Ann M West, Nicole L Schueler, Rachel A Moody, Merissa T Andersen, Jill S Dinsmore, David B Miller, Charles D Wickens, Pauline M Byom, Andrea Y Lehnertz, Kannan Ramar
{"title":"Applying High-reliability Principles to Infusion Pump Safety: A Case Study at a Multisite Health System.","authors":"Ann M West, Nicole L Schueler, Rachel A Moody, Merissa T Andersen, Jill S Dinsmore, David B Miller, Charles D Wickens, Pauline M Byom, Andrea Y Lehnertz, Kannan Ramar","doi":"10.1097/PTS.0000000000001395","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>A framework of high-reliability principles was used to identify, investigate, and mitigate infusion pump safety concerns at a large, multisite health care system. We developed a systematic approach to address challenges associated with overinfusions, underinfusions, and the inability to clear upstream occlusion alarms. We identified 112,875 upstream occlusion events for 389,604 infusion starts (failure rate, 29%) within 6 months.</p><p><strong>Methods: </strong>Five high-reliability principles were applied to infusion pump management. Preoccupation with failure emphasized reporting safety concerns and performing appropriate clinical and bench tests. Deference to expertise prompted the development of a multidisciplinary team with internal and external partners. Sensitivity to operations prompted assessment of human factors design, with simulations and analysis of medication-specific factors (e.g., viscosity; infusion rates). Reluctance to simplify prompted assessment of tubing characteristics (e.g., concentricity; wall thickness) with micro-computed tomography and process development for removing faulty equipment. Practice of resilience ensured ongoing engagement and commitment to a culture of safety and patient advocacy.</p><p><strong>Results: </strong>The multidisciplinary oversight team prompted a national recall, removal of malfunctioning pumps, and development of system-wide training and mitigation protocols. Despite ongoing pump challenges, our team optimized internal patient safety systems and processes. The cause of these malfunctions remains under investigation, but serious patient harm has been prevented.</p><p><strong>Conclusions: </strong>Key strategies for enhancing patient safety were continuous vigilance, interdisciplinary collaboration, and embracing complexity in a large health care organization. Future directions involve deeply integrating these high-reliability principles across all aspects of health care delivery to continue improving safety and quality outcomes.</p>","PeriodicalId":48901,"journal":{"name":"Journal of Patient Safety","volume":" ","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-09-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Patient Safety","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/PTS.0000000000001395","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"HEALTH CARE SCIENCES & SERVICES","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: A framework of high-reliability principles was used to identify, investigate, and mitigate infusion pump safety concerns at a large, multisite health care system. We developed a systematic approach to address challenges associated with overinfusions, underinfusions, and the inability to clear upstream occlusion alarms. We identified 112,875 upstream occlusion events for 389,604 infusion starts (failure rate, 29%) within 6 months.
Methods: Five high-reliability principles were applied to infusion pump management. Preoccupation with failure emphasized reporting safety concerns and performing appropriate clinical and bench tests. Deference to expertise prompted the development of a multidisciplinary team with internal and external partners. Sensitivity to operations prompted assessment of human factors design, with simulations and analysis of medication-specific factors (e.g., viscosity; infusion rates). Reluctance to simplify prompted assessment of tubing characteristics (e.g., concentricity; wall thickness) with micro-computed tomography and process development for removing faulty equipment. Practice of resilience ensured ongoing engagement and commitment to a culture of safety and patient advocacy.
Results: The multidisciplinary oversight team prompted a national recall, removal of malfunctioning pumps, and development of system-wide training and mitigation protocols. Despite ongoing pump challenges, our team optimized internal patient safety systems and processes. The cause of these malfunctions remains under investigation, but serious patient harm has been prevented.
Conclusions: Key strategies for enhancing patient safety were continuous vigilance, interdisciplinary collaboration, and embracing complexity in a large health care organization. Future directions involve deeply integrating these high-reliability principles across all aspects of health care delivery to continue improving safety and quality outcomes.
期刊介绍:
Journal of Patient Safety (ISSN 1549-8417; online ISSN 1549-8425) is dedicated to presenting research advances and field applications in every area of patient safety. While Journal of Patient Safety has a research emphasis, it also publishes articles describing near-miss opportunities, system modifications that are barriers to error, and the impact of regulatory changes on healthcare delivery. This mix of research and real-world findings makes Journal of Patient Safety a valuable resource across the breadth of health professions and from bench to bedside.