Toshihiko Goto, Masashi Yokoi, Kento Mori, Yoshihiro Seo
{"title":"Late Dislodgement of a Leadless Pacemaker: Potential Role of Impedance Decline as an Early Warning Sign","authors":"Toshihiko Goto, Masashi Yokoi, Kento Mori, Yoshihiro Seo","doi":"10.1002/joa3.70187","DOIUrl":null,"url":null,"abstract":"<p>Pacing impedance declined rapidly 1 week after implantation, followed by a significant increase in threshold at the 2-week follow-up, resulting in pacing failure.\n <figure>\n <div><picture>\n <source></source></picture><p></p>\n </div>\n </figure></p>","PeriodicalId":15174,"journal":{"name":"Journal of Arrhythmia","volume":"41 5","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-09-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/joa3.70187","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Arrhythmia","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/joa3.70187","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Pacing impedance declined rapidly 1 week after implantation, followed by a significant increase in threshold at the 2-week follow-up, resulting in pacing failure.