{"title":"Comparison of two dual-energy CT-based methods for proton stopping-power ratio estimation in brain cancer patients with metal implants.","authors":"Ivanka Sojat Tarp, Vicki Trier Taasti, Maria Fuglsang Jensen, Ludvig Paul Muren, Kenneth Jensen","doi":"10.2340/1651-226X.2025.43930","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong>Accurate stopping-power ratio (SPR) estimation is crucial for proton therapy planning. In brain cancer patients with metal clips, SPR accuracy may be affected by high-density materials and imaging artefacts. Dual-energy CT (DECT)-based methods have been shown to improve SPR accuracy. This study evaluated the consistency between two SPR estimation methods in brain cancer patients: (1) a Hounsfield look-up table (HLUT) for DECT-generated virtual monoenergetic images (VMIs) and (2) the DirectSPR algorithm (Siemens Healthineers). Patient/material and methods: DECT scans were acquired for 11 brain cancer patients. Two SPR maps were generated: one using a 90 keV VMI with a HLUT and the other via the DirectSPR algorithm. The VMI HLUT was adjusted in high-density regions to align with the SPR of titanium. Clinically applied proton therapy plans were recalculated on both SPR maps and dose distributions were compared using dose-volume histograms. Furthermore, a voxel-wise SPR comparison and a separate titanium implant analysis were performed.</p><p><strong>Results: </strong>Dose differences between the SPR methods were minimal for organs-at-risk. DirectSPR showed strong SPR agreement with the VMI HLUT approach for CT numbers up to 1500 HU (SPR~1.9). Beyond this, especially in regions with titanium implants, DirectSPR yielded higher SPR values than the VMI HLUT, suggesting an adjustment may also be needed for DirectSPR.</p><p><strong>Interpretation: </strong>DirectSPR was consistent with the VMI HLUT up to 1500 HU but deviated at higher CT numbers. These deviations had limited impact on dose metrics, but they should be considered when choosing beam orientations.</p>","PeriodicalId":7110,"journal":{"name":"Acta Oncologica","volume":"64 ","pages":"1168-1175"},"PeriodicalIF":2.7000,"publicationDate":"2025-09-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12421333/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Oncologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2340/1651-226X.2025.43930","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background and purpose: Accurate stopping-power ratio (SPR) estimation is crucial for proton therapy planning. In brain cancer patients with metal clips, SPR accuracy may be affected by high-density materials and imaging artefacts. Dual-energy CT (DECT)-based methods have been shown to improve SPR accuracy. This study evaluated the consistency between two SPR estimation methods in brain cancer patients: (1) a Hounsfield look-up table (HLUT) for DECT-generated virtual monoenergetic images (VMIs) and (2) the DirectSPR algorithm (Siemens Healthineers). Patient/material and methods: DECT scans were acquired for 11 brain cancer patients. Two SPR maps were generated: one using a 90 keV VMI with a HLUT and the other via the DirectSPR algorithm. The VMI HLUT was adjusted in high-density regions to align with the SPR of titanium. Clinically applied proton therapy plans were recalculated on both SPR maps and dose distributions were compared using dose-volume histograms. Furthermore, a voxel-wise SPR comparison and a separate titanium implant analysis were performed.
Results: Dose differences between the SPR methods were minimal for organs-at-risk. DirectSPR showed strong SPR agreement with the VMI HLUT approach for CT numbers up to 1500 HU (SPR~1.9). Beyond this, especially in regions with titanium implants, DirectSPR yielded higher SPR values than the VMI HLUT, suggesting an adjustment may also be needed for DirectSPR.
Interpretation: DirectSPR was consistent with the VMI HLUT up to 1500 HU but deviated at higher CT numbers. These deviations had limited impact on dose metrics, but they should be considered when choosing beam orientations.
期刊介绍:
Acta Oncologica is a journal for the clinical oncologist and accepts articles within all fields of clinical cancer research. Articles on tumour pathology, experimental oncology, radiobiology, cancer epidemiology and medical radio physics are also welcome, especially if they have a clinical aim or interest. Scientific articles on cancer nursing and psychological or social aspects of cancer are also welcomed. Extensive material may be published as Supplements, for which special conditions apply.