Technical performance of dual-contrast-agent K-edge imaging with four energy thresholds on a commercial photon counting detector CT.

IF 3.2
Medical physics Pub Date : 2025-10-01 DOI:10.1002/mp.70063
Kevin J Treb, Jeffrey F Marsh, Elisabeth R Shanblatt, Tristan Nowak, Chelsea A S Dunning, George S K Fung, Bernhard T Schmidt, Cynthia H McCollough, Shuai Leng
{"title":"Technical performance of dual-contrast-agent K-edge imaging with four energy thresholds on a commercial photon counting detector CT.","authors":"Kevin J Treb, Jeffrey F Marsh, Elisabeth R Shanblatt, Tristan Nowak, Chelsea A S Dunning, George S K Fung, Bernhard T Schmidt, Cynthia H McCollough, Shuai Leng","doi":"10.1002/mp.70063","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Photon counting detector CT (PCD-CT) has the potential to sort detected x-ray photons into more than two energy bins to perform multi-material decomposition using K-edge materials without additional mathematical constraints/assumptions. The availability of multiple energy thresholds in PCD-CT opens the possibility for unique clinical applications such as simultaneous multi-contrast-agent imaging.</p><p><strong>Purpose: </strong>To evaluate the technical performance of a PCD-CT system with four energy thresholds for conducting multi-material decomposition of iodine, gadolinium, and water after a comprehensive system calibration procedure.</p><p><strong>Methods: </strong>Experiments were performed on a commercial PCD-CT scanner (NAEOTOM Alpha, Siemens Healthineers) using a research mode with four energy thresholds. Material calibration was performed by scanning contrast materials with known concentration placed in various sizes of cylindrical water-equivalent phantoms at various tube currents and positions. From these data, per-voxel contrast models were derived for each material by multiple linear regressions up to second order, where the predictor variables were combinations of tube current, phantom diameter, and distance to isocenter. After calibration, small (20 cm diameter) and large (30 cm × 40 cm) uniform multi-energy CT phantoms and an anthropomorphic thorax phantom (31 cm × 39 cm), each containing bone, soft tissue, and several quantitative inserts of iodine (0.2-5.0 mg/ml), gadolinium (0.3-5.0 mg/ml), and a mixture of both (5.0 mg/ml each) were scanned using the four-energy threshold mode at 140 kV with energy thresholds set at 20, 52, 75, and 82 keV. Images for each energy threshold were reconstructed with a filtered back-projection algorithm and a quantitative kernel (Qr44). Subsequent image-based material decomposition was performed on individual voxels of the energy threshold images using the models derived from the calibration process to generate water, iodine, and gadolinium basis images. Quantitative material concentrations and image noise magnitudes, and noise power spectra (NPS) were measured.</p><p><strong>Results: </strong>Iodine and gadolinium concentrations above 1.25 mg/ml were visible in the material basis images in all phantoms. Averaged quantitative accuracy for iodine and gadolinium in the small, large, and thorax phantoms were 0.39 ± 0.30 mg/ml, 0.96 ± 0.53 mg/ml, and 0.53 ± 0.35 mg/ml, respectively. In the water basis images, accuracy was within 24 HU in all phantoms. Noise magnitudes in iodine and gadolinium basis images were 2.91 and 3.52 times that of the lowest energy threshold image for the small phantom, and were 6.92 and 7.17 times that of the lowest threshold image for the large phantom. Noise texture did not change between the energy threshold and material basis images for either size phantom as measured by the NPS.</p><p><strong>Conclusions: </strong>After a comprehensive system calibration, a commercial PCD-CT with four energy thresholds performed multiple material decomposition of iodine, gadolinium, and water for concentrations above 1.25 mg/ml in human-sized phantoms. Noise amplification and quantitative accuracy from the decomposition greatly depended on object size.</p>","PeriodicalId":94136,"journal":{"name":"Medical physics","volume":"52 10","pages":"e70063"},"PeriodicalIF":3.2000,"publicationDate":"2025-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/mp.70063","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Photon counting detector CT (PCD-CT) has the potential to sort detected x-ray photons into more than two energy bins to perform multi-material decomposition using K-edge materials without additional mathematical constraints/assumptions. The availability of multiple energy thresholds in PCD-CT opens the possibility for unique clinical applications such as simultaneous multi-contrast-agent imaging.

Purpose: To evaluate the technical performance of a PCD-CT system with four energy thresholds for conducting multi-material decomposition of iodine, gadolinium, and water after a comprehensive system calibration procedure.

Methods: Experiments were performed on a commercial PCD-CT scanner (NAEOTOM Alpha, Siemens Healthineers) using a research mode with four energy thresholds. Material calibration was performed by scanning contrast materials with known concentration placed in various sizes of cylindrical water-equivalent phantoms at various tube currents and positions. From these data, per-voxel contrast models were derived for each material by multiple linear regressions up to second order, where the predictor variables were combinations of tube current, phantom diameter, and distance to isocenter. After calibration, small (20 cm diameter) and large (30 cm × 40 cm) uniform multi-energy CT phantoms and an anthropomorphic thorax phantom (31 cm × 39 cm), each containing bone, soft tissue, and several quantitative inserts of iodine (0.2-5.0 mg/ml), gadolinium (0.3-5.0 mg/ml), and a mixture of both (5.0 mg/ml each) were scanned using the four-energy threshold mode at 140 kV with energy thresholds set at 20, 52, 75, and 82 keV. Images for each energy threshold were reconstructed with a filtered back-projection algorithm and a quantitative kernel (Qr44). Subsequent image-based material decomposition was performed on individual voxels of the energy threshold images using the models derived from the calibration process to generate water, iodine, and gadolinium basis images. Quantitative material concentrations and image noise magnitudes, and noise power spectra (NPS) were measured.

Results: Iodine and gadolinium concentrations above 1.25 mg/ml were visible in the material basis images in all phantoms. Averaged quantitative accuracy for iodine and gadolinium in the small, large, and thorax phantoms were 0.39 ± 0.30 mg/ml, 0.96 ± 0.53 mg/ml, and 0.53 ± 0.35 mg/ml, respectively. In the water basis images, accuracy was within 24 HU in all phantoms. Noise magnitudes in iodine and gadolinium basis images were 2.91 and 3.52 times that of the lowest energy threshold image for the small phantom, and were 6.92 and 7.17 times that of the lowest threshold image for the large phantom. Noise texture did not change between the energy threshold and material basis images for either size phantom as measured by the NPS.

Conclusions: After a comprehensive system calibration, a commercial PCD-CT with four energy thresholds performed multiple material decomposition of iodine, gadolinium, and water for concentrations above 1.25 mg/ml in human-sized phantoms. Noise amplification and quantitative accuracy from the decomposition greatly depended on object size.

背景:光子计数检测器CT (PCD-CT)有潜力将检测到的x射线光子分类到两个以上的能量箱中,使用k边材料进行多材料分解,而无需额外的数学约束/假设。在PCD-CT中多重能量阈值的可用性为独特的临床应用提供了可能性,例如同时使用多种造影剂成像。目的:在一个全面的系统校准程序后,评估具有四个能量阈值的PCD-CT系统进行碘、钆和水的多物质分解的技术性能。方法:采用四能量阈值研究模式,在商用pc - ct扫描仪(NAEOTOM Alpha, Siemens Healthineers)上进行实验。材料校准是通过扫描已知浓度的对比材料,在不同的管电流和位置下,放置在不同尺寸的圆柱形水等效模型中。从这些数据中,通过二次多元线性回归得到了每种材料的逐体素对比模型,其中预测变量是管电流、模体直径和到等中心的距离的组合。校正后,使用四能阈值模式在140 kV下扫描小(直径20 cm)和大(30 cm × 40 cm)均匀多能CT幻象和拟人胸腔幻象(31 cm × 39 cm),每个幻象均含有骨骼、软组织和若干定量插入物碘(0.2-5.0 mg/ml)、钆(0.3-5.0 mg/ml),以及两者的混合物(各5.0 mg/ml),能量阈值设置为20、52、75和82 keV。利用滤波后的反投影算法和定量核(Qr44)对每个能量阈值的图像进行重构。随后的基于图像的材料分解使用从校准过程中得到的模型对能量阈值图像的单个体素进行,以生成水、碘和钆基图像。定量测量了材料浓度、图像噪声大小和噪声功率谱(NPS)。结果:所有幻影的物质基础图像均可见碘、钆浓度高于1.25 mg/ml。碘和钆在小影、大影和胸影中的平均定量准确度分别为0.39±0.30 mg/ml、0.96±0.53 mg/ml和0.53±0.35 mg/ml。在水基图像中,所有幻影的精度都在24 HU以内。碘基和钆基图像的噪声值分别是小幻像最低能量阈值图像的2.91倍和3.52倍,是大幻像最低能量阈值图像的6.92倍和7.17倍。在NPS测量的两种尺寸的图像中,噪声纹理在能量阈值和物质基础图像之间没有变化。结论:经过全面的系统校准,具有四个能量阈值的商用PCD-CT对浓度高于1.25 mg/ml的碘、钆和水在人体大小的幻影中进行了多次物质分解。噪声放大和分解的定量精度很大程度上取决于物体的大小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信