Straightening Misfolded Microaxial Flow Pump in the Aorta Using a Snare Catheter

Q4 Medicine
Marko Šustić MD, PhD , Tomislav Jakljević MD, PhD , Zoran Miovski MD, PhD , Davorka Lulić MD , David Gobić MD, PhD , Luka Bastiančić MD , Josip Aničić MD , Vjekoslav Tomulić MD, PhD
{"title":"Straightening Misfolded Microaxial Flow Pump in the Aorta Using a Snare Catheter","authors":"Marko Šustić MD, PhD ,&nbsp;Tomislav Jakljević MD, PhD ,&nbsp;Zoran Miovski MD, PhD ,&nbsp;Davorka Lulić MD ,&nbsp;David Gobić MD, PhD ,&nbsp;Luka Bastiančić MD ,&nbsp;Josip Aničić MD ,&nbsp;Vjekoslav Tomulić MD, PhD","doi":"10.1016/j.jaccas.2025.105218","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>In patients with cardiogenic shock, misfolding of a microaxial flow pump within the aorta after deployment is a rare but serious complication that presents a significant therapeutic challenge.</div></div><div><h3>Case Summary</h3><div>A 35-year-old woman was hospitalized with influenza myocarditis. Although initially hemodynamically stable, she developed cardiogenic shock 24 hours later, requiring implantation of an Impella CP device. She was transferred to the intensive care unit, where signs of device malfunction soon appeared. Fluoroscopy revealed a distorted, Z-shaped configuration of the Impella CP in the ascending aorta. Using a snare technique, the device was straightened in the descending aorta and successfully retrieved via the femoral sheath.</div></div><div><h3>Discussion</h3><div>Currently, no established guidelines exist for managing microaxial flow pump misfolding. This case demonstrates a safe and effective percutaneous retrieval approach for this complication.</div></div><div><h3>Take-Home Message</h3><div>The snare technique provides a safe and effective percutaneous method for retrieving a kinked microaxial flow pump device.</div></div>","PeriodicalId":14792,"journal":{"name":"JACC. Case reports","volume":"30 30","pages":"Article 105218"},"PeriodicalIF":0.0000,"publicationDate":"2025-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JACC. Case reports","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666084925019989","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Background

In patients with cardiogenic shock, misfolding of a microaxial flow pump within the aorta after deployment is a rare but serious complication that presents a significant therapeutic challenge.

Case Summary

A 35-year-old woman was hospitalized with influenza myocarditis. Although initially hemodynamically stable, she developed cardiogenic shock 24 hours later, requiring implantation of an Impella CP device. She was transferred to the intensive care unit, where signs of device malfunction soon appeared. Fluoroscopy revealed a distorted, Z-shaped configuration of the Impella CP in the ascending aorta. Using a snare technique, the device was straightened in the descending aorta and successfully retrieved via the femoral sheath.

Discussion

Currently, no established guidelines exist for managing microaxial flow pump misfolding. This case demonstrates a safe and effective percutaneous retrieval approach for this complication.

Take-Home Message

The snare technique provides a safe and effective percutaneous method for retrieving a kinked microaxial flow pump device.
用圈套导管矫直主动脉错折微轴流泵
在心源性休克患者中,微轴流泵部署后在主动脉内错误折叠是一种罕见但严重的并发症,给治疗带来了重大挑战。病例总结:一名35岁女性因流行性心肌炎住院。虽然最初血流动力学稳定,但24小时后发生心源性休克,需要植入Impella CP装置。她被转移到重症监护室,很快就出现了设备故障的迹象。透视显示升主动脉Impella CP呈扭曲的z形构型。使用圈套技术,该装置在降主动脉中被拉直,并成功地通过股鞘取出。目前,没有既定的指导方针存在管理微轴流泵错误折叠。本病例展示了一种安全有效的经皮手术治疗该并发症的方法。圈套技术提供了一种安全有效的经皮方法来回收扭结的微轴流泵装置。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
JACC. Case reports
JACC. Case reports Medicine-Cardiology and Cardiovascular Medicine
CiteScore
1.30
自引率
0.00%
发文量
404
审稿时长
17 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信