Influence of the cut-out-rescan procedure and rubber dam color on the trueness of intraoral scanners: A laboratory evaluation.

IF 1.2 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE
American journal of dentistry Pub Date : 2025-08-01
Ayşe A Şenol, Bengü Doğu Kaya, Ece Bal, Hatice B Cengiz, Bora Korkut, Pinar Yilmaz Atali
{"title":"Influence of the cut-out-rescan procedure and rubber dam color on the trueness of intraoral scanners: A laboratory evaluation.","authors":"Ayşe A Şenol, Bengü Doğu Kaya, Ece Bal, Hatice B Cengiz, Bora Korkut, Pinar Yilmaz Atali","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the laboratory effect of the cut-out-rescan procedure and the rubber dam color on the trueness of two intraoral scanners (IOS).</p><p><strong>Methods: </strong>An onlay preparation was performed on a typodont maxillary left 1st molar. The left hemi-arch of the maxilla was scanned using two IOS (Trios 3, 3Shape; n=20). These digital scans were defined as references and exported. The prepared tooth and 1 mm proximal surfaces were cut-out using the IOS' software tools. Following the rubber dam (blue-color) isolation, the cut-out area was rescanned and recorded. Reference and rescan data in .STL were uploaded to the software programs and superimposed. The deviations were obtained selectively for the teeth. Also, additional hemi-arc rescans with different rubber dam sheet colors (blue, purple, and black) were obtained and superimposed with reference scan without cut-out procedure. Trueness was evaluated using root mean square (RMS) error by Geomagic Control X and mean distance (Md) by Oracheck. Data were analyzed using ANOVA, Shapiro-Wilk, and Independent Samples (P< 0.05).</p><p><strong>Results: </strong>The Md values by the Trios 3 were significantly higher than Omnicam for the 1st premolar and 1st molar (P< 0.05). Omnicam's RMS and Md values for the 2nd molar were significantly higher than 1st premolar and 1st molar (P< 0.001). Trios's RMS values were significantly higher for the 2nd molar than the 1st molar (P< 0.001). A lower deviation in Md values was obtained for black rubber dam color compared to the blue and purple in Omnicam and the purple in Trios (P< 0.005).</p><p><strong>Clinical significance: </strong>The cut-out-rescan procedure may influence the scanning trueness depending on the IOS and location of the cut-out area. The use of the cut-out-rescan procedure will be optimized by considering the technology and software of the IOS. The black rubber dam was considered the optimal choice when using the rescan procedure.</p>","PeriodicalId":7538,"journal":{"name":"American journal of dentistry","volume":"38 4","pages":"178-184"},"PeriodicalIF":1.2000,"publicationDate":"2025-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of dentistry","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the laboratory effect of the cut-out-rescan procedure and the rubber dam color on the trueness of two intraoral scanners (IOS).

Methods: An onlay preparation was performed on a typodont maxillary left 1st molar. The left hemi-arch of the maxilla was scanned using two IOS (Trios 3, 3Shape; n=20). These digital scans were defined as references and exported. The prepared tooth and 1 mm proximal surfaces were cut-out using the IOS' software tools. Following the rubber dam (blue-color) isolation, the cut-out area was rescanned and recorded. Reference and rescan data in .STL were uploaded to the software programs and superimposed. The deviations were obtained selectively for the teeth. Also, additional hemi-arc rescans with different rubber dam sheet colors (blue, purple, and black) were obtained and superimposed with reference scan without cut-out procedure. Trueness was evaluated using root mean square (RMS) error by Geomagic Control X and mean distance (Md) by Oracheck. Data were analyzed using ANOVA, Shapiro-Wilk, and Independent Samples (P< 0.05).

Results: The Md values by the Trios 3 were significantly higher than Omnicam for the 1st premolar and 1st molar (P< 0.05). Omnicam's RMS and Md values for the 2nd molar were significantly higher than 1st premolar and 1st molar (P< 0.001). Trios's RMS values were significantly higher for the 2nd molar than the 1st molar (P< 0.001). A lower deviation in Md values was obtained for black rubber dam color compared to the blue and purple in Omnicam and the purple in Trios (P< 0.005).

Clinical significance: The cut-out-rescan procedure may influence the scanning trueness depending on the IOS and location of the cut-out area. The use of the cut-out-rescan procedure will be optimized by considering the technology and software of the IOS. The black rubber dam was considered the optimal choice when using the rescan procedure.

切开-重新扫描程序和橡胶坝颜色对口腔内扫描仪准确性的影响:实验室评估。
目的:评价两种口腔内扫描仪(IOS)切开重新扫描程序和橡胶坝色对其准确性的实验室影响。方法:对上颌左第一磨牙进行嵌套制备。使用两台IOS (Trios 3, 3Shape; n=20)对上颌骨左半弓进行扫描。这些数字扫描被定义为参考并导出。利用IOS的软件工具将准备好的牙齿和近端1mm的表面切割出来。在橡胶坝(蓝色)隔离后,重新扫描并记录切割区域。中引用并重新扫描数据。将STL上传到软件程序中并进行叠加。这些偏差是选择性地获得的牙齿。此外,还获得了不同橡胶坝板颜色(蓝色,紫色和黑色)的额外半弧扫描,并与参考扫描叠加,而无需切割程序。用Geomagic Control X软件的均方根误差(RMS)和Oracheck软件的平均距离(Md)来评估准确性。数据采用方差分析、Shapiro-Wilk和独立样本分析(P< 0.05)。结果:Trios 3对第一前磨牙和第一磨牙的Md值显著高于Omnicam (P< 0.05)。第二磨牙Omnicam的RMS和Md值显著高于第一前磨牙和第一磨牙(P< 0.001)。第2磨牙Trios的RMS值显著高于第1磨牙(P< 0.001)。与Omnicam中的蓝色和紫色以及Trios中的紫色相比,黑色橡胶坝色的Md值偏差较小(P< 0.005)。临床意义:切割-重新扫描过程可能会影响扫描的准确性,这取决于IOS和切割区域的位置。考虑到IOS的技术和软件,将优化切割-重新扫描过程的使用。重新扫描时,认为黑色橡胶坝是最佳选择。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
American journal of dentistry
American journal of dentistry 医学-牙科与口腔外科
CiteScore
2.40
自引率
7.10%
发文量
57
审稿时长
1 months
期刊介绍: The American Journal of Dentistry, published by Mosher & Linder, Inc., provides peer-reviewed scientific articles with clinical significance for the general dental practitioner.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信