A Rare Presentation of Sense-B-Noise With High-Amplitude Signals in a Subcutaneous Implantable Cardioverter Defibrillator: Diagnostic Challenges and Management.
{"title":"A Rare Presentation of Sense-B-Noise With High-Amplitude Signals in a Subcutaneous Implantable Cardioverter Defibrillator: Diagnostic Challenges and Management.","authors":"Yukihiro Uehara, Nobuhiko Ueda, Akinori Wakamiya, Kohei Ishibashi, Kengo Kusano","doi":"10.1111/jce.70076","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Sense-B-noise is a rare cause of inappropriate shock (IAS) of subcutaneous implantable cardioverter defibrillator (S-ICD), which is an oversensing of the vectors using the Sense B and a factor of noncardiac signal oversensing. High-amplitude signals in Sense-B-noise have not been reported previously.</p><p><strong>Methods: </strong>We present a case of a 47-year-old man, who experienced IAS of S-ICD due to sense-B-noise followed by intense noise of primary and alternate vectors.</p><p><strong>Results: </strong>Analysis of the extracted lead revealed a slight tear in the connection of the Sense B electrode, but the non-physiological signal was not reproducible and no functional defects were identified in the device.</p><p><strong>Conclusions: </strong>High-amplitude signals after an inappropriate shock due to sense-B-noise are another presentation of Sense-B-noise.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":" ","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2025-08-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.70076","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Sense-B-noise is a rare cause of inappropriate shock (IAS) of subcutaneous implantable cardioverter defibrillator (S-ICD), which is an oversensing of the vectors using the Sense B and a factor of noncardiac signal oversensing. High-amplitude signals in Sense-B-noise have not been reported previously.
Methods: We present a case of a 47-year-old man, who experienced IAS of S-ICD due to sense-B-noise followed by intense noise of primary and alternate vectors.
Results: Analysis of the extracted lead revealed a slight tear in the connection of the Sense B electrode, but the non-physiological signal was not reproducible and no functional defects were identified in the device.
Conclusions: High-amplitude signals after an inappropriate shock due to sense-B-noise are another presentation of Sense-B-noise.
期刊介绍:
Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.