Enabling in vivo comparisons of different four-dimensional magnetic resonance imaging sequences for radiotherapy guidance using visual biofeedback

IF 3.3 Q2 ONCOLOGY
Katrinus Keijnemans , Tim Schakel , Bastien Lecoeur , Pim T.S. Borman , William A. Hall , Bas W. Raaymakers , Andreas Wetscherek , Eric S. Paulson , Martin F. Fast
{"title":"Enabling in vivo comparisons of different four-dimensional magnetic resonance imaging sequences for radiotherapy guidance using visual biofeedback","authors":"Katrinus Keijnemans ,&nbsp;Tim Schakel ,&nbsp;Bastien Lecoeur ,&nbsp;Pim T.S. Borman ,&nbsp;William A. Hall ,&nbsp;Bas W. Raaymakers ,&nbsp;Andreas Wetscherek ,&nbsp;Eric S. Paulson ,&nbsp;Martin F. Fast","doi":"10.1016/j.phro.2025.100815","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and Purpose:</h3><div>Managing respiratory motion is essential for effective radiotherapy in the abdominothoracic regions. Respiratory-correlated four-dimensional magnetic resonance imaging (4D-MRI) can provide accurate motion estimation to help define treatment volumes for adaptive radiotherapy. However, validating and comparing 4D-MRI sequences in vivo is challenging due to the presence of breathing variability. This study combines visual biofeedback (VBF) with 4D-MRI sequences to facilitate in vivo comparisons.</div></div><div><h3>Materials and Methods:</h3><div>Fourteen healthy volunteers and one patient were scanned on a 1.5 T Unity MR-linear accelerator (Elekta AB, Stockholm, Sweden) at two institutions. A radial stack-of-stars (SoS), a simultaneous multi-slice (SMS), and a Cartesian acquisition with spiral ordering (CASPR) 4D-MRI sequence were acquired. These acquisitions were performed without and with VBF based on an interleaved one-dimensional respiratory navigator (1D-RNAV) acquisition. Breathing variability across sequences was quantified using 1D-RNAV-derived breathing waveforms. Reconstructed 4D-MRI data were used to extract the motion amplitude, which was compared intra-volunteer across sequences and to the amplitudes of the breathing waveforms.</div></div><div><h3>Results:</h3><div>Breathing variability across sequences decreased by 37% (amplitude, <em>p</em> <span><math><mo>=</mo></math></span> 0.039) and 64% (period, <em>p</em> <span><math><mo>&lt;</mo></math></span> 0.003), and the median intra-volunteer 4D-MRI-derived motion amplitude agreement improved from 3.5 mm to 1.8 mm (<em>p</em> <span><math><mo>=</mo></math></span> 0.064) across sequences due to VBF guidance. Four-dimensional MRI-derived amplitudes were smaller than breathing waveform amplitudes, with median differences of -31% (SoS), -17% (SMS), and -9% (CASPR). The average breathing waveform amplitude was 8% larger than instructed.</div></div><div><h3>Conclusions:</h3><div>This methodology enables in vivo comparisons of 4D-MRI sequences for adaptive radiotherapy, with guidance improving anatomical consistency and ensuring more reliable comparisons.</div></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"35 ","pages":"Article 100815"},"PeriodicalIF":3.3000,"publicationDate":"2025-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631625001204","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background and Purpose:

Managing respiratory motion is essential for effective radiotherapy in the abdominothoracic regions. Respiratory-correlated four-dimensional magnetic resonance imaging (4D-MRI) can provide accurate motion estimation to help define treatment volumes for adaptive radiotherapy. However, validating and comparing 4D-MRI sequences in vivo is challenging due to the presence of breathing variability. This study combines visual biofeedback (VBF) with 4D-MRI sequences to facilitate in vivo comparisons.

Materials and Methods:

Fourteen healthy volunteers and one patient were scanned on a 1.5 T Unity MR-linear accelerator (Elekta AB, Stockholm, Sweden) at two institutions. A radial stack-of-stars (SoS), a simultaneous multi-slice (SMS), and a Cartesian acquisition with spiral ordering (CASPR) 4D-MRI sequence were acquired. These acquisitions were performed without and with VBF based on an interleaved one-dimensional respiratory navigator (1D-RNAV) acquisition. Breathing variability across sequences was quantified using 1D-RNAV-derived breathing waveforms. Reconstructed 4D-MRI data were used to extract the motion amplitude, which was compared intra-volunteer across sequences and to the amplitudes of the breathing waveforms.

Results:

Breathing variability across sequences decreased by 37% (amplitude, p = 0.039) and 64% (period, p < 0.003), and the median intra-volunteer 4D-MRI-derived motion amplitude agreement improved from 3.5 mm to 1.8 mm (p = 0.064) across sequences due to VBF guidance. Four-dimensional MRI-derived amplitudes were smaller than breathing waveform amplitudes, with median differences of -31% (SoS), -17% (SMS), and -9% (CASPR). The average breathing waveform amplitude was 8% larger than instructed.

Conclusions:

This methodology enables in vivo comparisons of 4D-MRI sequences for adaptive radiotherapy, with guidance improving anatomical consistency and ensuring more reliable comparisons.
利用视觉生物反馈对不同的四维磁共振成像序列进行放射治疗指导的体内比较
背景与目的:控制呼吸运动是腹胸区有效放疗的必要条件。呼吸相关四维磁共振成像(4D-MRI)可以提供准确的运动估计,以帮助确定适应性放疗的治疗量。然而,由于存在呼吸变异性,在体内验证和比较4D-MRI序列具有挑战性。本研究将视觉生物反馈(VBF)与4D-MRI序列相结合,以促进体内比较。材料和方法:在两个机构的1.5 T Unity磁共振直线加速器(Elekta AB, Stockholm, Sweden)上对14名健康志愿者和1名患者进行扫描。获得了径向星团(SoS)、同时多层(SMS)和螺旋有序笛卡尔采集(CASPR) 4D-MRI序列。这些采集是在无VBF和有VBF的情况下进行的,基于交错一维呼吸导航(1D-RNAV)采集。使用1d - rnav衍生的呼吸波形来量化不同序列的呼吸变异性。重建的4D-MRI数据用于提取运动幅度,并将其与志愿者内部序列和呼吸波形的幅度进行比较。结果:呼吸变异性在不同序列间分别下降了37%(幅度,p = 0.039)和64%(周期,p <;0.003),在VBF引导下,志愿者内部4d - mri得出的运动幅度一致性中位数从3.5 mm提高到1.8 mm (p = 0.064)。四维mri衍生振幅小于呼吸波形振幅,中位数差异为-31% (SoS), -17% (SMS)和-9% (CASPR)。平均呼吸波形幅度比指示大8%。结论:该方法能够在体内比较适应放疗的4D-MRI序列,并在指导下提高解剖一致性,确保更可靠的比较。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physics and Imaging in Radiation Oncology
Physics and Imaging in Radiation Oncology Physics and Astronomy-Radiation
CiteScore
5.30
自引率
18.90%
发文量
93
审稿时长
6 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信