Imaging system for real-time, full-field pulse-by-pulse surface dosimetry of UHDR electron beams.

Medical physics Pub Date : 2025-04-01 DOI:10.1002/mp.17784
Megan Clark, Noah Daniel, Petr Bruza, Rongxiao Zhang, Lesley Jarvis, P Jack Hoopes, David Gladstone
{"title":"Imaging system for real-time, full-field pulse-by-pulse surface dosimetry of UHDR electron beams.","authors":"Megan Clark, Noah Daniel, Petr Bruza, Rongxiao Zhang, Lesley Jarvis, P Jack Hoopes, David Gladstone","doi":"10.1002/mp.17784","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The interest in ultra-high dose rate (UHDR) radiation therapy (RT) has grown due to its potential to spare normal tissue. However, clinical application is hindered by dosimetry challenges, as current irradiators and dosimeters are not designed for UHDR's high fluence. To ensure safe treatment and accurate dose delivery, real-time dose and dose rate quantification methods are essential.</p><p><strong>Purpose: </strong>We propose a novel scintillation imaging system for in vivo, pulse-by-pulse surface dose monitoring during delivery with a UHDR-capable Mobetron (IntraOp LLC Sunnyvale, CA, USA) system. This setup aims to measure entrance beam dose with high 2D spatial and temporal resolution.</p><p><strong>Methods: </strong>A modified collimating cone was 3D printed to house the imaging lens. The system featured a 90° sinuscope endoscope attached to a CMOS camera, was gated by the Mobetron's magnetron output signal, and captured light from a scintillator placed on the treatment surface. Three scintillator types were tested for their emission intensity and decay time. Dose and dose rate linearity studies were performed using various pulse lengths and repetition frequencies, respectively, and the imaging data were compared to an EDGE diode detector (SunNuclear Melbourne, FL, USA) and the Mobetron beam-current transformer (BCT) measurements.</p><p><strong>Results: </strong>Dose (R<sup>2 </sup>= 0.993) and dose rate (within 2%) were linear, and the temporal beam structure agreed with the diode and BCT data, as evident by the fact that it was successfully gated such that it captured each pulse during testing. Dose per pulse measurements agreed with diode and BCT data within 2.0 ± 1.2 cGy (0.6% ± 0.3%) and 2.5 ± 1.0 cGy (1.1% ± 0.4%), respectively.</p><p><strong>Conclusions: </strong>The developed imaging system met the criteria for measuring entrance beam dose with high spatial and temporal resolution, offering a promising in vivo dosimetry method for UHDR RT in preclinical and clinical trials.</p>","PeriodicalId":94136,"journal":{"name":"Medical physics","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/mp.17784","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The interest in ultra-high dose rate (UHDR) radiation therapy (RT) has grown due to its potential to spare normal tissue. However, clinical application is hindered by dosimetry challenges, as current irradiators and dosimeters are not designed for UHDR's high fluence. To ensure safe treatment and accurate dose delivery, real-time dose and dose rate quantification methods are essential.

Purpose: We propose a novel scintillation imaging system for in vivo, pulse-by-pulse surface dose monitoring during delivery with a UHDR-capable Mobetron (IntraOp LLC Sunnyvale, CA, USA) system. This setup aims to measure entrance beam dose with high 2D spatial and temporal resolution.

Methods: A modified collimating cone was 3D printed to house the imaging lens. The system featured a 90° sinuscope endoscope attached to a CMOS camera, was gated by the Mobetron's magnetron output signal, and captured light from a scintillator placed on the treatment surface. Three scintillator types were tested for their emission intensity and decay time. Dose and dose rate linearity studies were performed using various pulse lengths and repetition frequencies, respectively, and the imaging data were compared to an EDGE diode detector (SunNuclear Melbourne, FL, USA) and the Mobetron beam-current transformer (BCT) measurements.

Results: Dose (R= 0.993) and dose rate (within 2%) were linear, and the temporal beam structure agreed with the diode and BCT data, as evident by the fact that it was successfully gated such that it captured each pulse during testing. Dose per pulse measurements agreed with diode and BCT data within 2.0 ± 1.2 cGy (0.6% ± 0.3%) and 2.5 ± 1.0 cGy (1.1% ± 0.4%), respectively.

Conclusions: The developed imaging system met the criteria for measuring entrance beam dose with high spatial and temporal resolution, offering a promising in vivo dosimetry method for UHDR RT in preclinical and clinical trials.

求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信