Spatio-Temporal Activation Wavefront Reconstruction From Sparsely Sampled Electrograms.

IF 4.4 2区 医学 Q2 ENGINEERING, BIOMEDICAL
Kostiantyn Ahapov, Peter Ruppersberg, Philip Haeusser, Melissa H Kong, Michael Moeller
{"title":"Spatio-Temporal Activation Wavefront Reconstruction From Sparsely Sampled Electrograms.","authors":"Kostiantyn Ahapov, Peter Ruppersberg, Philip Haeusser, Melissa H Kong, Michael Moeller","doi":"10.1109/TBME.2025.3552356","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Panoramic basket catheters offer a rapid alternative to conventional local activation time (LAT) mapping for visualizing conduction patterns in arrhythmia patients. However, their limited spatial resolution often makes producing highly interpretable visualizations of the underlying phenomena challenging. We propose a novel interpolation method that addresses uncertainties arising from limited spatial resolution and variable tissue contact associated with basket catheters. By leveraging high sampling rates in the time domain to impose spatio-temporal constraints on the reconstruction, we enable enhanced spatial resolution in visualizing wave propagation.</p><p><strong>Methods: </strong>We constructed overlapping triangles from adjacent electrodes. The local apparent conduction velocity (CV) was estimated from cross-correlations of signals, and the triangles were classified into linear conducting, singularity, and non-conducting types based on CV and goodness of fit. Within each triangle, a linear constraint was imposed on the reconstruction, depending on the triangle type, by projecting signals along the CV. A smoothness constraint was added to ensure consistency at clique boundaries. The hyperparameters were calibrated in an unsupervised manner.</p><p><strong>Results: </strong>The proposed method reduced the activation time estimation error by up to 15% compared to traditional interpolation methods in simulations and showed qualitative consistency with LAT maps in clinical cases.</p><p><strong>Conclusion: </strong>The method enhances spatial resolution in panoramic mapping, mitigating aliasing and signal artifacts, especially in regions with complex wavefronts or large inter-electrode distances.</p><p><strong>Significance: </strong>The proposed method enables fast and accurate visualizations of panoramic conduction patterns from a single atrial cycle, potentially reducing procedure times compared to sequential mapping.</p>","PeriodicalId":13245,"journal":{"name":"IEEE Transactions on Biomedical Engineering","volume":"PP ","pages":""},"PeriodicalIF":4.4000,"publicationDate":"2025-04-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1109/TBME.2025.3552356","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Panoramic basket catheters offer a rapid alternative to conventional local activation time (LAT) mapping for visualizing conduction patterns in arrhythmia patients. However, their limited spatial resolution often makes producing highly interpretable visualizations of the underlying phenomena challenging. We propose a novel interpolation method that addresses uncertainties arising from limited spatial resolution and variable tissue contact associated with basket catheters. By leveraging high sampling rates in the time domain to impose spatio-temporal constraints on the reconstruction, we enable enhanced spatial resolution in visualizing wave propagation.

Methods: We constructed overlapping triangles from adjacent electrodes. The local apparent conduction velocity (CV) was estimated from cross-correlations of signals, and the triangles were classified into linear conducting, singularity, and non-conducting types based on CV and goodness of fit. Within each triangle, a linear constraint was imposed on the reconstruction, depending on the triangle type, by projecting signals along the CV. A smoothness constraint was added to ensure consistency at clique boundaries. The hyperparameters were calibrated in an unsupervised manner.

Results: The proposed method reduced the activation time estimation error by up to 15% compared to traditional interpolation methods in simulations and showed qualitative consistency with LAT maps in clinical cases.

Conclusion: The method enhances spatial resolution in panoramic mapping, mitigating aliasing and signal artifacts, especially in regions with complex wavefronts or large inter-electrode distances.

Significance: The proposed method enables fast and accurate visualizations of panoramic conduction patterns from a single atrial cycle, potentially reducing procedure times compared to sequential mapping.

求助全文
约1分钟内获得全文 求助全文
来源期刊
IEEE Transactions on Biomedical Engineering
IEEE Transactions on Biomedical Engineering 工程技术-工程:生物医学
CiteScore
9.40
自引率
4.30%
发文量
880
审稿时长
2.5 months
期刊介绍: IEEE Transactions on Biomedical Engineering contains basic and applied papers dealing with biomedical engineering. Papers range from engineering development in methods and techniques with biomedical applications to experimental and clinical investigations with engineering contributions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信