Wonjoong Cheon, Young Kyu Lee, Yunji Seol, Chan-Beom Park, Hong Qi Tan, Kyu Hye Choi, Young-Nam Kang, Byung Ock Choi
{"title":"Log file-based quality assurance method for respiratory gating system.","authors":"Wonjoong Cheon, Young Kyu Lee, Yunji Seol, Chan-Beom Park, Hong Qi Tan, Kyu Hye Choi, Young-Nam Kang, Byung Ock Choi","doi":"10.1002/acm2.70101","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>As medical linear accelerator technology advances, enabling higher dose rate deliveries, hypofractionation regimens has increased. This necessitates respiratory gating systems that synchronize radiation delivery with tumor position, requiring simple rigorous quality assurance (QA) to ensure treatment accuracy and patient safety.</p><p><strong>Purpose: </strong>This study aimed to propose log-based QA for respiratory-gated radiation therapy using the respiratory gating system and treatment machine.</p><p><strong>Methods: </strong>4D CT scans were performed with a Varian motion phantom using a Varian Respiratory Gating for Scanner (RGSC). A treatment plan using 25%-75% respiratory phases with 100 MU was created and delivered to a solid water phantom. Treatment logs containing respiratory signals, beam on/off flags, and frame information were extracted from the treatment planning system's offline review. Log file analyses were conducted using in-house softwares to assess temporal synchronization between respiratory phases and beam triggers. Output measurements using a calibrated ion chamber (FC65G) were performed to evaluate dosimetric accuracy. Additionally, EPID images were acquired in cine mode and analyzed frame-by-frame to independently verify beam delivery timing.</p><p><strong>Results: </strong>Log file analysis revealed precise temporal synchronization, with mean time differences of 0.03 s ± 0.05 s between the planned 25% phase and beam-on, and -0.04 s ± 0.05 s between 75% phase and beam-off. The log-derived beam-on duration (2.61 s ± 0.02 s) closely matched the planned duration (2.66 s ± 0.00 s). Three-month log data showed consistent temporal accuracy, with trigger-on times remaining stable at 2.60 s ± 0.01 s across all measurements. Supporting ion chamber measurements confirmed dosimetric agreement between gating and non-gating modes (difference: 0.05 cGy ± 0.09 cGy) CONCLUSIONS: The proposed log file-based QA method demonstrated high accuracy and reproducibility in assessing respiratory gating performance. This approach provides an efficient, objective method for standardizing QA procedures in respiratory-gated radiation therapy, enhancing treatment accuracy and patient safety.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e70101"},"PeriodicalIF":2.0000,"publicationDate":"2025-04-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.70101","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: As medical linear accelerator technology advances, enabling higher dose rate deliveries, hypofractionation regimens has increased. This necessitates respiratory gating systems that synchronize radiation delivery with tumor position, requiring simple rigorous quality assurance (QA) to ensure treatment accuracy and patient safety.
Purpose: This study aimed to propose log-based QA for respiratory-gated radiation therapy using the respiratory gating system and treatment machine.
Methods: 4D CT scans were performed with a Varian motion phantom using a Varian Respiratory Gating for Scanner (RGSC). A treatment plan using 25%-75% respiratory phases with 100 MU was created and delivered to a solid water phantom. Treatment logs containing respiratory signals, beam on/off flags, and frame information were extracted from the treatment planning system's offline review. Log file analyses were conducted using in-house softwares to assess temporal synchronization between respiratory phases and beam triggers. Output measurements using a calibrated ion chamber (FC65G) were performed to evaluate dosimetric accuracy. Additionally, EPID images were acquired in cine mode and analyzed frame-by-frame to independently verify beam delivery timing.
Results: Log file analysis revealed precise temporal synchronization, with mean time differences of 0.03 s ± 0.05 s between the planned 25% phase and beam-on, and -0.04 s ± 0.05 s between 75% phase and beam-off. The log-derived beam-on duration (2.61 s ± 0.02 s) closely matched the planned duration (2.66 s ± 0.00 s). Three-month log data showed consistent temporal accuracy, with trigger-on times remaining stable at 2.60 s ± 0.01 s across all measurements. Supporting ion chamber measurements confirmed dosimetric agreement between gating and non-gating modes (difference: 0.05 cGy ± 0.09 cGy) CONCLUSIONS: The proposed log file-based QA method demonstrated high accuracy and reproducibility in assessing respiratory gating performance. This approach provides an efficient, objective method for standardizing QA procedures in respiratory-gated radiation therapy, enhancing treatment accuracy and patient safety.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic