Lingling Liu, Zhenle Fei, Jie Li, Jiong Shu, Jingyuan Shao, Jianguang Zhang, Xiangli Cui, Hongzhi Wang
{"title":"Impact of respiratory motion on dose distribution in SIB-SBRT for lung cancer.","authors":"Lingling Liu, Zhenle Fei, Jie Li, Jiong Shu, Jingyuan Shao, Jianguang Zhang, Xiangli Cui, Hongzhi Wang","doi":"10.1002/acm2.70136","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Respiratory motion is a major source of dose uncertainty in lung cancer radiotherapy. The dose distribution of simultaneous integrated boost-stereotactic body radiotherapy (SIB-SBRT) is inhomogeneous and is significantly impacted by respiratory motion for lung cancer. The effect of respiratory motion on SIB-SBRT was investigated with a four-dimensional (4D) dose calculation method.</p><p><strong>Methods: </strong>Nineteen previously treated lung cancer patients were selected for this planning study. All patients underwent four-dimensional CT (4D-CT) scanning, and volumetric modulated arc therapy (VMAT) treatments were planned with internal target volume (ITV) and planning target volume (PTV). Dose distributions (3D-plan) were calculated on the average reconstruction of the 4D-CT. 4D dose distributions (4D-plan) were calculated to evaluate respiratory motion effects. These calculations were performed on the CT images of related respiratory phase with a respiration-correlated assignment of the 3D plan's monitor units to the respiratory phases of the 4D-CT. Subsequently, the accumulative 4D dose based on deformable registrations of the CT series was generated and compared to the 3D dose distribution. Dosimetric deviations in targets and organs at risk (OARs) were analyzed with dosimetric parameters, and correlations between dose deviations (ΔV<sub>100</sub> (ITV, PTV)) and patient characteristics (left-right, SI, anterior-posterior, S, L, Volume (ITV, PTV)) were explored.</p><p><strong>Results: </strong>With deformable registrations, the median values of relative differences between 3D-plan and 4D-plan 0 were found to be from -6.6% to 12.1% for all targets dosimetric parameters, and from -4.2% to 1.4% for OAR parameters. It was also shown that PTV coverage dropped more significantly than that of ITV with respiratory motion. Strong correlations were observed between the ΔV<sub>100</sub> (ITV, PTV) and patient characteristic (SI, S, L).</p><p><strong>Conclusion: </strong>Respiratory motion effects during SIB-SBRT treatment resulted in non-negligible dose variability. Furthermore, with the correlation relationship and respiratory motion parameters, the dose coverage reduction of targets could be predicted.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e70136"},"PeriodicalIF":2.0000,"publicationDate":"2025-06-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.70136","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Respiratory motion is a major source of dose uncertainty in lung cancer radiotherapy. The dose distribution of simultaneous integrated boost-stereotactic body radiotherapy (SIB-SBRT) is inhomogeneous and is significantly impacted by respiratory motion for lung cancer. The effect of respiratory motion on SIB-SBRT was investigated with a four-dimensional (4D) dose calculation method.
Methods: Nineteen previously treated lung cancer patients were selected for this planning study. All patients underwent four-dimensional CT (4D-CT) scanning, and volumetric modulated arc therapy (VMAT) treatments were planned with internal target volume (ITV) and planning target volume (PTV). Dose distributions (3D-plan) were calculated on the average reconstruction of the 4D-CT. 4D dose distributions (4D-plan) were calculated to evaluate respiratory motion effects. These calculations were performed on the CT images of related respiratory phase with a respiration-correlated assignment of the 3D plan's monitor units to the respiratory phases of the 4D-CT. Subsequently, the accumulative 4D dose based on deformable registrations of the CT series was generated and compared to the 3D dose distribution. Dosimetric deviations in targets and organs at risk (OARs) were analyzed with dosimetric parameters, and correlations between dose deviations (ΔV100 (ITV, PTV)) and patient characteristics (left-right, SI, anterior-posterior, S, L, Volume (ITV, PTV)) were explored.
Results: With deformable registrations, the median values of relative differences between 3D-plan and 4D-plan 0 were found to be from -6.6% to 12.1% for all targets dosimetric parameters, and from -4.2% to 1.4% for OAR parameters. It was also shown that PTV coverage dropped more significantly than that of ITV with respiratory motion. Strong correlations were observed between the ΔV100 (ITV, PTV) and patient characteristic (SI, S, L).
Conclusion: Respiratory motion effects during SIB-SBRT treatment resulted in non-negligible dose variability. Furthermore, with the correlation relationship and respiratory motion parameters, the dose coverage reduction of targets could be predicted.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic