Early Deterioration of the Right Ventricular Pacing Threshold Predicts the Increase in His-Bundle Pacing Threshold During the Chronic Phase: A Single-Center Retrospective Study
{"title":"Early Deterioration of the Right Ventricular Pacing Threshold Predicts the Increase in His-Bundle Pacing Threshold During the Chronic Phase: A Single-Center Retrospective Study","authors":"Kenji Baba, Shogo Sakamoto, Moritoshi Irishio, Tomotaka Yoshiyama, Yusuke Hayashi, Toru Kataoka, Daiju Fukuda","doi":"10.1002/joa3.70144","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>His-bundle pacing (HBP) facilitates physiological ventricular activation. However, concerns about long-term threshold deterioration persist. The predictors of chronic threshold elevation are not yet well established.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Between February 2018 and December 2021, HBP was attempted in 95 patients undergoing pacemaker implantation. Strict success criteria (threshold ≤ 2.0 V/0.4 ms) were applied, and 47 patients with successful HBP were monitored for over 3 years. We assessed pacing thresholds for both the His-bundle and right ventricle (RV) at implantation, 1 week, 1 month, and annually thereafter. Lead shape was evaluated using chest radiography. Patients were categorized into two groups based on their His-bundle pacing threshold: stable (≤ 3.5 V/0.4 ms) and deteriorated (> 3.5 V/0.4 ms).</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Deterioration in His-bundle pacing thresholds was associated with increased RV pacing thresholds. Deterioration in RV pacing occurred earlier, with significant differences observed at 1 week post-implantation (median RV: 1.87 vs. 3.25 V/0.4 ms, <i>p</i> = 0.032; His-bundle: 1.0 vs. 1.25 V/0.4 ms, <i>p</i> = 0.212). Multivariate analysis identified an RV threshold ≥ 3.0 V/0.4 ms at 1 week (OR 10.7, <i>p</i> = 0.036) and lead bending on chest radiography (OR 12.8, <i>p</i> = 0.018) as independent predictors of chronic His-bundle pacing threshold deterioration.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>An elevated RV pacing threshold at 1 week post-implantation and lead flexion at implantation may serve as early indicators of long-term deterioration in His-bundle pacing thresholds. When the RV pacing threshold increase is detected, it is important to closely monitor the patient and frequently adjust the output settings to prevent pacing failure.</p>\n </section>\n </div>","PeriodicalId":15174,"journal":{"name":"Journal of Arrhythmia","volume":"41 4","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-07-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/joa3.70144","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Arrhythmia","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/joa3.70144","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
His-bundle pacing (HBP) facilitates physiological ventricular activation. However, concerns about long-term threshold deterioration persist. The predictors of chronic threshold elevation are not yet well established.
Methods
Between February 2018 and December 2021, HBP was attempted in 95 patients undergoing pacemaker implantation. Strict success criteria (threshold ≤ 2.0 V/0.4 ms) were applied, and 47 patients with successful HBP were monitored for over 3 years. We assessed pacing thresholds for both the His-bundle and right ventricle (RV) at implantation, 1 week, 1 month, and annually thereafter. Lead shape was evaluated using chest radiography. Patients were categorized into two groups based on their His-bundle pacing threshold: stable (≤ 3.5 V/0.4 ms) and deteriorated (> 3.5 V/0.4 ms).
Results
Deterioration in His-bundle pacing thresholds was associated with increased RV pacing thresholds. Deterioration in RV pacing occurred earlier, with significant differences observed at 1 week post-implantation (median RV: 1.87 vs. 3.25 V/0.4 ms, p = 0.032; His-bundle: 1.0 vs. 1.25 V/0.4 ms, p = 0.212). Multivariate analysis identified an RV threshold ≥ 3.0 V/0.4 ms at 1 week (OR 10.7, p = 0.036) and lead bending on chest radiography (OR 12.8, p = 0.018) as independent predictors of chronic His-bundle pacing threshold deterioration.
Conclusion
An elevated RV pacing threshold at 1 week post-implantation and lead flexion at implantation may serve as early indicators of long-term deterioration in His-bundle pacing thresholds. When the RV pacing threshold increase is detected, it is important to closely monitor the patient and frequently adjust the output settings to prevent pacing failure.
他束起搏(HBP)促进生理心室激活。然而,对长期阈值恶化的担忧仍然存在。慢性阈值升高的预测因素尚未很好地建立。方法2018年2月至2021年12月,对95例接受心脏起搏器植入的患者进行降压治疗。采用严格的成功标准(阈值≤2.0 V/0.4 ms),对47例HBP成功患者进行了3年以上的监测。我们评估了他束和右心室(RV)在植入时、1周、1个月和之后每年的起搏阈值。使用胸片评估铅形态。根据患者的his束起搏阈值分为稳定组(≤3.5 V/0.4 ms)和恶化组(≤3.5 V/0.4 ms)。结果his束起搏阈值恶化与RV起搏阈值升高相关。RV起搏恶化发生较早,在植入后1周观察到显著差异(中位RV: 1.87 vs. 3.25 V/0.4 ms, p = 0.032;his束:1.0 vs. 1.25 V/0.4 ms, p = 0.212)。多因素分析发现,1周时RV阈值≥3.0 V/0.4 ms (OR 10.7, p = 0.036)和胸片导联弯曲(OR 12.8, p = 0.018)是慢性his束起搏阈值恶化的独立预测因素。结论植入1周后RV起搏阈值升高和植入时导联屈曲可作为his束起搏阈值长期恶化的早期指标。当检测到心室起搏阈值升高时,应密切监测患者并经常调整输出设置以防止起搏失败。