Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates.

Neonatology Pub Date : 2025-05-19 DOI:10.1159/000546187
R Brandon Hunter, Michele S Saruwatari, Nathan Daniel Bliss, Sarah Kim, Rodica Monica Turcu, Brooke Ann Krbec, Kamran Yusuf, Anoop Rao, Xina Quan, Lamia Soghier
{"title":"Accuracy and Safety of a Continuous Noninvasive Blood Pressure Monitor in Neonates.","authors":"R Brandon Hunter, Michele S Saruwatari, Nathan Daniel Bliss, Sarah Kim, Rodica Monica Turcu, Brooke Ann Krbec, Kamran Yusuf, Anoop Rao, Xina Quan, Lamia Soghier","doi":"10.1159/000546187","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Accurate and continuous blood pressure (BP) monitoring in neonates is crucial in the Intensive Care Unit. Invasive arterial lines, oscillometric cuffs, and current noninvasive continuous BP monitoring devices have significant limitations. The Boppli® device is a novel, continuous, noninvasive BP device that requires no calibration, designed for neonates. Aim &amp; Methods: This prospective, multicenter study evaluated the performance, usability, and safety of the Boppli device in neonates <5 kg. We compared Mean Arterial Pressure (MAP), Systolic Blood Pressure (SBP) and Diastolic Blood Pressure (DBP) measurements from the Boppli with IAL reference values by calculating average values of Mean Average Error (MAE) and Standard Deviation (SD) for each patient, then averaging those means. Safety and usability were evaluated by analysis of adverse events and survey data, respectively.</p><p><strong>Results: </strong>The Boppli device demonstrated good performance, meeting the FDA requirements of MAE and SD of the entire cohort: MAE (SD) 0.7 (5.3) mmHg for MAP, -0.8 (7.7) mmHg for SBP, and 1.4 (4.7) mmHg for DBP. Patients with elevated MAPs, Asian ethnicity, and lower extremity IALs were the subgroups with MAE >±5 mmHg. Various subgroups had standard deviations >8 mmHg attributed to low sample sizes. The device received high usability scores from clinicians and parents. No serious adverse events were reported.</p><p><strong>Conclusions: </strong>The Boppli device is a promising alternative for continuous noninvasive BP monitoring in neonates, offering good accuracy and usability. The device, which received 510(k) clearance in September of 2023, was well-received by clinicians and parents, with a low risk profile.</p>","PeriodicalId":94152,"journal":{"name":"Neonatology","volume":" ","pages":"1-19"},"PeriodicalIF":0.0000,"publicationDate":"2025-05-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Neonatology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1159/000546187","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Accurate and continuous blood pressure (BP) monitoring in neonates is crucial in the Intensive Care Unit. Invasive arterial lines, oscillometric cuffs, and current noninvasive continuous BP monitoring devices have significant limitations. The Boppli® device is a novel, continuous, noninvasive BP device that requires no calibration, designed for neonates. Aim & Methods: This prospective, multicenter study evaluated the performance, usability, and safety of the Boppli device in neonates <5 kg. We compared Mean Arterial Pressure (MAP), Systolic Blood Pressure (SBP) and Diastolic Blood Pressure (DBP) measurements from the Boppli with IAL reference values by calculating average values of Mean Average Error (MAE) and Standard Deviation (SD) for each patient, then averaging those means. Safety and usability were evaluated by analysis of adverse events and survey data, respectively.

Results: The Boppli device demonstrated good performance, meeting the FDA requirements of MAE and SD of the entire cohort: MAE (SD) 0.7 (5.3) mmHg for MAP, -0.8 (7.7) mmHg for SBP, and 1.4 (4.7) mmHg for DBP. Patients with elevated MAPs, Asian ethnicity, and lower extremity IALs were the subgroups with MAE >±5 mmHg. Various subgroups had standard deviations >8 mmHg attributed to low sample sizes. The device received high usability scores from clinicians and parents. No serious adverse events were reported.

Conclusions: The Boppli device is a promising alternative for continuous noninvasive BP monitoring in neonates, offering good accuracy and usability. The device, which received 510(k) clearance in September of 2023, was well-received by clinicians and parents, with a low risk profile.

新生儿连续无创血压监测仪的准确性和安全性。
背景:准确和持续的新生儿血压监测在重症监护病房是至关重要的。有创动脉线、振荡袖带和目前的无创连续血压监测设备有明显的局限性。Boppli®装置是一种新颖、连续、无创的BP装置,无需校准,专为新生儿设计。的目标,方法:这项前瞻性、多中心研究评估了Boppli装置在新生儿中的性能、可用性和安全性。结果:Boppli装置表现出良好的性能,满足FDA对整个队列的MAE和SD的要求:MAP MAE (SD)为0.7 (5.3)mmHg,收缩压MAE (SD)为-0.8 (7.7)mmHg,舒张压MAE (SD)为1.4 (4.7)mmHg。map升高的患者、亚洲人种和下肢患者是MAE bb0±5 mmHg的亚组。由于样本量小,各个亚组的标准差为80毫米汞柱。该设备从临床医生和家长那里获得了很高的可用性分数。无严重不良事件报告。结论:Boppli装置具有良好的准确性和可用性,是一种有前景的无创新生儿血压持续监测的替代方案。该设备于2023年9月获得510(k)批准,受到临床医生和家长的好评,风险低。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信