First experimental test of a coded-mask gamma camera for proton therapy monitoring.

IF 3.3 3区 医学 Q2 ENGINEERING, BIOMEDICAL
Magdalena Kołodziej, Stephan Brons, Mikołaj Dubiel, George N Farah, Alexander Fenger, Ronja Hetzel, Jonas Kasper, Monika Kercz, Barbara Kołodziej, Linn Mielke, Gabriel Ostrzołek, Magdalena Rafecas, Jorge Roser, Katarzyna Rusiecka, Achim Stahl, Vitalii Urbanevych, Ming-Liang Wong, Aleksandra Wrońska
{"title":"First experimental test of a coded-mask gamma camera for proton therapy monitoring.","authors":"Magdalena Kołodziej, Stephan Brons, Mikołaj Dubiel, George N Farah, Alexander Fenger, Ronja Hetzel, Jonas Kasper, Monika Kercz, Barbara Kołodziej, Linn Mielke, Gabriel Ostrzołek, Magdalena Rafecas, Jorge Roser, Katarzyna Rusiecka, Achim Stahl, Vitalii Urbanevych, Ming-Liang Wong, Aleksandra Wrońska","doi":"10.1088/1361-6560/adc96b","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective.</i>The objective of the presented study was to evaluate the feasibility of a coded-mask (CM) gamma camera for real-time range verification in proton therapy, addressing the need for a precise and efficient method of treatment monitoring.<i>Approach.</i>A CM gamma camera prototype was tested in clinical conditions. The setup incorporated a scintillator-based detection system and a structured tungsten collimator. The experiment consisted of the irradiation of PMMA phantom with proton beams of energies ranging from 70.51 to 108.15 MeV. Experimental data were benchmarked against Monte Carlo simulations. The distal falloff position (DFP) was determined for both experimental data and simulations.<i>Main results.</i>The tested CM camera achieved a statistical precision of DFP determination of 1.7 mm for 10<sup>8</sup>protons, which is consistent with simulation predictions, despite hardware limitations such as non-functional detector pixels. Simulations indicated that a fully operational setup would further improve the performance of the detector. The system demonstrated rate capability sufficient for clinical proton beam intensities and maintained performance without significant dead time.<i>Significance.</i>This study validates the potential of the CM gamma camera for real-time proton therapy monitoring. The technology promises to enhance treatment accuracy and patient safety, offering a competitive alternative to existing approaches such as single-slit and multi-slit systems.</p>","PeriodicalId":20185,"journal":{"name":"Physics in medicine and biology","volume":" ","pages":""},"PeriodicalIF":3.3000,"publicationDate":"2025-05-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics in medicine and biology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6560/adc96b","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objective.The objective of the presented study was to evaluate the feasibility of a coded-mask (CM) gamma camera for real-time range verification in proton therapy, addressing the need for a precise and efficient method of treatment monitoring.Approach.A CM gamma camera prototype was tested in clinical conditions. The setup incorporated a scintillator-based detection system and a structured tungsten collimator. The experiment consisted of the irradiation of PMMA phantom with proton beams of energies ranging from 70.51 to 108.15 MeV. Experimental data were benchmarked against Monte Carlo simulations. The distal falloff position (DFP) was determined for both experimental data and simulations.Main results.The tested CM camera achieved a statistical precision of DFP determination of 1.7 mm for 108protons, which is consistent with simulation predictions, despite hardware limitations such as non-functional detector pixels. Simulations indicated that a fully operational setup would further improve the performance of the detector. The system demonstrated rate capability sufficient for clinical proton beam intensities and maintained performance without significant dead time.Significance.This study validates the potential of the CM gamma camera for real-time proton therapy monitoring. The technology promises to enhance treatment accuracy and patient safety, offering a competitive alternative to existing approaches such as single-slit and multi-slit systems.

用于质子治疗监测的编码掩模伽马相机的首次实验测试。
目的:本研究的目的是评估一种编码掩模(CM)伽马相机在质子治疗中用于实时距离验证的可行性,解决了对精确有效的治疗监测方法的需求。CM伽玛相机样机在临床条件下进行了测试。该装置包括一个基于闪烁体的探测系统和一个结构化的钨准直器。实验采用能量为70.51 ~ 108.15 MeV的质子束辐照PMMA模体。实验数据以蒙特卡罗模拟为基准。根据实验数据和仿真结果确定了远端脱落位置。 ;经过测试的CM相机在10^8个质子的情况下实现了1.7 mm的远端衰减位置测定的统计精度,这与模拟预测一致,尽管存在硬件限制,如无功能检测器像素。仿真结果表明,完全可操作的设置将进一步提高探测器的性能。该系统显示出足以满足临床质子束强度的速率能力,并在没有明显死区时间的情况下保持性能。这项研究验证了CM伽马相机在实时质子治疗监测方面的潜力。该技术有望提高治疗的准确性和患者的安全性,为现有的方法(如单缝和多缝系统)提供了一种有竞争力的替代方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Physics in medicine and biology
Physics in medicine and biology 医学-工程:生物医学
CiteScore
6.50
自引率
14.30%
发文量
409
审稿时长
2 months
期刊介绍: The development and application of theoretical, computational and experimental physics to medicine, physiology and biology. Topics covered are: therapy physics (including ionizing and non-ionizing radiation); biomedical imaging (e.g. x-ray, magnetic resonance, ultrasound, optical and nuclear imaging); image-guided interventions; image reconstruction and analysis (including kinetic modelling); artificial intelligence in biomedical physics and analysis; nanoparticles in imaging and therapy; radiobiology; radiation protection and patient dose monitoring; radiation dosimetry
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信