Quantification of Cannula Influence on Negative Pressure in Syringe Assisted Liposuction.

IF 3.4 2区 医学 Q1 SURGERY
Alexander Bartsch, Andreas Eigenberger, Oliver Felthaus, Brian W Booth, Lars Krenkel, Thomas Schratzenstaller, Lukas Prantl
{"title":"Quantification of Cannula Influence on Negative Pressure in Syringe Assisted Liposuction.","authors":"Alexander Bartsch, Andreas Eigenberger, Oliver Felthaus, Brian W Booth, Lars Krenkel, Thomas Schratzenstaller, Lukas Prantl","doi":"10.1097/PRS.0000000000012477","DOIUrl":null,"url":null,"abstract":"<p><strong>Summary: </strong>To make the results of the autologous fat transfer more satisfactory, all parameters of suction, preparation, and reapplication must be well understood. Since syringe-assisted liposuction is frequently used for small-volume procedures, we quantified the influence of the cannula on negative pressure.For these experiments, we used a digital manometer to which a syringe and one of seven different cannulas were connected. Data was collected and analyzed using LabVIEW and Matlab. The volume of our measurement setup was calculated using Boyle-Mariotte's law to ease interpretation, which revealed a volume of 2.82 ± 0.03 mL.We demonstrate that the volume of the measurement setup significantly affects the measurement results. Consequently, we accounted for this factor and calculated the theoretical values accordingly. Statistical analysis shows a significant influence on negative pressure across different cannulas, except for two pairs, which had minimal volume differences. Exemplary, for cannulas with volumes under 1.75 mL, a 2 mL syringe plunger pull was sufficient to create a - 0.5 bar (- 14.8 inHg) negative pressure.The present data indicate that the size of the suction cannula must be considered in syringe-assisted suction to ensure a specific negative pressure is not exceeded. Other research groups have demonstrated that adipocyte vitality and the amount of stromal vascular fraction are reduced by excessively strong negative pressure.</p>","PeriodicalId":20128,"journal":{"name":"Plastic and reconstructive surgery","volume":" ","pages":""},"PeriodicalIF":3.4000,"publicationDate":"2025-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Plastic and reconstructive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/PRS.0000000000012477","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 0

Abstract

Summary: To make the results of the autologous fat transfer more satisfactory, all parameters of suction, preparation, and reapplication must be well understood. Since syringe-assisted liposuction is frequently used for small-volume procedures, we quantified the influence of the cannula on negative pressure.For these experiments, we used a digital manometer to which a syringe and one of seven different cannulas were connected. Data was collected and analyzed using LabVIEW and Matlab. The volume of our measurement setup was calculated using Boyle-Mariotte's law to ease interpretation, which revealed a volume of 2.82 ± 0.03 mL.We demonstrate that the volume of the measurement setup significantly affects the measurement results. Consequently, we accounted for this factor and calculated the theoretical values accordingly. Statistical analysis shows a significant influence on negative pressure across different cannulas, except for two pairs, which had minimal volume differences. Exemplary, for cannulas with volumes under 1.75 mL, a 2 mL syringe plunger pull was sufficient to create a - 0.5 bar (- 14.8 inHg) negative pressure.The present data indicate that the size of the suction cannula must be considered in syringe-assisted suction to ensure a specific negative pressure is not exceeded. Other research groups have demonstrated that adipocyte vitality and the amount of stromal vascular fraction are reduced by excessively strong negative pressure.

注射器辅助吸脂术中插管对负压影响的定量分析。
摘要:为了使自体脂肪移植的结果更令人满意,必须充分了解吸引、准备和再应用的所有参数。由于注射器辅助吸脂经常用于小体积手术,我们量化了套管对负压的影响。在这些实验中,我们使用了一个数字压力计,注射器和七个不同的套管中的一个连接。使用LabVIEW和Matlab对数据进行采集和分析。为了便于解释,我们使用Boyle-Mariotte定律计算了测量装置的体积,结果显示测量装置的体积为2.82±0.03 ml。我们证明了测量装置的体积对测量结果有显著影响。因此,我们考虑了这一因素,并据此计算了理论值。统计分析表明,除两对套管的容积差异很小外,不同套管对负压的影响显著。例如,对于体积小于1.75 mL的套管,2 mL的注射器柱塞拉力足以产生- 0.5 bar (- 14.8 inHg)的负压。目前的数据表明,在注射器辅助抽吸中必须考虑吸管的大小,以确保不超过特定的负压。其他研究小组已经证明,脂肪细胞的活力和间质血管分数的数量会因过度强烈的负压而降低。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.00
自引率
13.90%
发文量
1436
审稿时长
1.5 months
期刊介绍: For more than 70 years Plastic and Reconstructive Surgery® has been the one consistently excellent reference for every specialist who uses plastic surgery techniques or works in conjunction with a plastic surgeon. Plastic and Reconstructive Surgery® , the official journal of the American Society of Plastic Surgeons, is a benefit of Society membership, and is also available on a subscription basis. Plastic and Reconstructive Surgery® brings subscribers up-to-the-minute reports on the latest techniques and follow-up for all areas of plastic and reconstructive surgery, including breast reconstruction, experimental studies, maxillofacial reconstruction, hand and microsurgery, burn repair, cosmetic surgery, as well as news on medicolegal issues. The cosmetic section provides expanded coverage on new procedures and techniques and offers more cosmetic-specific content than any other journal. All subscribers enjoy full access to the Journal''s website, which features broadcast quality videos of reconstructive and cosmetic procedures, podcasts, comprehensive article archives dating to 1946, and additional benefits offered by the newly-redesigned website.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信