[National Survey on System Settings of Angiography for Endovascular Treatment for Lower Extremity Arterial Disease].

Yuji Sakai, Tomomichi Ishibashi, Yoshinori Takao, Toru Ishibashi, Yudai Fujimoto, Mitsuharu Osawa, Akihiro Shinobu, Kenji Hasegawa, Hajime Sakamoto
{"title":"[National Survey on System Settings of Angiography for Endovascular Treatment for Lower Extremity Arterial Disease].","authors":"Yuji Sakai, Tomomichi Ishibashi, Yoshinori Takao, Toru Ishibashi, Yudai Fujimoto, Mitsuharu Osawa, Akihiro Shinobu, Kenji Hasegawa, Hajime Sakamoto","doi":"10.6009/jjrt.25-1530","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The objective of this study was to determine the standard settings of the angiography system during EVT in Japan.</p><p><strong>Methods: </strong>Survey requests were mailed to 493 facilities across Japan. The survey items included in the survey were system display dose setting of fluoroscopy dose (FL dose)/digital angiography dose (DA dose)/digital subtraction angiography dose (DSA dose), detector size, fluoroscopy pulse rate, DA/DSA frame rate, and standard additional filter settings. The number of valid responses for each item was system display dose setting: 187 systems (128 institutions), detector size: 187 systems (128 institutions), fluoroscopy pulse rate and DA/DSA frame rate: 185 systems (127 institutions), standard additional filter settings: 182 systems (125 institutions).</p><p><strong>Results: </strong>Median (interquartile range) of each system display dose setting was FL: 5.9 (4.2-7.8) mGy/min, DA: 0.10 (0.06-0.16) mGy/frame, DSA: 0.82 (0.64-1.40) mGy/frame. The detector size for dose measurements was 103 systems (55%) with a diagonal of 35 cm or less and 84 systems (45%) with a diagonal of more than 35 cm. The most used FL pulse rate was 7.5 pulse/s (132 systems: 71%), DA frame rate was 7.5 frame/s (72 systems: 39%), and DSA frame rate was 3 frame/s (97 systems: 52%). More than 90% of the institutions used some kind of additional filter for imaging.</p><p><strong>Conclusion: </strong>This survey showed the standard setting of the angiography system during EVT in Japan.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":"81 6","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.25-1530","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The objective of this study was to determine the standard settings of the angiography system during EVT in Japan.

Methods: Survey requests were mailed to 493 facilities across Japan. The survey items included in the survey were system display dose setting of fluoroscopy dose (FL dose)/digital angiography dose (DA dose)/digital subtraction angiography dose (DSA dose), detector size, fluoroscopy pulse rate, DA/DSA frame rate, and standard additional filter settings. The number of valid responses for each item was system display dose setting: 187 systems (128 institutions), detector size: 187 systems (128 institutions), fluoroscopy pulse rate and DA/DSA frame rate: 185 systems (127 institutions), standard additional filter settings: 182 systems (125 institutions).

Results: Median (interquartile range) of each system display dose setting was FL: 5.9 (4.2-7.8) mGy/min, DA: 0.10 (0.06-0.16) mGy/frame, DSA: 0.82 (0.64-1.40) mGy/frame. The detector size for dose measurements was 103 systems (55%) with a diagonal of 35 cm or less and 84 systems (45%) with a diagonal of more than 35 cm. The most used FL pulse rate was 7.5 pulse/s (132 systems: 71%), DA frame rate was 7.5 frame/s (72 systems: 39%), and DSA frame rate was 3 frame/s (97 systems: 52%). More than 90% of the institutions used some kind of additional filter for imaging.

Conclusion: This survey showed the standard setting of the angiography system during EVT in Japan.

全国下肢动脉疾病血管内治疗血管造影系统设置调查
目的:本研究的目的是确定日本EVT期间血管造影系统的标准设置。方法:向全日本493家机构邮寄调查请求。调查项目包括透视剂量(FL剂量)/数字血管造影剂量(DA剂量)/数字减影血管造影剂量(DSA剂量)的系统显示剂量设置、检测器尺寸、透视脉冲率、DA/DSA帧率、标准附加滤光器设置。每个项目的有效回复数为:系统显示剂量设置:187个系统(128个机构),检测器大小:187个系统(128个机构),透视脉冲率和DA/DSA帧率:185个系统(127个机构),标准附加过滤器设置:182个系统(125个机构)。结果:各系统显示剂量设置的中位数(四分位数范围)为:FL: 5.9 (4.2-7.8) mGy/min, DA: 0.10 (0.06-0.16) mGy/frame, DSA: 0.82 (0.64-1.40) mGy/frame。用于剂量测量的探测器尺寸为对角线为35厘米或更小的103个系统(55%)和对角线大于35厘米的84个系统(45%)。最常用的FL帧率为7.5帧/s(132个系统:71%),DA帧率为7.5帧/s(72个系统:39%),DSA帧率为3帧/s(97个系统:52%)。超过90%的机构使用某种额外的滤光片进行成像。结论:本调查显示了日本EVT期间血管造影系统的标准设置。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信