Procalcitonin assay variation in an Australasian external quality assurance program.

IF 1 4区 医学 Q3 MEDICAL LABORATORY TECHNOLOGY
Alexander Richardson, Kristie Chapman, Peter Graham, Tony Badrick
{"title":"Procalcitonin assay variation in an Australasian external quality assurance program.","authors":"Alexander Richardson, Kristie Chapman, Peter Graham, Tony Badrick","doi":"10.1177/00045632251372749","DOIUrl":null,"url":null,"abstract":"<p><p>Sepsis accounts for approximately 20% of global deaths, and early diagnosis is a critical factor in intervention. In 2017, the FDA approved procalcitonin (PCT) to guide antibiotic use for patients with suspected sepsis, and there are recognized intervals for clinical interpretation. Therapeutic algorithms incorporating PCT measurement have implications for antibiotic stewardship in the age of antibiotic resistance. A 2021 review of external quality assurance programs for PCT, including the Royal College of Pathologists of Australasia Quality Assurance Programs (RCPAQAP) PCT program, highlighted variable performance between assays. We reviewed the RCPAQAP's PCT program results from 2019 to 2021 to analyse any variation in reported results. The RCPAQAP's PCT program is conducted annually and consists of a lyophilized human/bovine serum albumin base with added recombinant PCT sent to participating laboratories. Results received for the 2019, 2020 and 2021 PCT programs were analysed using two-way ANOVA with Tukey's multiple comparison test and a Student's <i>t</i>-test to investigate variation in assay performance. We found significant variation between the different assay manufacturers at all PCT concentrations analysed. Additionally, bimodal reporting was observed, where bioMerieux/Beckman Coulter/Siemens methods had significantly higher results when compared to Roche/Abbott methods. There was also a significant increase in the average coefficient of variation between 2019 and 2020/2021, coinciding with reported method changes. Finally, variable performance of the semi-quantitative PCT method at both low and high PCT concentrations was detected. These findings suggest that clinical decision cut-offs must be validated for each assay. However, the commutability of the program material has yet to be determined.</p>","PeriodicalId":8005,"journal":{"name":"Annals of Clinical Biochemistry","volume":" ","pages":"45632251372749"},"PeriodicalIF":1.0000,"publicationDate":"2025-08-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Clinical Biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/00045632251372749","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Sepsis accounts for approximately 20% of global deaths, and early diagnosis is a critical factor in intervention. In 2017, the FDA approved procalcitonin (PCT) to guide antibiotic use for patients with suspected sepsis, and there are recognized intervals for clinical interpretation. Therapeutic algorithms incorporating PCT measurement have implications for antibiotic stewardship in the age of antibiotic resistance. A 2021 review of external quality assurance programs for PCT, including the Royal College of Pathologists of Australasia Quality Assurance Programs (RCPAQAP) PCT program, highlighted variable performance between assays. We reviewed the RCPAQAP's PCT program results from 2019 to 2021 to analyse any variation in reported results. The RCPAQAP's PCT program is conducted annually and consists of a lyophilized human/bovine serum albumin base with added recombinant PCT sent to participating laboratories. Results received for the 2019, 2020 and 2021 PCT programs were analysed using two-way ANOVA with Tukey's multiple comparison test and a Student's t-test to investigate variation in assay performance. We found significant variation between the different assay manufacturers at all PCT concentrations analysed. Additionally, bimodal reporting was observed, where bioMerieux/Beckman Coulter/Siemens methods had significantly higher results when compared to Roche/Abbott methods. There was also a significant increase in the average coefficient of variation between 2019 and 2020/2021, coinciding with reported method changes. Finally, variable performance of the semi-quantitative PCT method at both low and high PCT concentrations was detected. These findings suggest that clinical decision cut-offs must be validated for each assay. However, the commutability of the program material has yet to be determined.

降钙素原测定在澳大利亚外部质量保证计划中的变化。
脓毒症约占全球死亡人数的20%,早期诊断是干预的关键因素。2017年,FDA批准降钙素原(PCT)用于指导疑似脓毒症患者的抗生素使用,并且有公认的临床解释间隔。结合PCT测量的治疗算法对抗生素耐药性时代的抗生素管理具有重要意义。2021年对PCT外部质量保证项目的审查,包括澳大利亚皇家病理学家学院质量保证项目(RCPAQAP) PCT项目,强调了不同检测方法之间的不同表现。我们回顾了RCPAQAP 2019-2021年的PCT项目结果,以分析报告结果的任何变化。RCPAQAP的PCT项目每年进行一次,由冻干的人/牛血清白蛋白碱基和添加的重组PCT组成,并送到参与实验室。对2019年、2020年和2021年PCT项目收到的结果进行双向方差分析,采用Tukey多重比较检验和学生t检验来调查分析性能的变化。我们发现在所有PCT浓度分析中,不同检测制造商之间存在显著差异。此外,还观察到双峰报告,其中生物梅里埃/贝克曼-库尔特/西门子方法的结果明显高于罗氏/雅培方法。2019年至2020/2021年期间的平均变异系数也显着增加,与报告的方法变化相一致。最后,检测了半定量PCT方法在低浓度和高浓度PCT下的可变性能。这些发现表明,临床决策的切断必须验证每个分析。然而,程序材料的可交换性还有待确定。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of Clinical Biochemistry
Annals of Clinical Biochemistry Biochemistry, Genetics and Molecular Biology-Clinical Biochemistry
CiteScore
5.20
自引率
4.50%
发文量
61
期刊介绍: Annals of Clinical Biochemistry is the fully peer reviewed international journal of the Association for Clinical Biochemistry and Laboratory Medicine. Annals of Clinical Biochemistry accepts papers that contribute to knowledge in all fields of laboratory medicine, especially those pertaining to the understanding, diagnosis and treatment of human disease. It publishes papers on clinical biochemistry, clinical audit, metabolic medicine, immunology, genetics, biotechnology, haematology, microbiology, computing and management where they have both biochemical and clinical relevance. Papers describing evaluation or implementation of commercial reagent kits or the performance of new analysers require substantial original information. Unless of exceptional interest and novelty, studies dealing with the redox status in various diseases are not generally considered within the journal''s scope. Studies documenting the association of single nucleotide polymorphisms (SNPs) with particular phenotypes will not normally be considered, given the greater strength of genome wide association studies (GWAS). Research undertaken in non-human animals will not be considered for publication in the Annals. Annals of Clinical Biochemistry is also the official journal of NVKC (de Nederlandse Vereniging voor Klinische Chemie) and JSCC (Japan Society of Clinical Chemistry).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信