Quantification of iron in soft tissues through fast kV-switching dual-energy CT imaging: What calibration data are required?

IF 3.3 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Kostas Perisinakis , Angeliki Ntouli , Thomas G. Maris , Apostolos Karantanas
{"title":"Quantification of iron in soft tissues through fast kV-switching dual-energy CT imaging: What calibration data are required?","authors":"Kostas Perisinakis ,&nbsp;Angeliki Ntouli ,&nbsp;Thomas G. Maris ,&nbsp;Apostolos Karantanas","doi":"10.1016/j.ejmp.2025.104973","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To provide recommendation on the type of calibration data needed for quantification of iron content in soft tissues through fast kV-switching dual-energy CT (DECT).</div></div><div><h3>Methods</h3><div>Tissue-specific liquid surrogates mimicking human liver, spleen, kidney and muscles with iron concentration of 0–7 mg/ml were prepared and attached circumferentially to a 16-cm polymethylmethacrylate CT phantom. Soft tissue-equivalent gel boluses were employed to create different in size and configuration phantom-vials setups. Each phantom-vials setup was subjected to fast kV-switching DECT imaging with different acquisition protocols. The virtual iron concentration (VIC) in mg/ml was determined for each vial through the iron(water) material density images. VIC-to-true iron concentration (TIC) curves were derived for four phantom-vials setups and three different acquisition protocols. The applicability of derived VIC-to-TIC calibration data was tested in ten DECT examinations from our institution’s archive.</div></div><div><h3>Results</h3><div>A linear relationship between TIC and VIC values was observed for all tissue-surrogates and phantom-vials setups (<em>R<sup>2</sup></em> &gt; 0.94). The VIC-to-TIC regression-lines derived for different tissues were found to differ significantly (<em>p</em> &lt; 0.05). The regression-lines derived for the same tissue type, but different in size phantom-vials setups were also found to differ significantly (<em>p</em> &lt; 0.05). The effects of DECT acquisition protocol and different vials positioning within the phantom-vials setup on derived regression-lines were found to be minor (<em>p</em> &gt; 0.05).</div></div><div><h3>Conclusions</h3><div>Quantification of iron content through DECT imaging requires tissue- and patient size- specific calibration data. The presented DECT imaging-based method might be useful for monitoring iron levels in patients suspected of iron mis-regulation.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"133 ","pages":"Article 104973"},"PeriodicalIF":3.3000,"publicationDate":"2025-04-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179725000833","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

To provide recommendation on the type of calibration data needed for quantification of iron content in soft tissues through fast kV-switching dual-energy CT (DECT).

Methods

Tissue-specific liquid surrogates mimicking human liver, spleen, kidney and muscles with iron concentration of 0–7 mg/ml were prepared and attached circumferentially to a 16-cm polymethylmethacrylate CT phantom. Soft tissue-equivalent gel boluses were employed to create different in size and configuration phantom-vials setups. Each phantom-vials setup was subjected to fast kV-switching DECT imaging with different acquisition protocols. The virtual iron concentration (VIC) in mg/ml was determined for each vial through the iron(water) material density images. VIC-to-true iron concentration (TIC) curves were derived for four phantom-vials setups and three different acquisition protocols. The applicability of derived VIC-to-TIC calibration data was tested in ten DECT examinations from our institution’s archive.

Results

A linear relationship between TIC and VIC values was observed for all tissue-surrogates and phantom-vials setups (R2 > 0.94). The VIC-to-TIC regression-lines derived for different tissues were found to differ significantly (p < 0.05). The regression-lines derived for the same tissue type, but different in size phantom-vials setups were also found to differ significantly (p < 0.05). The effects of DECT acquisition protocol and different vials positioning within the phantom-vials setup on derived regression-lines were found to be minor (p > 0.05).

Conclusions

Quantification of iron content through DECT imaging requires tissue- and patient size- specific calibration data. The presented DECT imaging-based method might be useful for monitoring iron levels in patients suspected of iron mis-regulation.
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信