Felice Lorusso, Sergio Alexandre Gehrke, Sergio Rexhep Tari, Antonio Scarano
{"title":"Fracture Strength of Conometric Joint Implants Vs. Internal Hexagon Abutment Joint Design: An In Vitro Study.","authors":"Felice Lorusso, Sergio Alexandre Gehrke, Sergio Rexhep Tari, Antonio Scarano","doi":"10.1563/aaid-joi-D-25-00001","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The dental implant is an effective long-term rehabilitation procedure for oral edentulism and supports functional masticatory loading forces. The implant prosthetic joint is considered a key factor for interface stability due to its related biological and biomechanical implications-the present investigation aimed to evaluate the fracture strength of two different implant prosthetic joints.</p><p><strong>Material and methods: </strong>This investigation tested 10 implants for each group: econometric implant joint (Group I) and internal hexagon implant (Group II). The implant abutment joint was coupled using a calibrated torque meter. The samples were assessed using a loading fracture test and radiographically evaluated to investigate the interface changes and deformations.</p><p><strong>Results: </strong>The means and standard deviations of the group I and group II maximum force (N) were 553±51N and 432±43N. The Young elastic modulus of group I and group II implants were 183.97±11.71 GPa and 143.72±15.93 GPa.</p><p><strong>Conclusions: </strong>The conometric joint was reported to have a higher strength than the regular internal hexagon implant connection. The study findings could have clinical implications for implant durability and peri-implant tissue stability in favor of the conical joint design.</p>","PeriodicalId":519890,"journal":{"name":"The Journal of oral implantology","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-05-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of oral implantology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1563/aaid-joi-D-25-00001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: The dental implant is an effective long-term rehabilitation procedure for oral edentulism and supports functional masticatory loading forces. The implant prosthetic joint is considered a key factor for interface stability due to its related biological and biomechanical implications-the present investigation aimed to evaluate the fracture strength of two different implant prosthetic joints.
Material and methods: This investigation tested 10 implants for each group: econometric implant joint (Group I) and internal hexagon implant (Group II). The implant abutment joint was coupled using a calibrated torque meter. The samples were assessed using a loading fracture test and radiographically evaluated to investigate the interface changes and deformations.
Results: The means and standard deviations of the group I and group II maximum force (N) were 553±51N and 432±43N. The Young elastic modulus of group I and group II implants were 183.97±11.71 GPa and 143.72±15.93 GPa.
Conclusions: The conometric joint was reported to have a higher strength than the regular internal hexagon implant connection. The study findings could have clinical implications for implant durability and peri-implant tissue stability in favor of the conical joint design.