{"title":"Low Power Long Duration Ablation for High Impedance Left Ventricular Summit PVCs","authors":"Yutsuki Tsukagoshi, Keijiro Nakamura, Shoma Kitano, Naohiko Sahara, Hidehiko Hara","doi":"10.1002/joa3.70185","DOIUrl":null,"url":null,"abstract":"<p>Low-power, long-duration (LPLD) ablation (≤ 25 W for ≥ 60 s with 30 mL/min irrigation) was applied to high-impedance (≥ 150 Ω) left ventricular summit PVC sites via the anterior interventricular vein. Power was escalated stepwise (10 W → 12 W → 15 W → 18 W) under strict safety criteria. Computer simulations demonstrated safe lesion formation at 20 W, whereas 30 W exceeded 100°C, predicting steam-pop risk. This approach achieved 66.7% acute success without complications, highlighting LPLD as a promising option for challenging high-impedance coronary venous ablation sites.\n <figure>\n <div><picture>\n <source></source></picture><p></p>\n </div>\n </figure></p>","PeriodicalId":15174,"journal":{"name":"Journal of Arrhythmia","volume":"41 5","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/joa3.70185","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Arrhythmia","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/joa3.70185","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Low-power, long-duration (LPLD) ablation (≤ 25 W for ≥ 60 s with 30 mL/min irrigation) was applied to high-impedance (≥ 150 Ω) left ventricular summit PVC sites via the anterior interventricular vein. Power was escalated stepwise (10 W → 12 W → 15 W → 18 W) under strict safety criteria. Computer simulations demonstrated safe lesion formation at 20 W, whereas 30 W exceeded 100°C, predicting steam-pop risk. This approach achieved 66.7% acute success without complications, highlighting LPLD as a promising option for challenging high-impedance coronary venous ablation sites.