X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer.

IF 1.1 4区 医学 Q4 ONCOLOGY
Journal of Contemporary Brachytherapy Pub Date : 2025-08-01 Epub Date: 2025-08-29 DOI:10.5114/jcb.2025.154104
Jaroslav Šolc, Gustavo Kertzscher, Jana Šmoldasová, Vladimír Sochor, Peter Georgi
{"title":"X-ray spectra emitted by the electronic brachytherapy unit Papillon 50 measured with a CdTe spectrometer.","authors":"Jaroslav Šolc, Gustavo Kertzscher, Jana Šmoldasová, Vladimír Sochor, Peter Georgi","doi":"10.5114/jcb.2025.154104","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aim of this paper was to present the measured photon fluence spectra emitted by the electronic brachytherapy (eBT) unit Papillon 50.</p><p><strong>Material and methods: </strong>Spectra were collected with a compact CdTe spectrometer in air at a distance of 52 cm from the unit output window, for three different applicators attached to the unit and at several angles with respect to the beam axis.</p><p><strong>Results: </strong>The fluence spectra recorded for this device showed lower energy of the photon spectrum end-point than anticipated. The end-point energy was found to be (47.6 ±0.2) keV for the nominal high voltage bias of 50 kV set by the operator. Full-energy peaks in the spectra identified characteristic photons emitted by the rhenium anode, and by molybdenum and silver supposedly present in the inner coating of the applicators. An asymmetry in the half-value layer values of measured spectra was observed.</p><p><strong>Conclusions: </strong>The presented work provides a comprehensive dataset of photon fluence spectra for traceable dosimetry using the Papillon 50 eBT device, which is required, for example, for correcting the energy dependence of detectors or as an input for Monte Carlo calculations. Spectrometry and dose characterization studies are needed for eBT systems.</p>","PeriodicalId":51305,"journal":{"name":"Journal of Contemporary Brachytherapy","volume":"17 4","pages":"267-274"},"PeriodicalIF":1.1000,"publicationDate":"2025-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12489539/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Contemporary Brachytherapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5114/jcb.2025.154104","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/8/29 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The aim of this paper was to present the measured photon fluence spectra emitted by the electronic brachytherapy (eBT) unit Papillon 50.

Material and methods: Spectra were collected with a compact CdTe spectrometer in air at a distance of 52 cm from the unit output window, for three different applicators attached to the unit and at several angles with respect to the beam axis.

Results: The fluence spectra recorded for this device showed lower energy of the photon spectrum end-point than anticipated. The end-point energy was found to be (47.6 ±0.2) keV for the nominal high voltage bias of 50 kV set by the operator. Full-energy peaks in the spectra identified characteristic photons emitted by the rhenium anode, and by molybdenum and silver supposedly present in the inner coating of the applicators. An asymmetry in the half-value layer values of measured spectra was observed.

Conclusions: The presented work provides a comprehensive dataset of photon fluence spectra for traceable dosimetry using the Papillon 50 eBT device, which is required, for example, for correcting the energy dependence of detectors or as an input for Monte Carlo calculations. Spectrometry and dose characterization studies are needed for eBT systems.

Abstract Image

Abstract Image

Abstract Image

电子近距离放射治疗装置Papillon 50发射的x射线光谱用CdTe光谱仪测量。
目的:研究电子近距离放射治疗(eBT)装置Papillon 50发射的光子通量光谱。材料和方法:在距离单元输出窗口52厘米的空气中,使用紧凑型碲化镉光谱仪收集光谱,将三种不同的涂抹器连接到单元上,并在相对于光束轴的几个角度上收集光谱。结果:该装置记录的通量光谱显示光子光谱端点的能量低于预期。在操作者设定的50kv标称高压偏置下,终点能量为(47.6±0.2)keV。光谱中的全能峰确定了由铼阳极发射的特征光子,以及据推测存在于涂抹器内涂层中的钼和银。测量光谱的半值层值存在不对称性。结论:所提出的工作为使用Papillon 50 eBT装置进行可追溯剂量测定提供了一个全面的光子通量光谱数据集,例如,用于纠正探测器的能量依赖性或作为蒙特卡罗计算的输入。需要对eBT系统进行光谱和剂量表征研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Contemporary Brachytherapy
Journal of Contemporary Brachytherapy ONCOLOGY-RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
CiteScore
2.40
自引率
14.30%
发文量
54
审稿时长
16 weeks
期刊介绍: The “Journal of Contemporary Brachytherapy” is an international and multidisciplinary journal that will publish papers of original research as well as reviews of articles. Main subjects of the journal include: clinical brachytherapy, combined modality treatment, advances in radiobiology, hyperthermia and tumour biology, as well as physical aspects relevant to brachytherapy, particularly in the field of imaging, dosimetry and radiation therapy planning. Original contributions will include experimental studies of combined modality treatment, tumor sensitization and normal tissue protection, molecular radiation biology, and clinical investigations of cancer treatment in brachytherapy. Another field of interest will be the educational part of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信