than in NC images for both <sup>99m</sup>Tc and <sup>123</sup>I. In SC images, <sup>99m</sup>Tc counts in the inferoseptal myocardial wall slightly decreased along with the <sup>99m</sup>Tc/<sup>123</sup>I ratio, whereas <sup>123</sup>I counts in the inferolateral walls declined as the <sup>99m</sup>Tc/<sup>123</sup>I ratio increased to 6.5. Higher kernel and Gaussian standard settings resulted in lower %CV values in both NC and SC <sup>99m</sup>Tc and <sup>123</sup>I images, improving image uniformity. Overall, NC images showed more homogeneous myocardial distribution in the inferior wall compared with SC images.</p><p><strong>Conclusion: </strong>Image quality and correction effects are influenced by <sup>99m</sup>Tc/<sup>123</sup>I ratios, resulting in heterogeneity in the inferior wall of the normal myocardium due to SC. Adjusting kernel parameters of the Gaussian filter or using NC images may help improve image quality when <sup>99m</sup>Tc/<sup>123</sup>I ratios are very high or low.</p>","PeriodicalId":519909,"journal":{"name":"EJNMMI reports","volume":"9 1","pages":"36"},"PeriodicalIF":1.7000,"publicationDate":"2025-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12535947/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41824-025-00272-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Cadmium-zinc-telluride (CZT) SPECT cameras offer high energy resolution. This allows simultaneous dual-isotope (SDI) acquisition even when 123I and 99mTc photon peaks are close. This study aimed to reveal the impact of image quality for 99mTc/123I ratios on SDI images.

Methods: We created normal and inferior wall defect myocardial models using an anthropomorphic myocardial phantom. The doses of 99mTc/123I ratios were set to 1.0, 3.5, 5.0 and 6.5 (referred to as Tc/I_1.0, Tc/I_3.5, Tc/I_5.0 and Tc/I_6.5 conditions). We acquired SPECT images using D-SPECT Cardio. The acquisition time was adjusted to achieve 1.5 × 106 left ventricular 123I counts as a clinical reference. Short-axis images were reconstructed with (SC) and without (NC) scatter correction. Kernel and Gaussian-standard values were set to 1 and 1.0 as default conditions. The filter parameters were changed to 1, 3 and 5 for the kernel settings, and 0.25-1.0 for Gaussian standards. The image quality of normal and defective myocardia at different 99mTc/123I ratios was evaluated as cavity contrast on short-axis images, the percent coefficient of variance (%CV) and %uptake on polar maps. In addition, these reconstruction conditions were applied to a patient with acute coronary syndrome, supporting the phantom-based findings.

Results: In the normal myocardial phantom, %uptake in the inferior and inferolateral myocardial walls was significantly lower in SC images than in NC images for both 99mTc and 123I. In SC images, 99mTc counts in the inferoseptal myocardial wall slightly decreased along with the 99mTc/123I ratio, whereas 123I counts in the inferolateral walls declined as the 99mTc/123I ratio increased to 6.5. Higher kernel and Gaussian standard settings resulted in lower %CV values in both NC and SC 99mTc and 123I images, improving image uniformity. Overall, NC images showed more homogeneous myocardial distribution in the inferior wall compared with SC images.

Conclusion: Image quality and correction effects are influenced by 99mTc/123I ratios, resulting in heterogeneity in the inferior wall of the normal myocardium due to SC. Adjusting kernel parameters of the Gaussian filter or using NC images may help improve image quality when 99mTc/123I ratios are very high or low.

Abstract Image

Abstract Image

Abstract Image

同时双同位素SPECT成像使用心脏镉锌碲化照相机与不同的锝-99米碘-123比率。
目的:研究碲化镉锌(CZT) SPECT相机的高能量分辨率。这允许同时获得双同位素(SDI),即使在123I和99mTc光子峰接近的情况下。本研究旨在揭示99mTc/123I比例下图像质量对SDI图像的影响。方法:采用拟人心肌模型建立正常和下壁缺损心肌模型。99mTc/123I的剂量分别设定为1.0、3.5、5.0和6.5(简称Tc/I_1.0、Tc/I_3.5、Tc/I_5.0和Tc/I_6.5)。我们使用D-SPECT Cardio获取SPECT图像。调整采集时间,达到1.5 × 106左心室123I计数作为临床参考。用(SC)和不(NC)散射校正重建短轴图像。内核和高斯标准值分别设置为1和1.0作为默认条件。对于内核设置,过滤器参数更改为1,3和5,对于高斯标准,过滤器参数更改为0.25-1.0。在不同99mTc/123I比值下,对正常心肌和缺损心肌的图像质量进行短轴造影、百分比方差系数(%CV)和极坐标摄动率的评价。此外,这些重建条件应用于急性冠状动脉综合征患者,支持基于幻像的发现。结果:在正常心肌幻像中,99mTc和123I的SC图像中,下壁和外壁的%摄取明显低于NC图像。SC图像中,室间隔内心肌壁99mTc计数随99mTc/123I比值略有下降,而外壁123I计数随99mTc/123I比值升高至6.5而下降。较高的核和高斯标准设置导致NC和SC 99mTc和123I图像的%CV值较低,从而提高了图像的均匀性。总的来说,NC图像显示下壁心肌分布比SC图像更均匀。结论:99mTc/123I比值影响图像质量和校正效果,导致正常心肌下壁因SC而呈现异质性,当99mTc/123I比值过高或过低时,调整高斯滤波核参数或采用NC图像均可改善图像质量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书