{"title":"Accuracy and reproducibility of a single-pose image-to-robot registration method for mobile C-arm cone beam CT guided histotripsy.","authors":"Grace M Minesinger, Paul F Laeseke, Katrina L Falk, Claire E Hennen, Michael A Speidel, Martin G Wagner","doi":"10.1002/acm2.70132","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Histotripsy is a focal tumor therapy that utilizes focused ultrasound (US) to mechanically destroy tissue. To overcome visualization limitations of diagnostic US-guidance, C-arm cone beam CT (CBCT)-guided histotripsy is being developed, for which a mobile C-arm could increase accessibility. CBCT-guided histotripsy uses a phantom with a helical fiducial pattern to determine the CBCT-to-histotripsy robot coordinate transformation. This study presents an image-to-robot registration method requiring only one phantom CBCT, evaluated for accuracy and reproducibility using a mobile C-arm.</p><p><strong>Methods: </strong>The phantom is attached to a robotic arm (replacing the histotripsy transducer) and positioned at isocenter. A CBCT is acquired and image-to-robot registration performed by registering a digital model of the phantom to observed fiducials in CBCT coordinates. Registration was performed by one user (n = 8/day, 2 days) and by 12 different users (n = 4/day, 3 days) with fiducial registration errors (FREs) calculated. After each registration, the transducer was reattached to the robot and a treatment delivered in a multi-layered, agar-based phantom. Directional and target registration errors (TREs) were calculated as directional and Euclidean distances between planned and observed treatments. Directional error inter-day differences for the single-user experiment were evaluated for significance using 2-tailed unpaired Student t-tests. The effect of user variability on variability of FRE and directional error was evaluated for significance using 2-tailed F-tests.</p><p><strong>Results: </strong>Registrations yielded FRE of 0.12 ± 0.03 mm and TRE of 1.51 ± 0.83 mm. Targeting error significantly increased along the transducer's short axis between days (0.88 ± 0.60 vs 1.43 ± 0.18 mm, p = 0.025) for the single user, with a similar trend for the multi-user experiment (1.45 ± 0.79, 2.70 ± 0.19, and 2.83 ± 0.40 mm). User variability, and thus robot pose variability, did not significantly affect variability of FRE or directional error.</p><p><strong>Conclusions: </strong>Mobile C-arm CBCT-guided histotripsy showed high accuracy with minimal yet nonnegligible TREs, consistent within but not between days, demonstrating that errors can be measured and accounted for, ideally near treatment day to maximize accuracy.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e70132"},"PeriodicalIF":2.0000,"publicationDate":"2025-05-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.70132","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Histotripsy is a focal tumor therapy that utilizes focused ultrasound (US) to mechanically destroy tissue. To overcome visualization limitations of diagnostic US-guidance, C-arm cone beam CT (CBCT)-guided histotripsy is being developed, for which a mobile C-arm could increase accessibility. CBCT-guided histotripsy uses a phantom with a helical fiducial pattern to determine the CBCT-to-histotripsy robot coordinate transformation. This study presents an image-to-robot registration method requiring only one phantom CBCT, evaluated for accuracy and reproducibility using a mobile C-arm.
Methods: The phantom is attached to a robotic arm (replacing the histotripsy transducer) and positioned at isocenter. A CBCT is acquired and image-to-robot registration performed by registering a digital model of the phantom to observed fiducials in CBCT coordinates. Registration was performed by one user (n = 8/day, 2 days) and by 12 different users (n = 4/day, 3 days) with fiducial registration errors (FREs) calculated. After each registration, the transducer was reattached to the robot and a treatment delivered in a multi-layered, agar-based phantom. Directional and target registration errors (TREs) were calculated as directional and Euclidean distances between planned and observed treatments. Directional error inter-day differences for the single-user experiment were evaluated for significance using 2-tailed unpaired Student t-tests. The effect of user variability on variability of FRE and directional error was evaluated for significance using 2-tailed F-tests.
Results: Registrations yielded FRE of 0.12 ± 0.03 mm and TRE of 1.51 ± 0.83 mm. Targeting error significantly increased along the transducer's short axis between days (0.88 ± 0.60 vs 1.43 ± 0.18 mm, p = 0.025) for the single user, with a similar trend for the multi-user experiment (1.45 ± 0.79, 2.70 ± 0.19, and 2.83 ± 0.40 mm). User variability, and thus robot pose variability, did not significantly affect variability of FRE or directional error.
Conclusions: Mobile C-arm CBCT-guided histotripsy showed high accuracy with minimal yet nonnegligible TREs, consistent within but not between days, demonstrating that errors can be measured and accounted for, ideally near treatment day to maximize accuracy.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic